{
  "abstract": "Purpose\n\nUnintended Luer connector disconnection during hemodialysis can cause blood loss and venous air embolism and persists despite Luer-lock mechanisms. The mechanical relationship between insertion force (Fin ) and separation force (Fsp ) remains unclear.Patients and methods\n\nFin and Fsp were measured under dry, axial loading using an automated linear stage and a six-axis force sensor, and a mechanical model based on Coulomb's law of friction was derived. Five male bloodline connector types and seven female needle or cannula connector types, including check valve-equipped cannulas, were tested in 35 combinations. Three female specimens per type were reused across male connectors, yielding 2001 cycles with insertion forces below 50 N. Linear mixed-effects models accounted for repeated measurements within specimens.Results\n\nFsp increased approximately linearly with Fin across all combinations (per-combination slopes, 0.57-0.99). Without a check valve, the pooled slope was 0.855 (95% confidence interval, 0.817-0.894); however, Fsp exceeded Fin in 20% of cycles, primarily at low insertion forces. With a check valve, the relationship had a negative intercept (Fsp = 0.758 Fin - 4.07 N), and certain combinations showed unstable engagement even at insertion forces near 10 N. The model yielded Fsp /Fin = tan(ρ₀ - θ)/tan(ρ' + θ), where ρ' is the insertion friction angle, ρ₀ denotes the static friction angle during engagement, and θ is the Luer taper angle. This ratio is below 1 when friction remains constant. The observed slopes corresponded to static friction coefficients of approximately 0.1-0.4, comparable to reported values for plastics.Conclusion\n\nClinical disconnection risk was not directly quantified. Under controlled dry, axial loading, Luer connectors generally separated at lower forces than the preceding insertion force. Axial retention differed among connector combinations and was lower in check valve-equipped cannulas. Evaluating connection stability may require consideration of mechanical engagement characteristics alongside dimensional standards.",
  "abstract_shown": true,
  "affiliations": [
    "Department of Clinical Engineering, Faculty of Health Care Sciences, Jikei University of Health Care Sciences, Osaka, Japan.",
    "Department of Clinical Engineering, Hanwa Memorial Hospital, Osaka, Japan.",
    "Independent Researcher, Osaka, Japan."
  ],
  "article_type": "research article",
  "arxiv": null,
  "author_total": 2,
  "authors": [
    {
      "affs": [
        0,
        1
      ],
      "name": "Tomohiro Matsui",
      "orcid": "0009-0006-9853-6189"
    },
    {
      "affs": [
        2
      ],
      "name": "Atsushi Sakurai"
    }
  ],
  "bibtex": "https://research.draconitic.com/paper/2d5neb9kbred4.bib",
  "cited_by": 0,
  "doi": "10.2147/mder.s644728",
  "fulltext_url": "https://europepmc.org/articles/PMC13795242",
  "id": "2d5neb9kbred4",
  "issns": [
    "1179-1470"
  ],
  "landing_url": "https://europepmc.org/article/MED/43081112",
  "language": "eng",
  "licence": "CC-BY-NC",
  "licence_label": "CC BY-NC",
  "licence_source": "article",
  "licence_url": "https://creativecommons.org/licenses/by-nc/4.0/",
  "metadata_licence": null,
  "native_subjects": [
    [
      "openalex",
      "2706|Central Venous Catheters and Hemodialysis"
    ]
  ],
  "openalex": "W7236812023",
  "pdf_url": "https://europepmc.org/articles/PMC13795242?pdf=render",
  "pmcid": "PMC13795242",
  "pmid": "43081112",
  "published": "2026-11-20",
  "publisher": null,
  "ris": "https://research.draconitic.com/paper/2d5neb9kbred4.ris",
  "sources": [
    {
      "first_seen": "2026-11-26T15:22:40Z",
      "native_id": "MED:43081112",
      "source": "europepmc",
      "url": "https://europepmc.org/article/MED/43081112"
    }
  ],
  "subjects": [
    {
      "field": "medicine",
      "origin": "openalex:2706|Central Venous Catheters and Hemodialysis",
      "subfield": "general"
    }
  ],
  "title": "Highly Repeatable Measurement and Mechanical Modeling of Axial Insertion and Separation Forces in Hemodialysis Luer Connectors.",
  "title_text": "Highly Repeatable Measurement and Mechanical Modeling of Axial Insertion and Separation Forces in Hemodialysis Luer Connectors.",
  "type": "peer-reviewed",
  "updated_at": "2026-11-26T15:22:40Z",
  "url": "https://research.draconitic.com/paper/2d5neb9kbred4",
  "venue": "Medical devices (Auckland, N.Z.)",
  "versions": []
}