  "abstract": "We report the case of a 73-year-old man with prior endovascular aneurysm repair (EVAR) of the abdominal aorta complicated by type Ia endoleak requiring two proximal aortic cuffs. Ten years later, he underwent transfemoral transcatheter aortic valve replacement (TAVR) for severe aortic stenosis. During the delivery of the transcatheter aortic valve replacement device, one of the aortic cuffs was displaced into the distal thoracic aorta. This case report describes a rare complication of delayed device displacement during TAVR and demonstrates a novel strategy for endovascular salvage. It highlights the opportunity for endovascular repair during TAVR planning in patients with prior aortic stent grafts. In addition, close multidisciplinary collaboration is valuable for managing unexpected device interactions.",
  "abstract_shown": true,
  "affiliations": [
    "Vanderbilt University Medical Center, Nashville, TN",
    "Vanderbilt University School of Medicine, Nashville, TN",
    "Division of Cardiovascular Medicine, Vanderbilt University Medical Center, Nashville, TN",
    "Department of Cardiothoracic Surgery, Vanderbilt University Medical Center, Nashville, TN",
    "Tennessee Valley VA Health System, Nashville, TN",
    "Meharry Medical College, Nashville, TN",
    "Department of Vascular Surgery, Vanderbilt University Medical Center, Nashville, TN"
  ],
  "article_type": "case report",
  "arxiv": null,
  "author_total": 6,
  "authors": [
    {
      "affs": [
        0
      ],
      "name": "Punit Vyas"
    },
    {
      "affs": [
        0
      ],
      "name": "Grant Woodruff"
    },
    {
      "affs": [
        1
      ],
      "name": "Andre Hall"
    },
    {
      "affs": [
        2
      ],
      "name": "Kashish Goel"
    },
    {
      "affs": [
        3
      ],
      "name": "Brian Lima"
    },
    {
      "affs": [
        4,
        5,
        6
      ],
      "name": "John Curci"
    }
  ],
  "bibtex": "https://research.draconitic.com/paper/2fe24abzswdrt.bib",
  "cited_by": 0,
  "doi": "10.1016/j.jvscit.2026.102534",
  "fulltext_url": "https://pmc.ncbi.nlm.nih.gov/articles/PMC13744836/",
  "id": "2fe24abzswdrt",
  "issns": [
    "2468-4287"
  ],
  "landing_url": "https://doi.org/10.1016/j.jvscit.2026.102534",
  "language": "en",
  "licence": "CC-BY-NC-ND-4.0",
  "licence_label": "CC BY-NC-ND 4.0",
  "licence_source": "article",
  "licence_url": "https://creativecommons.org/licenses/by-nc-nd/4.0/",
  "metadata_licence": null,
  "native_subjects": [
    [
      "heading",
      "Complication"
    ],
    [
      "openalex",
      "2705|Cardiac Valve Diseases and Treatments"
    ]
  ],
  "openalex": "W7214784996",
  "pdf_url": "https://www.jvscit.org/article/S2468-4287(26)00418-1/pdf",
  "pmcid": "PMC13744836",
  "pmid": "43001207",
  "published": "2026-09-28",
  "publisher": "Elsevier",
  "ris": "https://research.draconitic.com/paper/2fe24abzswdrt.ris",
  "sources": [
    {
      "first_seen": "2026-11-11T12:22:18Z",
      "native_id": "PMC13744836",
      "source": "pmc",
      "url": "https://doi.org/10.1016/j.jvscit.2026.102534"
    }
  ],
  "subjects": [
    {
      "field": "medicine",
      "origin": "openalex:2705|Cardiac Valve Diseases and Treatments",
      "subfield": "cardiovascular"
    }
  ],
  "title": "Novel endovascular management of a displaced aortic endograft component during transcatheter aortic valve replacement",
  "title_text": "Novel endovascular management of a displaced aortic endograft component during transcatheter aortic valve replacement",
  "type": "peer-reviewed",
  "updated_at": "2026-11-12T12:21:15Z",
  "url": "https://research.draconitic.com/paper/2fe24abzswdrt",
  "venue": "Journal of Vascular Surgery Cases, Innovations and Techniques",
  "versions": []
}