{
  "abstract": "Purpose: Treatment planning systems (TPSs) are critical for ensuring the safety and accuracy of radiation therapy. The International Atomic Energy Agency (IAEA) has developed comprehensive guidelines, such as Technical Reports Series No. 430 and the international Atomic Energy Agency (IAEA) Technical Document 1583 (TECDOC-1583), to support the dosimetric quality control and clinical validation of TPSs. This study aims to validate a TPS using the IAEA’s dosimetric quality assurance methodology and evaluates the feasibility and effectiveness of using locally developed phantoms as alternatives to commercially available phantoms for dosimetric verification of 3-dimensional conformal radiation therapy (3D-CRT) plans. This could provide a cost-effective solution for low-income countries, where the high cost of commercial phantoms can be a barrier.\n\nMethods and Materials: An Elekta Versa HD linear accelerator operating at 6 MV and the Monaco TPS were employed in this study. Dose verification was performed using 3 different phantoms: a commercially available intensity modulated radiation therapy thorax phantom, a custom-made acrylic phantom, and a locally developed beeswax phantom. Supporting tools included an electron density phantom for computed tomography calibration and a calibrated ionization chamber for absolute dose measurements. All testing procedures adhered to IAEA protocols for evaluating dose distribution accuracy in 3D-CRT plans. A comprehensive comparison was conducted to assess the feasibility of using the locally fabricated phantoms as standalone verification tools.\n\nResults: The TPS successfully passed the IAEA dosimetric test cases, with dose measurements generally falling within the IAEA’s ±4% tolerance limit. Both the acrylic and beeswax phantoms showed performance comparable to the commercial phantom, indicating their potential as effective, low-cost alternatives.\n\nConclusions: The IAEA test methodology effectively supports the validation of TPSs for 3D-CRT. The Monaco TPS demonstrated consistent accuracy across all 3 phantoms. Locally fabricated phantoms, particularly in resource-limited settings, can serve as practical and economical tools for TPS quality assurance when implemented in accordance with international guidelines.",
  "abstract_shown": true,
  "affiliations": [
    "Physiology Department, Medical Physics Unit, Faculty of Medicine & Health Sciences, Aden University, Aden, Yemen",
    "Biomedical Physics Department, Faculty of Science & Technology, Al Neelain University, Khartoum, Sudan",
    "Sudan University of Science & Technology, University of Medical Sciences & Technology, Department of Radiology, Khartoum, Sudan",
    "Medical Physics Unit, Radiation Oncology & Nuclear Medicine Department, National Cancer Institute, Cairo University, Cairo, Egypt"
  ],
  "article_type": "research article",
  "arxiv": null,
  "author_total": 5,
  "authors": [
    {
      "affs": [
        0
      ],
      "name": "Naayem Abdulraheem Salem Bakran"
    },
    {
      "affs": [
        1
      ],
      "name": "Ali Sid Ahmed Mohammed Ali"
    },
    {
      "affs": [
        2
      ],
      "name": "Al-Nazer Osman Mohammed Hamza"
    },
    {
      "affs": [
        3
      ],
      "name": "Maha Hassan Mokhtar"
    },
    {
      "affs": [
        3
      ],
      "name": "Nashaat Ahmed Deiab"
    }
  ],
  "bibtex": "https://research.draconitic.com/paper/46xfen3zss93q.bib",
  "cited_by": 0,
  "doi": "10.1016/j.adro.2026.102217",
  "fulltext_url": "https://pmc.ncbi.nlm.nih.gov/articles/PMC13755471/",
  "id": "46xfen3zss93q",
  "issns": [
    "2452-1094"
  ],
  "landing_url": "https://doi.org/10.1016/j.adro.2026.102217",
  "language": "en",
  "licence": "CC-BY-NC-ND-4.0",
  "licence_label": "CC BY-NC-ND 4.0",
  "licence_source": "article",
  "licence_url": "https://creativecommons.org/licenses/by-nc-nd/4.0/",
  "metadata_licence": null,
  "native_subjects": [
    [
      "heading",
      "Scientific Article"
    ],
    [
      "openalex",
      "3108|Advanced Radiotherapy Techniques"
    ]
  ],
  "openalex": "W7213551814",
  "pdf_url": "https://www.advancesradonc.org/article/S2452-1094(26)00223-X/pdf",
  "pmcid": "PMC13755471",
  "pmid": "43022073",
  "published": "2026-09-18",
  "publisher": "Elsevier",
  "ris": "https://research.draconitic.com/paper/46xfen3zss93q.ris",
  "sources": [
    {
      "first_seen": "2026-11-16T12:21:33Z",
      "native_id": "PMC13755471",
      "source": "pmc",
      "url": "https://doi.org/10.1016/j.adro.2026.102217"
    }
  ],
  "subjects": [
    {
      "field": "physics",
      "origin": "openalex:3108|Advanced Radiotherapy Techniques",
      "subfield": "applied-instrumentation"
    }
  ],
  "title": "Dosimetric Verification Using Tissue-Equivalent Phantoms",
  "title_text": "Dosimetric Verification Using Tissue-Equivalent Phantoms",
  "type": "peer-reviewed",
  "updated_at": "2026-11-16T12:21:33Z",
  "url": "https://research.draconitic.com/paper/46xfen3zss93q",
  "venue": "Advances in Radiation Oncology",
  "versions": []
}