  "abstract": "Introduction: The Varian Ethos adaptive radiotherapy system was upgraded from version 1.0 to 2.0, introducing an enhanced multi-leaf collimator (MLC) dose calculation model and a new optimization engine. We evaluated whether this affected agreement between calculated and measured dose, and plan complexity.\n\nMaterial and methods: Twenty-four plans (14 intensity-modulated radiotherapy [IMRT], 10 volumetric modulated arc therapy [VMAT]; 26 measurement configurations) covering 10 anatomical sites were analyzed in three matched configurations: original Ethos 1.0 plans (E1.0), the same plans recalculated in Ethos 2.0 with the new MLC model only (E1.0→2.0), and plans re-optimized in Ethos 2.0 (E2.0). Quality assurance comprised ionization chamber point-dose measurement and global gamma analysis on an ArcCHECK array at 3%/3 mm, 3%/2 mm and 2%/2 mm (10% threshold). Complexity was quantified with UCoMX v1.1; groups were compared with the Wilcoxon signed-rank test.\n\nResults: Mean signed chamber deviation was −1.59 ± 0.54% (E1.0), −1.54 ± 0.57% (E1.0→2.0) and −1.39 ± 0.63% (E2.0), with no significant differences (p > 0.6); all met the ±3% criterion. Gamma pass rates at 3%/3 mm were 99.2 ± 0.9%, 98.7 ± 1.0% (p < 0.001) and 98.6 ± 2.1% (p = 0.059). Re-optimized plans used 12% fewer monitor units (1,169 ± 395 vs. 1,338 ± 471, p < 0.001) and were simpler (MCSv 0.31 to 0.35; SAS10 mm 0.27 to 0.20; both p < 0.001).\n\nConclusion: The upgrade preserved quality-assurance compliance and produced simpler, lower-monitor-unit plans, but the systematic −1.5% offset between calculated and measured point dose persisted, showing the new MLC model alone does not remove it.",
  "abstract_shown": true,
  "affiliations": [
    "Medical Physics Department, Maria Skłodowska-Curie National Research Institute of Oncology, Warsaw, Poland"
  ],
  "article_type": "research article",
  "arxiv": null,
  "author_total": 8,
  "authors": [
    {
      "affs": [
        0
      ],
      "name": "Paulina Wesołowska",
      "orcid": "0000-0003-0889-3408"
    },
    {
      "affs": [
        0
      ],
      "name": "Joanna Iwanowska-Hanke",
      "orcid": "0000-0003-3573-5413"
    },
    {
      "affs": [
        0
      ],
      "name": "Anna Paciorkiewicz",
      "orcid": "0009-0003-4179-6228"
    },
    {
      "affs": [
        0
      ],
      "name": "Sandra Maluszczak"
    },
    {
      "affs": [
        0
      ],
      "name": "Mikołaj Tarchalski"
    },
    {
      "affs": [
        0
      ],
      "name": "Bartłomiej Sadowski",
      "orcid": "0000-0002-7694-6623"
    },
    {
      "affs": [
        0
      ],
      "name": "Marta Fillmann"
    },
    {
      "affs": [
        0
      ],
      "name": "Agnieszka Walewska",
      "orcid": "0000-0002-3645-3736"
    }
  ],
  "bibtex": "https://research.draconitic.com/paper/6mng4f5r2dz32.bib",
  "cited_by": 0,
  "doi": "10.2340/1651-226x.2026.46866",
  "fulltext_url": "https://pmc.ncbi.nlm.nih.gov/articles/PMC13786067/",
  "id": "6mng4f5r2dz32",
  "issns": [
    "0284-186X",
    "1651-226X"
  ],
  "landing_url": "https://doi.org/10.2340/1651-226x.2026.46866",
  "language": "en",
  "licence": "CC-BY-4.0",
  "licence_label": "CC BY 4.0",
  "licence_source": "article",
  "licence_url": "https://creativecommons.org/licenses/by/4.0/",
  "metadata_licence": null,
  "native_subjects": [
    [
      "heading",
      "Original Article"
    ],
    [
      "openalex",
      "3108|Advanced Radiotherapy Techniques"
    ]
  ],
  "openalex": "W7235872529",
  "pdf_url": "https://medicaljournalssweden.se/actaoncologica/article/download/46866/56159",
  "pmcid": "PMC13786067",
  "pmid": "43048850",
  "published": "2026-11-18",
  "publisher": "MJS Publishing",
  "ris": "https://research.draconitic.com/paper/6mng4f5r2dz32.ris",
  "sources": [
    {
      "first_seen": "2026-11-24T12:21:11Z",
      "native_id": "PMC13786067",
      "source": "pmc",
      "url": "https://doi.org/10.2340/1651-226x.2026.46866"
    }
  ],
  "subjects": [
    {
      "field": "physics",
      "origin": "openalex:3108|Advanced Radiotherapy Techniques",
      "subfield": "applied-instrumentation"
    }
  ],
  "title": "Comparison of dosimetric verification results and plan complexity for treatment plans generated in Ethos 1.0 and Ethos 2.0 systems",
  "title_text": "Comparison of dosimetric verification results and plan complexity for treatment plans generated in Ethos 1.0 and Ethos 2.0 systems",
  "type": "peer-reviewed",
  "updated_at": "2026-11-24T12:21:11Z",
  "url": "https://research.draconitic.com/paper/6mng4f5r2dz32",
  "venue": "Acta Oncologica",
  "versions": []
}