  "abstract": "Purpose\n\nTo compare implant positioning and clinical outcomes of MAKO robotic arm-assisted total hip arthroplasty (R-THA) with conventional THA (C-THA) and computed tomography (CT)-based navigated THA (N-THA) in developmental dysplasia of the hip (DDH).Methods\n\nThis systematic review and meta-analysis followed Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) 2020 and the Cochrane Handbook. PubMed, Scopus, Web of Science and Embase were searched up to 30 October 2025. Eligible studies compared MAKO R-THA with C-THA and/or N-THA in DDH patients. Risk of bias was assessed using the Risk of Bias in Non-Randomized Studies of Interventions (ROBINS-I) tool. Meta-analysis was performed with STATA 18.0 using random-effects models. Pooled mean differences (MDs) and risk ratios (RRs) were calculated. Heterogeneity was assessed using the I ² statistic, and the certainty of evidence was evaluated using the Grading of Recommendations, Assessment, Development and Evaluations (GRADE) framework.Results\n\nNine observational studies involving 1130 hips (562 MAKO R-THA, 442 C-THA and 126 N-THA) were included. Compared with C-THA, MAKO R-THA significantly improved horizontal center of rotation (COR) restoration (MD = -1.01 mm, 95% confidence interval [CI] [-1.49 to -0.54]), while no significant differences were observed in operative time, intraoperative blood loss, vertical COR and overall complication rate. Compared with N-THA, MAKO R-THA achieved significantly smaller deviations from planned acetabular inclination (MD = -1.47°, 95% CI [-2.11 to -0.83]) and anteversion (MD = -1.60°, 95% CI [-2.21 to -0.99]).Conclusions\n\nIn patients with DDH, MAKO R-THA demonstrated small, statistically significant improvements in specific radiographic parameters (lower horizontal COR deviation vs. C-THA and smaller inclination and anteversion deviations vs. CT-based N-THA). However, these modest radiographic differences have not been shown to translate into superior clinical function, patient-reported outcomes or lower complication rates. Long-term studies are needed to determine whether these technical precision gains affect implant survivorship.Level of evidence\n\nLevel III.",
  "abstract_shown": true,
  "affiliations": [
    "Student Research Committee, School of Medicine Shahid Beheshti University of Medical Sciences Tehran Iran.",
    "School of Medicine Tehran University of Medical Sciences Tehran Iran."
  ],
  "article_type": "review article",
  "arxiv": null,
  "author_total": 3,
  "authors": [
    {
      "affs": [
        0
      ],
      "name": "Amirhossein Amiri"
    },
    {
      "affs": [
        0
      ],
      "name": "Ariyan Ayati Firoozabadi",
      "orcid": "0009-0000-4363-8010"
    },
    {
      "affs": [
        1
      ],
      "name": "Sadra Mohebbi"
    }
  ],
  "bibtex": "https://research.draconitic.com/paper/8f6acfty9n1ec.bib",
  "cited_by": 0,
  "doi": "10.1002/jeo2.70999",
  "fulltext_url": "https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13794306/?tool=EBI",
  "id": "8f6acfty9n1ec",
  "issns": [
    "2197-1153"
  ],
  "landing_url": "https://europepmc.org/article/PMC/PMC13794306",
  "language": "eng",
  "licence": "CC-BY",
  "licence_label": "CC BY",
  "licence_source": "article",
  "licence_url": "https://creativecommons.org/licenses/by/4.0/",
  "metadata_licence": null,
  "native_subjects": [
    [
      "openalex",
      "2746|Orthopaedic implants and arthroplasty"
    ]
  ],
  "openalex": "W7239352302",
  "pdf_url": "https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13794306/pdf/?tool=EBI",
  "pmcid": "PMC13794306",
  "pmid": "43081034",
  "published": "2026-10-01",
  "publisher": null,
  "ris": "https://research.draconitic.com/paper/8f6acfty9n1ec.ris",
  "sources": [
    {
      "first_seen": "2026-11-26T03:21:04Z",
      "native_id": "PMC:PMC13794306",
      "source": "europepmc",
      "url": "https://europepmc.org/article/PMC/PMC13794306"
    },
    {
      "first_seen": "2026-11-26T15:23:43Z",
      "native_id": "MED:43081034",
      "source": "europepmc",
      "url": "https://europepmc.org/article/MED/43081034"
    }
  ],
  "subjects": [
    {
      "field": "medicine",
      "origin": "openalex:2746|Orthopaedic implants and arthroplasty",
      "subfield": "surgery"
    }
  ],
  "title": "MAKO robotic‐assisted versus computed tomography‐based navigated and conventional total hip arthroplasty in patients with developmental dysplasia of the hip: A systematic review and meta‐analysis",
  "title_text": "MAKO robotic‐assisted versus computed tomography‐based navigated and conventional total hip arthroplasty in patients with developmental dysplasia of the hip: A systematic review and meta‐analysis",
  "type": "peer-reviewed",
  "updated_at": "2026-11-26T15:23:43Z",
  "url": "https://research.draconitic.com/paper/8f6acfty9n1ec",
  "venue": "Journal of experimental orthopaedics",
  "versions": []
}