  "abstract": "Background\n\nBalloon catheter fracture during percutaneous coronary intervention (PCI) is rare but potentially catastrophic. The literature describes three distinct forms: shaft fracture, undeflatable balloon, and entrapped balloon. Published bailout strategies typically address each problem in isolation; however, no specific guidance exists when multiple forms occur simultaneously.Case summary\n\nA 62-year-old man with multivessel coronary disease underwent staged PCI of a left circumflex-obtuse marginal bifurcation. During lesion preparation, a reused semi-compliant balloon (2.5 × 14 mm) fractured after multiple inflations, leaving an inflated, entrapped balloon fragment at the bifurcation with impaired antegrade flow. A stepwise percutaneous bailout was performed: (1) length comparison with an intact reference balloon suggested that a long fragment was partly retained inside the guiding catheter; (2) a guide extension catheter was advanced to displace the fragment distally, achieving passive deflation and restoring flow; and (3) the deflated fragment was retrieved using a 2.0 mm trapping balloon straddling the tip of the guiding catheter. The procedure was completed successfully with stenting, and the patient recovered uneventfully.Discussion\n\nSequential management converted a high-risk situation into a tractable single problem. Distal displacement of the inflated fragment relieved ischaemia before any retrieval attempt was made. Trapping with a second balloon straddling the guide catheter tip may improve grip when the fragment partly protrudes from the guiding catheter.",
  "abstract_shown": true,
  "affiliations": [
    "Department of Interventional Cardiology, Nhan dan Gia Dinh Hospital, 1 No Trang Long Street, Gia Dinh Ward, Ho Chi Minh City 700000, Vietnam."
  ],
  "article_type": null,
  "arxiv": null,
  "author_total": 3,
  "authors": [
    {
      "affs": [
        0
      ],
      "name": "Anh Do Nguyen",
      "orcid": "0009-0003-0631-1555"
    },
    {
      "affs": [
        0
      ],
      "name": "Tien Hoa Phan",
      "orcid": "0009-0006-0852-293X"
    },
    {
      "affs": [
        0
      ],
      "name": "Son Truong Do Chau",
      "orcid": "0009-0001-5610-9967"
    }
  ],
  "bibtex": "https://research.draconitic.com/paper/91sgvrbe6qzff.bib",
  "cited_by": 0,
  "doi": "10.1093/ehjcr/ytag804",
  "fulltext_url": "https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13795251/?tool=EBI",
  "id": "91sgvrbe6qzff",
  "issns": [
    "2514-2119"
  ],
  "landing_url": "https://europepmc.org/article/PMC/PMC13795251",
  "language": "eng",
  "licence": "CC-BY-NC",
  "licence_label": "CC BY-NC",
  "licence_source": "article",
  "licence_url": "https://creativecommons.org/licenses/by-nc/4.0/",
  "metadata_licence": null,
  "native_subjects": [
    [
      "openalex",
      "2705|Coronary Interventions and Diagnostics"
    ]
  ],
  "openalex": "W7223939440",
  "pdf_url": "https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13795251/pdf/?tool=EBI",
  "pmcid": "PMC13795251",
  "pmid": "43080779",
  "published": "2026-10-27",
  "publisher": null,
  "ris": "https://research.draconitic.com/paper/91sgvrbe6qzff.ris",
  "sources": [
    {
      "first_seen": "2026-11-26T03:21:04Z",
      "native_id": "PMC:PMC13795251",
      "source": "europepmc",
      "url": "https://europepmc.org/article/PMC/PMC13795251"
    },
    {
      "first_seen": "2026-11-26T15:23:43Z",
      "native_id": "MED:43080779",
      "source": "europepmc",
      "url": "https://europepmc.org/article/MED/43080779"
    }
  ],
  "subjects": [
    {
      "field": "medicine",
      "origin": "openalex:2705|Coronary Interventions and Diagnostics",
      "subfield": "cardiovascular"
    }
  ],
  "title": "Balloon shaft fracture, deflation failure, and intracoronary entrapment: a stepwise percutaneous bailout",
  "title_text": "Balloon shaft fracture, deflation failure, and intracoronary entrapment: a stepwise percutaneous bailout",
  "type": "peer-reviewed",
  "updated_at": "2026-11-26T15:23:43Z",
  "url": "https://research.draconitic.com/paper/91sgvrbe6qzff",
  "venue": "European heart journal. Case reports",
  "versions": []
}