  "abstract": "Background: Pulse oximetry is widely used to assess sleep disordered breathing and nocturnal hypoxia. Although variability between devices can influence clinical decision-making, agreement during polysomnography (PSG) remains poorly characterised. This study aims to describe agreement and clinical interchangeability between two common pulse oximeters.\n\nMethods: A retrospective cohort study of adults undergoing overnight PSG to determine the agreement between oximeters regarding oxygen desaturation index (ODI) and percentage of total sleep time with oxygen saturation (SpO2) ≤88% (%TST≤88%). Secondary outcomes were agreement of SpO2 (baseline awake, mean sleep and nadir) and the proportion of participants meeting criteria for nocturnal supplemental oxygen with each device. Bland-Altman analyses were conducted with predefined clinically acceptable differences of ±5 events/h for ODI and ±3% for SpO2 determined a priori.\n\nResults: Data from 335 polysomnographic studies (316 participants) were included. Agreement between oximeters for ODI exceeded the maximum acceptable difference (mean difference (MD) 4.8 ±6.7/h, limits of agreement (LOA) -8.4 to 18 events/h). Agreement for %TST ≤88% also fell outside the maximum acceptable difference (MD 6.1 ± 12.5%, LOA -18.5 to 30.5%). Individuals with baseline SpO2 ≤92% measured with either oximeter, were more likely to meet criteria for nocturnal supplemental oxygen therapy with the Nonin WristOx2 compared with the Masimo Radical-7 oximeter (baseline SpO2 ≤92% using Nonin: 37.3% vs 11.9%, p<.001; baseline SpO2 ≤92% using Masimo: 60% vs 24%, p<.004, respectively).\n\nConclusion: Agreement between oximeters was poor. These oximeters should not be used interchangeably as measurement differences may alter classification of disease severity, impacting clinical management.",
  "abstract_shown": true,
  "affiliations": [],
  "article_type": "abstract",
  "arxiv": null,
  "author_total": 8,
  "authors": [
    {
      "name": "J Matthews"
    },
    {
      "name": "O McGuiness"
    },
    {
      "name": "M Milross"
    },
    {
      "name": "A Piper"
    },
    {
      "name": "K Melehan"
    },
    {
      "name": "F Lowrie"
    },
    {
      "name": "A Myint"
    },
    {
      "name": "C Menadue"
    }
  ],
  "bibtex": "https://research.draconitic.com/paper/bv7aapheyf1cd.bib",
  "cited_by": 0,
  "doi": "10.1093/sleepadvances/zpag103.082",
  "fulltext_url": "https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13783417/?tool=EBI",
  "id": "bv7aapheyf1cd",
  "issns": [
    "2632-5012"
  ],
  "landing_url": "https://europepmc.org/article/PMC/PMC13783417",
  "language": "eng",
  "licence": "CC-BY-NC-ND",
  "licence_label": "CC BY-NC-ND",
  "licence_source": "article",
  "licence_url": "https://creativecommons.org/licenses/by-nc-nd/4.0/",
  "metadata_licence": null,
  "native_subjects": [
    [
      "heading",
      "Abstract"
    ],
    [
      "heading",
      "AcademicSubjects/MED00310"
    ],
    [
      "heading",
      "AcademicSubjects/MED00370"
    ],
    [
      "heading",
      "AcademicSubjects/MED00385"
    ],
    [
      "heading",
      "AcademicSubjects/SCI01870"
    ],
    [
      "openalex",
      "2737|Obstructive Sleep Apnea Research"
    ]
  ],
  "openalex": "W7236694716",
  "pdf_url": "https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13783417/pdf/?tool=EBI",
  "pmcid": "PMC13783417",
  "pmid": null,
  "published": "2026-11-01",
  "publisher": "Oxford University Press",
  "ris": "https://research.draconitic.com/paper/bv7aapheyf1cd.ris",
  "sources": [
    {
      "first_seen": "2026-11-21T15:23:17Z",
      "native_id": "PMC:PMC13783417",
      "source": "europepmc",
      "url": "https://europepmc.org/article/PMC/PMC13783417"
    },
    {
      "first_seen": "2026-11-23T12:22:09Z",
      "native_id": "PMC13783417",
      "source": "pmc",
      "url": "https://doi.org/10.1093/sleepadvances/zpag103.082"
    }
  ],
  "subjects": [
    {
      "field": "medicine",
      "origin": "openalex:2737|Obstructive Sleep Apnea Research",
      "subfield": "physiology"
    }
  ],
  "title": "P017. Agreement between two commonly used pulse oximeters during in-laboratory polysomnography",
  "title_text": "P017. Agreement between two commonly used pulse oximeters during in-laboratory polysomnography",
  "type": "peer-reviewed",
  "updated_at": "2026-11-23T12:22:09Z",
  "url": "https://research.draconitic.com/paper/bv7aapheyf1cd",
  "venue": "Sleep advances : a journal of the Sleep Research Society",