  "abstract": "Background: Although implantation strategies for dual-chamber leadless pacemakers are well documented, long-term management, particularly concerning battery longevity, remains under observation. Atrial leadless pacemaker battery life is shorter than that of its ventricular counterpart in real-world observations without clear guidelines established for postimplantation programming.\n\nProject Rationale: We undertook a quality improvement initiative in our outpatient device clinic to address battery longevity in leadless pacemakers via programming approaches targeting 6 optimization areas. A step-by-step algorithm was developed for device clinic staff.\n\nProject Summary: Twenty patients underwent algorithm-driven reprogramming by device clinic staff, with the average battery life increasing by 1.41 years to 5.48 years, a 33.7% improvement. Repeat interrogation at 90 days showed consistent longevity readings (5.4 years).\n\nTake-Home Messages: Dual-chamber leadless pacemakers require refined programming strategies. Implementing a structured algorithm in an outpatient device clinic enabled clinic staff to tailor settings for each patient, resulting in a substantial increase in battery life.",
  "abstract_shown": true,
  "affiliations": [
    "HonorHealth Research Institute and HonorHealth, Scottsdale, Arizona, USA",
    "Arizona State University Edson College of Nursing and Health Innovation, Phoenix, Arizona, USA",
    "Arizona State University John Shufeldt School of Medicine and Medical Engineering, Phoenix, Arizona, USA"
  ],
  "article_type": "case report",
  "arxiv": null,
  "author_total": 4,
  "authors": [
    {
      "affs": [
        0
      ],
      "name": "Amy Kleinhans"
    },
    {
      "affs": [
        0,
        1
      ],
      "name": "Heather M. Ross"
    },
    {
      "affs": [
        0
      ],
      "name": "Fantasia Hudson"
    },
    {
      "affs": [
        0,
        2
      ],
      "name": "Rahul N. Doshi"
    }
  ],
  "bibtex": "https://research.draconitic.com/paper/db6rnef3racq2.bib",
  "cited_by": 0,
  "doi": "10.1016/j.jaccas.2026.109730",
  "fulltext_url": "https://pmc.ncbi.nlm.nih.gov/articles/PMC13749539/",
  "id": "db6rnef3racq2",
  "issns": [
    "2666-0849"
  ],
  "landing_url": "https://doi.org/10.1016/j.jaccas.2026.109730",
  "language": "en",
  "licence": "CC-BY-NC-ND-4.0",
  "licence_label": "CC BY-NC-ND 4.0",
  "licence_source": "article",
  "licence_url": "https://creativecommons.org/licenses/by-nc-nd/4.0/",
  "metadata_licence": null,
  "native_subjects": [
    [
      "heading",
      "Outcomes and Quality"
    ],
    [
      "heading",
      "Quality Improvement Projects"
    ],
    [
      "openalex",
      "2705|Cardiac pacing and defibrillation studies"
    ]
  ],
  "openalex": "W7213348941",
  "pdf_url": null,
  "pmcid": "PMC13749539",
  "pmid": "42747358",
  "published": "2026-09-16",
  "publisher": "Elsevier",
  "ris": "https://research.draconitic.com/paper/db6rnef3racq2.ris",
  "sources": [
    {
      "first_seen": "2026-11-14T12:22:17Z",
      "native_id": "PMC13749539",
      "source": "pmc",
      "url": "https://doi.org/10.1016/j.jaccas.2026.109730"
    }
  ],
  "subjects": [
    {
      "field": "medicine",
      "origin": "openalex:2705|Cardiac pacing and defibrillation studies",
      "subfield": "cardiovascular"
    }
  ],
  "title": "Optimizing Leadless Atrial Pacemaker Performance Through Strategic Reprogramming",
  "title_text": "Optimizing Leadless Atrial Pacemaker Performance Through Strategic Reprogramming",
  "type": "peer-reviewed",
  "updated_at": "2026-11-14T12:22:17Z",
  "url": "https://research.draconitic.com/paper/db6rnef3racq2",
  "venue": "JACC Case Reports",
  "versions": []
}