  "abstract": "A 75‐year‐old man with chronic obstructive pulmonary disease underwent endobronchial valve (EBV) placement in the left lower lobe and expectorated the valve from left B10 1 month later. Retrospective analysis of preprocedural dynamic computed tomography (CT) images showed that the median B10 luminal diameter increased from 4.25 mm at end‐expiration to 6.55 mm at end‐inspiration, approaching the upper limit of the recommended working range (4.0–7.0 mm) for the initially placed size 4.0 valve. A size 5.5 valve was subsequently placed without further expectoration. Bronchoscopic sizing under general anaesthesia with neuromuscular blockade and positive‐pressure ventilation may underestimate the maximum airway calibre encountered during spontaneous breathing, particularly in dependent bronchi. Inspiratory CT‐based airway assessment may provide useful complementary information for planning EBV placement and selecting an appropriate valve size.",
  "abstract_shown": true,
  "affiliations": [
    "Department of Respiratory Medicine Shiga University of Medical Science Otsu Shiga Japan",
    "Department of Radiology Shiga University of Medical Science Otsu Shiga Japan",
    "Health Administration Center Shiga University of Medical Science Otsu Shiga Japan"
  ],
  "article_type": "case study",
  "arxiv": null,
  "author_total": 13,
  "authors": [
    {
      "affs": [
        0
      ],
      "name": "Daisuke Kinose",
      "orcid": "0000-0003-1182-5844"
    },
    {
      "affs": [
        1
      ],
      "name": "Yukihiro Nagatani"
    },
    {
      "affs": [
        0
      ],
      "name": "Hiroaki Nakagawa",
      "orcid": "0000-0003-3830-765X"
    },
    {
      "affs": [
        0
      ],
      "name": "Mayu Fujino"
    },
    {
      "affs": [
        0
      ],
      "name": "Amane Ichida"
    },
    {
      "affs": [
        0
      ],
      "name": "Reo Tanaka"
    },
    {
      "affs": [
        0
      ],
      "name": "Shogo Okuda"
    },
    {
      "affs": [
        0
      ],
      "name": "Yasumitsu Ueki"
    },
    {
      "affs": [
        0
      ],
      "name": "Masashi Misonou"
    },
    {
      "affs": [
        0
      ],
      "name": "Akio Yamazaki"
    },
    {
      "affs": [
        2
      ],
      "name": "Emiko Ogawa"
    },
    {
      "affs": [
        0
      ],
      "name": "Masafumi Yamaguchi",
      "orcid": "0000-0002-0369-324X"
    },
    {
      "affs": [
        0
      ],
      "name": "Yasutaka Nakano",
      "orcid": "0000-0001-7516-8690"
    }
  ],
  "bibtex": "https://research.draconitic.com/paper/drwy4xr8y8wep.bib",
  "cited_by": 0,
  "doi": "10.1002/rcr2.70817",
  "fulltext_url": "https://pmc.ncbi.nlm.nih.gov/articles/PMC13747632/",
  "id": "drwy4xr8y8wep",
  "issns": [
    "2051-3380"
  ],
  "landing_url": "https://doi.org/10.1002/rcr2.70817",
  "language": "en",
  "licence": "CC-BY-4.0",
  "licence_label": "CC BY 4.0",
  "licence_source": "article",
  "licence_url": "https://creativecommons.org/licenses/by/4.0/",
  "metadata_licence": null,
  "native_subjects": [
    [
      "heading",
      "Case Report"
    ],
    [
      "openalex",
      "2740|Chronic Obstructive Pulmonary Disease (COPD) Research"
    ]
  ],
  "openalex": "W7225626373",
  "pdf_url": "https://onlinelibrary.wiley.com/doi/pdfdirect/10.1002/rcr2.70817",
  "pmcid": "PMC13747632",
  "pmid": "43006128",
  "published": "2026-11-09",
  "publisher": "Wiley",
  "ris": "https://research.draconitic.com/paper/drwy4xr8y8wep.ris",
  "sources": [
    {
      "first_seen": "2026-11-12T12:23:39Z",
      "native_id": "PMC13747632",
      "source": "pmc",
      "url": "https://doi.org/10.1002/rcr2.70817"
    },
    {
      "first_seen": "2026-11-12T15:21:47Z",
      "native_id": "MED:43006128",
      "source": "europepmc",
      "url": "https://europepmc.org/article/MED/43006128"
    }
  ],
  "subjects": [
    {
      "field": "medicine",
      "origin": "openalex:2740|Chronic Obstructive Pulmonary Disease (COPD) Research",
      "subfield": "general"
    }
  ],
  "title": "Valve Expectoration After Endobronchial Valve Placement: Potential Utility of Computed Tomography‐Based Airway Sizing—A Case Report",
  "title_text": "Valve Expectoration After Endobronchial Valve Placement: Potential Utility of Computed Tomography‐Based Airway Sizing—A Case Report",
  "type": "peer-reviewed",
  "updated_at": "2026-11-13T12:24:38Z",
  "url": "https://research.draconitic.com/paper/drwy4xr8y8wep",
  "venue": "Respirology Case Reports",
  "versions": []
}