  "abstract": "We describe the case of a 73-year-old man who had a subcutaneous implantable cardioverter-defibrillator (S-ICD) implanted for the secondary prevention of ventricular fibrillation. He received inappropriate shock therapy for S-ICD while using a Japanese electric bath (denki buro). The records of S-ICDs obtained by device interrogation showed that inappropriate therapy was induced by electric waves in the bath. S-ICDs might be more sensitive than transvenous ICD because of the nature of the anatomic position of the electrode. Patients should therefore be carefully educated regarding the risk of inappropriate S-ICD therapy induced by the use of electric baths.",
  "abstract_shown": true,
  "affiliations": [],
  "article_type": null,
  "arxiv": null,
  "author_total": 14,
  "authors": [
    {
      "name": "Motohiro Shingu"
    },
    {
      "name": "Tatsuya Nishikawa"
    },
    {
      "name": "Ryouta Yoshioka"
    },
    {
      "name": "Go Nakano"
    },
    {
      "name": "Atsushi Iseki"
    },
    {
      "name": "Tatsuya Matsubara"
    },
    {
      "name": "Yuya Oga"
    },
    {
      "name": "Hibiki Kadohara"
    },
    {
      "name": "Yuto Osumi"
    },
    {
      "name": "Kenta Ishibashi"
    },
    {
      "name": "Mana Hiraishi"
    },
    {
      "name": "Mitsuo Kinugasa"
    },
    {
      "name": "Yasutaka Hirayama"
    },
    {
      "name": "Koichi Tamita"
    }
  ],
  "bibtex": "https://research.draconitic.com/paper/f2wk7vm56yqzk.bib",
  "cited_by": 0,
  "doi": "10.2169/internalmedicine.6607-25",
  "fulltext_url": "https://www.jstage.jst.go.jp/article/internalmedicine/65/20/65_6607-25/_article",
  "id": "f2wk7vm56yqzk",
  "issns": [
    "1349-7235",
    "0918-2918"
  ],
  "landing_url": "https://www.jstage.jst.go.jp/article/internalmedicine/65/20/65_6607-25/_article",
  "language": null,
  "licence": "CC-BY-NC-ND",
  "licence_label": "CC BY-NC-ND",
  "licence_source": "journal",
  "licence_url": "https://creativecommons.org/licenses/by-nc-nd/4.0/",
  "metadata_licence": null,
  "native_subjects": [
    [
      "jstage-journal",
      "medicine/general"
    ],
    [
      "openalex",
      "2705|Cardiac pacing and defibrillation studies"
    ]
  ],
  "openalex": "W7137045756",
  "pdf_url": null,
  "pmcid": null,
  "pmid": null,
  "published": "2026-10-15",
  "publisher": null,
  "ris": "https://research.draconitic.com/paper/f2wk7vm56yqzk.ris",
  "sources": [
    {
      "first_seen": "2026-10-30T12:17:31Z",
      "native_id": "10.2169/internalmedicine.6607-25",
      "source": "jstage",
      "url": "https://www.jstage.jst.go.jp/article/internalmedicine/65/20/65_6607-25/_article"
    }
  ],
  "subjects": [
    {
      "field": "medicine",
      "origin": "openalex:2705|Cardiac pacing and defibrillation studies",
      "subfield": "cardiovascular"
    }
  ],
  "title": "Inappropriate Therapy of Subcutaneous Implantable Cardioverter-Defibrillator Induced by an Electrical Bath",
  "title_text": "Inappropriate Therapy of Subcutaneous Implantable Cardioverter-Defibrillator Induced by an Electrical Bath",
  "type": "peer-reviewed",
  "updated_at": "2026-10-30T12:17:31Z",
  "url": "https://research.draconitic.com/paper/f2wk7vm56yqzk",
  "venue": "Internal Medicine",
  "versions": []
}