Early Post-operative Fracture of Titanium Cementless Tibial Baseplate: Case Report and Review of the Literature.

  1. 1Department of Orthopaedics and Physical Performance, University of Rochester Medical Center, Rochester, New York, USA.
  2. 2Department of Orthopaedic Surgery, University of Rochester School of Medicine and Dentistry, Rochester, New York, USA.

Abstract

Modern improvements in design and materials have led to a recent resurgence in the popularity of cementless total knee arthroplasty (TKA). Although early catastrophic failure is known to infrequently occur in cemented constructs, it is uncommon in modern designs of cementless knee systems. Here, we report a 70-year-old female with a case of early fracture of the tibial component in a cementless, highly porous titanium-coated baseplate (Triathlon Tritanium, Stryker Orthopaedics), occurring within 20 months of the index procedure. Our case was successfully managed with a cemented tibial cone and stem construct; however, the etiology of failure remains uncertain. Tibial baseplate fracture following TKA with modern cementless titanium three-dimensional -printed porous implants is scarcely reported in the literature, and a high index of suspicion should be maintained in the setting of unexplained pain even at short-term follow-up.

Related

  1. Primary Total Knee Arthroplasty Using a Cementless Highly Porous Titanium Tibial Baseplate in Patients Under the Age of 50: An Average 7.6-year Follow-up
    Arun Nadar, Essa Gul, Ryan Pattyn +4
    Arthroplasty TodayPeer-reviewedCC BY-NC-NDSurgery & anaesthesiaHAL
  2. Fracture Morphology and Extraction-Related Femoral Damages in Short Versus Standard Cementless Stems: Analyses of Vancouver B2 Fractures and Well-Fixed Nonfracture Revisions
    Dominic Simon, Gautier Beckers, Florian Pachmann +5
    Arthroplasty todayPeer-reviewedCC BYFull textSurgery & anaesthesiaEurope PMC
  3. Association of Surgical Approach with Intraoperative Femoral Fracture Patterns During Cementless Hemiarthroplasty for Femoral Neck Fracture
    Ali Tecirli, Seyyid Serif Unsal, Tugrul Yildirim +1
    Journal of clinical medicinePeer-reviewedCC BYFull textSurgery & anaesthesiaEurope PMCCrossref
  4. Recurrent trauma and non-operative management of femoral fracture in fibrodysplasia ossificans progressiva: a pediatric case report
    Bicheng Liu, Xiangyu Lv, Yu Wang +2
    Frontiers in pediatricsPeer-reviewedCC BYFull textGeneral medicineEurope PMC
  5. Management of post-operative delayed proximal tibia and fibula fracture with Ayurvedic medicines and Panchakarma: A case report.
    Ranjit G Nimbalkar, Pratik J Parekar, Anil T Pawar +3
    Journal of Ayurveda and integrative medicinePeer-reviewedCC BY-NC-NDFull textEpidemiology & public healthEurope PMC
BibTeX
@article{jubril2026early,
  title = {{Early Post-operative Fracture of Titanium Cementless Tibial Baseplate: Case Report and Review of the Literature.}},
  author = {Ayodeji Jubril and Adwin Denasty and Ravi Krishnan and Nathan Kaplan and Benjamin Ricciardi},
  journal = {Journal of orthopaedic case reports},
  year = {2026},
  month = nov,
  doi = {10.13107/jocr.2026.v16.i11.8436},
  url = {https://europepmc.org/article/MED/42997051},
  note = {CC BY-NC},
}
RIS
TY  - JOUR
TI  - Early Post-operative Fracture of Titanium Cementless Tibial Baseplate: Case Report and Review of the Literature.
AU  - Ayodeji Jubril
AU  - Adwin Denasty
AU  - Ravi Krishnan
AU  - Nathan Kaplan
AU  - Benjamin Ricciardi
PY  - 2026
DA  - 2026/11/01
JO  - Journal of orthopaedic case reports
DO  - 10.13107/jocr.2026.v16.i11.8436
UR  - https://europepmc.org/article/MED/42997051
AB  - Modern improvements in design and materials have led to a recent resurgence in the popularity of cementless total knee arthroplasty (TKA). Although early catastrophic failure is known to infrequently occur in cemented constructs, it is uncommon in modern designs of cementless knee systems. Here, we report a 70-year-old female with a case of early fracture of the tibial component in a cementless, highly porous titanium-coated baseplate (Triathlon Tritanium, Stryker Orthopaedics), occurring within 20 months of the index procedure. Our case was successfully managed with a cemented tibial cone and stem construct; however, the etiology of failure remains uncertain. Tibial baseplate fracture following TKA with modern cementless titanium three-dimensional -printed porous implants is scarcely reported in the literature, and a high index of suspicion should be maintained in the setting of unexplained pain even at short-term follow-up.
SN  - 2250-0685
SN  - 2321-3817
ER  -