Elevated carbon dioxide in exhaled breath induces false-positive DOB results during Helicobacter pylori detection with ¹³C-urea breath test using infrared spectroscopy.
4 affiliations
- 1Department of Gastroenterology, The First Affiliated Hospital with Nanjing Medical University, No.300 Guangzhou Road, Nanjing, 210029, Jiangsu Province, China.
- 2Department of Gastroenterology, The Fourth Affiliated Hospital of Nanjing Medical University, Nanjing, Jiangsu Province, China.
- 3Department of Gastroenterology, The First Affiliated Hospital with Nanjing Medical University, No.300 Guangzhou Road, Nanjing, 210029, Jiangsu Province, China. [email protected].
- 4Department of Gastroenterology, The First Affiliated Hospital with Nanjing Medical University, No.300 Guangzhou Road, Nanjing, 210029, Jiangsu Province, China. [email protected].
Abstract
Background
The urea breath test (UBT) detects Helicobacter pylori (H. pylori) by measuring ¹³CO₂ in exhaled air after the ingestion of a ¹³C-urea metabolized by bacterial urease. However, the influence of CO₂ concentrations in exhaled breath on UBT results remains unclear. This study aimed to evaluate the effect of varying exhaled CO₂ levels on the outcomes of infrared spectroscopy-based ¹³C-UBT.Methods
Six volunteers confirmed to be H. pylori-negative via ¹³C-UBT and stool antigen assay were enrolled. All participants completed overnight fasting and oral rinsing before specimen collection, followed by standardized breath sampling with three breath-holding durations (5 s, 10 s, and 20 s). Baseline samples were collected at 0 min, followed by a 30-min rest without ¹³C-urea administered, then test samples were collected. Following each sampling cycle, participants rested with spontaneous quiet breathing for 20-30 s before performing the next breath-holding procedure. Each participant completed 30 sampling cycles, yielding 30 paired baseline-test samples per individual. CO₂ concentrations and delta over baseline (DOB) values were measured using three independent infrared spectrometers under calibration settings of 5.0 and 6.0. A DOB value ≥ 4 was defined as a positive result. High CO₂ was defined as a test-sample CO₂ concentration greater than or equal to the corresponding calibration concentration plus 0.2.Results
A total of 180 paired baseline-test datasets were included in the analysis. At calibration concentrations of 5.0, positive results were observed in 104 of 118 high CO₂ datasets (88.14%; 95% exact CI, 80.90%-93.36%), whereas none of the 62 low CO₂ datasets showed positive result (0.00%; 95% exact CI, 0.00%-5.78%). At a calibration concentration of 6.0, positive results occurred in 20 of 26 high-CO₂ datasets (76.92%; 95% exact CI, 56.35%-91.03%), whereas all 154 low CO₂ datasets tested negative (0.00%; 95% exact CI, 0.00%-2.37%). Sample-level Fisher's exact tests were significant at both settings (both P < 0.001).Conclusion
Elevated CO₂ concentrations may be associated with false-positive DOB results in infrared spectroscopy-based ¹³C-UBT, particularly when test-bag CO₂ exceeds the calibration value by more than 0.2 units. Monitoring of exhaled CO₂ during specimen collection is essential to guarantee accurate results.
Related
- Detection of Helicobacter pylori in dental plaque among children with gastrointestinal disorders in a hospital-based cross-sectional study
- Near-Infrared Spectroscopy-Guided Selective Distal Perfusion in Femoral Venoarterial Extracorporeal Membrane Oxygenation: A Feasibility Study
- Near-infrared spectroscopy for rapid and non-destructive assessment of PFAS effects in blue mussels (Mytilus spp.): a proof-of-concept study
- Effects of single-modality sensory stimulation on brain function in patients with prolonged disorders of consciousness: a near-infrared spectroscopy-based study
- Near infrared spectroscopy during non-ischaemic handgrip exercise reveals divergent muscle oxygenation signatures in mitochondrial disease and sarcopenia
BibTeX
@article{fan2026elevated,
title = {{Elevated carbon dioxide in exhaled breath induces false-positive DOB results during Helicobacter pylori detection with ¹³C-urea breath test using infrared spectroscopy.}},
author = {Wentao Fan and Yuwen Tao and Jinjin Shi and Zixiang Huang and MeiHong Chen and Guoxin Zhang and Bo Hao and Feng Ye},
journal = {BMC gastroenterology},
year = {2026},
month = jul,
doi = {10.1186/s12876-026-05169-7},
url = {https://europepmc.org/article/MED/42897461},
note = {CC BY-NC-ND},
}
RIS
TY - JOUR
TI - Elevated carbon dioxide in exhaled breath induces false-positive DOB results during Helicobacter pylori detection with ¹³C-urea breath test using infrared spectroscopy.
AU - Wentao Fan
AU - Yuwen Tao
AU - Jinjin Shi
AU - Zixiang Huang
AU - MeiHong Chen
AU - Guoxin Zhang
AU - Bo Hao
AU - Feng Ye
PY - 2026
DA - 2026/07/01
JO - BMC gastroenterology
DO - 10.1186/s12876-026-05169-7
UR - https://europepmc.org/article/MED/42897461
AB - Background The urea breath test (UBT) detects Helicobacter pylori (H. pylori) by measuring ¹³CO₂ in exhaled air after the ingestion of a ¹³C-urea metabolized by bacterial urease. However, the influence of CO₂ concentrations in exhaled breath on UBT results remains unclear. This study aimed to evaluate the effect of varying exhaled CO₂ levels on the outcomes of infrared spectroscopy-based ¹³C-UBT.Methods Six volunteers confirmed to be H. pylori-negative via ¹³C-UBT and stool antigen assay were enrolled. All participants completed overnight fasting and oral rinsing before specimen collection, followed by standardized breath sampling with three breath-holding durations (5 s, 10 s, and 20 s). Baseline samples were collected at 0 min, followed by a 30-min rest without ¹³C-urea administered, then test samples were collected. Following each sampling cycle, participants rested with spontaneous quiet breathing for 20-30 s before performing the next breath-holding procedure. Each participant completed 30 sampling cycles, yielding 30 paired baseline-test samples per individual. CO₂ concentrations and delta over baseline (DOB) values were measured using three independent infrared spectrometers under calibration settings of 5.0 and 6.0. A DOB value ≥ 4 was defined as a positive result. High CO₂ was defined as a test-sample CO₂ concentration greater than or equal to the corresponding calibration concentration plus 0.2.Results A total of 180 paired baseline-test datasets were included in the analysis. At calibration concentrations of 5.0, positive results were observed in 104 of 118 high CO₂ datasets (88.14%; 95% exact CI, 80.90%-93.36%), whereas none of the 62 low CO₂ datasets showed positive result (0.00%; 95% exact CI, 0.00%-5.78%). At a calibration concentration of 6.0, positive results occurred in 20 of 26 high-CO₂ datasets (76.92%; 95% exact CI, 56.35%-91.03%), whereas all 154 low CO₂ datasets tested negative (0.00%; 95% exact CI, 0.00%-2.37%). Sample-level Fisher's exact tests were significant at both settings (both P < 0.001).Conclusion Elevated CO₂ concentrations may be associated with false-positive DOB results in infrared spectroscopy-based ¹³C-UBT, particularly when test-bag CO₂ exceeds the calibration value by more than 0.2 units. Monitoring of exhaled CO₂ during specimen collection is essential to guarantee accurate results.
SN - 1471-230X
ER -