Elevated carbon dioxide in exhaled breath induces false-positive DOB results during Helicobacter pylori detection with ¹³C-urea breath test using infrared spectroscopy.

4 affiliations
  1. 1Department of Gastroenterology, The First Affiliated Hospital with Nanjing Medical University, No.300 Guangzhou Road, Nanjing, 210029, Jiangsu Province, China.
  2. 2Department of Gastroenterology, The Fourth Affiliated Hospital of Nanjing Medical University, Nanjing, Jiangsu Province, China.
  3. 3Department of Gastroenterology, The First Affiliated Hospital with Nanjing Medical University, No.300 Guangzhou Road, Nanjing, 210029, Jiangsu Province, China. [email protected].
  4. 4Department of Gastroenterology, The First Affiliated Hospital with Nanjing Medical University, No.300 Guangzhou Road, Nanjing, 210029, Jiangsu Province, China. [email protected].

Abstract

Background

The urea breath test (UBT) detects Helicobacter pylori (H. pylori) by measuring ¹³CO₂ in exhaled air after the ingestion of a ¹³C-urea metabolized by bacterial urease. However, the influence of CO₂ concentrations in exhaled breath on UBT results remains unclear. This study aimed to evaluate the effect of varying exhaled CO₂ levels on the outcomes of infrared spectroscopy-based ¹³C-UBT.Methods

Six volunteers confirmed to be H. pylori-negative via ¹³C-UBT and stool antigen assay were enrolled. All participants completed overnight fasting and oral rinsing before specimen collection, followed by standardized breath sampling with three breath-holding durations (5 s, 10 s, and 20 s). Baseline samples were collected at 0 min, followed by a 30-min rest without ¹³C-urea administered, then test samples were collected. Following each sampling cycle, participants rested with spontaneous quiet breathing for 20-30 s before performing the next breath-holding procedure. Each participant completed 30 sampling cycles, yielding 30 paired baseline-test samples per individual. CO₂ concentrations and delta over baseline (DOB) values were measured using three independent infrared spectrometers under calibration settings of 5.0 and 6.0. A DOB value ≥ 4 was defined as a positive result. High CO₂ was defined as a test-sample CO₂ concentration greater than or equal to the corresponding calibration concentration plus 0.2.Results

A total of 180 paired baseline-test datasets were included in the analysis. At calibration concentrations of 5.0, positive results were observed in 104 of 118 high CO₂ datasets (88.14%; 95% exact CI, 80.90%-93.36%), whereas none of the 62 low CO₂ datasets showed positive result (0.00%; 95% exact CI, 0.00%-5.78%). At a calibration concentration of 6.0, positive results occurred in 20 of 26 high-CO₂ datasets (76.92%; 95% exact CI, 56.35%-91.03%), whereas all 154 low CO₂ datasets tested negative (0.00%; 95% exact CI, 0.00%-2.37%). Sample-level Fisher's exact tests were significant at both settings (both P < 0.001).Conclusion

Elevated CO₂ concentrations may be associated with false-positive DOB results in infrared spectroscopy-based ¹³C-UBT, particularly when test-bag CO₂ exceeds the calibration value by more than 0.2 units. Monitoring of exhaled CO₂ during specimen collection is essential to guarantee accurate results.

Related

  1. Detection of Helicobacter pylori in dental plaque among children with gastrointestinal disorders in a hospital-based cross-sectional study
    Mehrzad Sadredinamin, Zahra Jamshidi, Rojin Samani +3
    Scientific ReportsPeer-reviewedCC BY-NC-ND 4.0Full textSurgery & anaesthesiaCrossref
  2. Near-Infrared Spectroscopy-Guided Selective Distal Perfusion in Femoral Venoarterial Extracorporeal Membrane Oxygenation: A Feasibility Study
    Woong-Su Yoon, Yerin Kim, Ju-Seung Kwun +3
    Diagnostics (Basel, Switzerland)Peer-reviewedCC BYFull textGeneral medicineEurope PMC
  3. Near-infrared spectroscopy for rapid and non-destructive assessment of PFAS effects in blue mussels (Mytilus spp.): a proof-of-concept study
    Colleen Guinle, Cécile Levasseur-Garcia, Laurence Poirier +2
    Marine Pollution BulletinPeer-reviewedCC BY 4.0Full textGeneral biochemistryGeneral biologyHAL
  4. Effects of single-modality sensory stimulation on brain function in patients with prolonged disorders of consciousness: a near-infrared spectroscopy-based study
    Juan Zhang, Bin Zhou, Huifang Tian +6
    Frontiers in neurologyPeer-reviewedCC BYFull textNeurologyEurope PMC
  5. Near infrared spectroscopy during non-ischaemic handgrip exercise reveals divergent muscle oxygenation signatures in mitochondrial disease and sarcopenia
    H. Devine, A. Brito da Silva, S. Hillman +12
    medRxivPreprintAll rights reservedFull textNeurologymedRxiv
BibTeX
@article{fan2026elevated,
  title = {{Elevated carbon dioxide in exhaled breath induces false-positive DOB results during Helicobacter pylori detection with ¹³C-urea breath test using infrared spectroscopy.}},
  author = {Wentao Fan and Yuwen Tao and Jinjin Shi and Zixiang Huang and MeiHong Chen and Guoxin Zhang and Bo Hao and Feng Ye},
  journal = {BMC gastroenterology},
  year = {2026},
  month = jul,
  doi = {10.1186/s12876-026-05169-7},
  url = {https://europepmc.org/article/MED/42897461},
  note = {CC BY-NC-ND},
}
RIS
TY  - JOUR
TI  - Elevated carbon dioxide in exhaled breath induces false-positive DOB results during Helicobacter pylori detection with ¹³C-urea breath test using infrared spectroscopy.
AU  - Wentao Fan
AU  - Yuwen Tao
AU  - Jinjin Shi
AU  - Zixiang Huang
AU  - MeiHong Chen
AU  - Guoxin Zhang
AU  - Bo Hao
AU  - Feng Ye
PY  - 2026
DA  - 2026/07/01
JO  - BMC gastroenterology
DO  - 10.1186/s12876-026-05169-7
UR  - https://europepmc.org/article/MED/42897461
AB  - Background  The urea breath test (UBT) detects Helicobacter pylori (H. pylori) by measuring ¹³CO₂ in exhaled air after the ingestion of a ¹³C-urea metabolized by bacterial urease. However, the influence of CO₂ concentrations in exhaled breath on UBT results remains unclear. This study aimed to evaluate the effect of varying exhaled CO₂ levels on the outcomes of infrared spectroscopy-based ¹³C-UBT.Methods  Six volunteers confirmed to be H. pylori-negative via ¹³C-UBT and stool antigen assay were enrolled. All participants completed overnight fasting and oral rinsing before specimen collection, followed by standardized breath sampling with three breath-holding durations (5 s, 10 s, and 20 s). Baseline samples were collected at 0 min, followed by a 30-min rest without ¹³C-urea administered, then test samples were collected. Following each sampling cycle, participants rested with spontaneous quiet breathing for 20-30 s before performing the next breath-holding procedure. Each participant completed 30 sampling cycles, yielding 30 paired baseline-test samples per individual. CO₂ concentrations and delta over baseline (DOB) values were measured using three independent infrared spectrometers under calibration settings of 5.0 and 6.0. A DOB value ≥ 4 was defined as a positive result. High CO₂ was defined as a test-sample CO₂ concentration greater than or equal to the corresponding calibration concentration plus 0.2.Results  A total of 180 paired baseline-test datasets were included in the analysis. At calibration concentrations of 5.0, positive results were observed in 104 of 118 high CO₂ datasets (88.14%; 95% exact CI, 80.90%-93.36%), whereas none of the 62 low CO₂ datasets showed positive result (0.00%; 95% exact CI, 0.00%-5.78%). At a calibration concentration of 6.0, positive results occurred in 20 of 26 high-CO₂ datasets (76.92%; 95% exact CI, 56.35%-91.03%), whereas all 154 low CO₂ datasets tested negative (0.00%; 95% exact CI, 0.00%-2.37%). Sample-level Fisher's exact tests were significant at both settings (both P < 0.001).Conclusion  Elevated CO₂ concentrations may be associated with false-positive DOB results in infrared spectroscopy-based ¹³C-UBT, particularly when test-bag CO₂ exceeds the calibration value by more than 0.2 units. Monitoring of exhaled CO₂ during specimen collection is essential to guarantee accurate results.
SN  - 1471-230X
ER  -