Skip to content
DraconiticResearch
Search Sources API
194,161 studies31 sourcesUpdated 4 min ago
Peer-reviewedCC BYAnnals of vascular diseases

Anatomical Risk Factors for Technical Failure of the Perclose Vascular Closure Device: A Single-Center Retrospective Study.

  1. 1Department of Vascular Surgery, Asahikawa Medical University, Asahikawa, Hokkaido, Japan.
  2. 2Department of Cardiovascular Surgery, Osaka University, Suita, Osaka, Japan.
  3. 3Department of Chronic Limb-Threatening Ischemia Research, Asahikawa Medical University, Asahikawa, Hokkaido, Japan.

Abstract

Objectives

Percutaneous large-bore femoral closure with the Perclose device is widely used, but technical failure remains clinically important. We identified anatomical risk factors for technical failure.Methods

We retrospectively analyzed endovascular aortic repair and endovascular therapy procedures using the Perclose device at the common femoral artery (CFA) between April 2020 and March 2025. Preoperative computed tomography (CT) measured CFA diameter, puncture depth, and a calcification score (0-4). Technical failure was defined as failure to achieve hemostasis with the intended deployment. Independent predictors were identified using patient-clustered generalized estimating equations.Results

Among 300 patients (485 limbs), technical failure occurred in 52 limbs (10.7%), of which 75.0% required surgical repair. Independent predictors were a higher calcification score (odds ratio [OR] 2.13 per 1-point increase; 95% confidence interval [CI] 1.37-3.31), an emergency procedure (OR 3.77; 1.55-9.14), a smaller CFA diameter (OR 1.16 per 1-mm decrease; 1.01-1.32), and a shallower puncture depth (OR 1.22 per 5-mm decrease; 1.01-1.47); the area under the curve was 0.72.Conclusions

Technical failure was driven by unfavorable CFA anatomy-heavy calcification, a small diameter, and a shallow puncture depth-and by emergency procedures. Preprocedural imaging assessment is essential; alternative closure or surgical cutdown should be considered in unfavorable or emergent cases.

Related

  1. Anatomical Predictors of Endovascular Technical Success in Chronic Limb-Threatening Ischemia: The Role of GLASS and Diabetes
    Anca Dumitrescu-Bordianu, Livia Genoveva Baroi, Radu Florin Popa +10
    Medical sciences (Basel, Switzerland)Peer-reviewedCC BYFull textGeneral medicineEurope PMC
  2. Technical feasibility of synchronized multispectral dual-site optical sensing during vascular occlusion in an acute porcine flap model
    Devang Vyas, Jonathan Bova, Michel Saint-Cyr +2
    Scientific ReportsPeer-reviewedCC BY-NC-ND 4.0Full textMedical imaging & radiologyCrossref
  3. Ultrasound-Guided Peristyloid Hydrodissection for Eagle Syndrome With 5% Dextrose: A Technical Note Integrating Computed Tomography-Defined Anatomy, Doppler Vascular Mapping, and Orthogonal Sonographic Guidance
    Yonghyun Yoon, Dongyeun Sung, Jonghyeok Lee +5
    CureusPeer-reviewedCC BYFull textGeneral medicineEurope PMC
  4. Anatomical and Technical Limitations of a Rabbit Median Sternotomy Model for Translational Mediastinal Adhesion Research
    Doğuş Hemşinli, Sedat Ozan Karakişi, Atilla Topçu +3
    Journal of Clinical MedicinePeer-reviewedCC BY 4.0General medicineCrossref
  5. Anatomical comparison populations and threshold transport in vascular transcriptomic classification
    L. You, Z. Peng, Y. Guo +2
    medRxivPreprintCC BYFull textCardiovascular medicinemedRxiv
BibTeX
@article{jinno2026anatomical,
  title = {{Anatomical Risk Factors for Technical Failure of the Perclose Vascular Closure Device: A Single-Center Retrospective Study.}},
  author = {Hirofumi Jinno and Shinsuke Kikuchi and Tsutomu Doita and Izumi Fukii and Kyokei Fuchizawa and Takayuki Uramoto and Seima Ohira and Keisuke Kamada and Naoya Kuriyama and Nobuyoshi Azuma},
  journal = {Annals of vascular diseases},
  year = {2026},
  month = oct,
  doi = {10.3400/avd.oa.26-00118},
  url = {https://europepmc.org/article/MED/42969012},
  note = {CC BY},
}
RIS
TY  - JOUR
TI  - Anatomical Risk Factors for Technical Failure of the Perclose Vascular Closure Device: A Single-Center Retrospective Study.
AU  - Hirofumi Jinno
AU  - Shinsuke Kikuchi
AU  - Tsutomu Doita
AU  - Izumi Fukii
AU  - Kyokei Fuchizawa
AU  - Takayuki Uramoto
AU  - Seima Ohira
AU  - Keisuke Kamada
AU  - Naoya Kuriyama
AU  - Nobuyoshi Azuma
PY  - 2026
DA  - 2026/10/27
JO  - Annals of vascular diseases
DO  - 10.3400/avd.oa.26-00118
UR  - https://europepmc.org/article/MED/42969012
AB  - Objectives  Percutaneous large-bore femoral closure with the Perclose device is widely used, but technical failure remains clinically important. We identified anatomical risk factors for technical failure.Methods  We retrospectively analyzed endovascular aortic repair and endovascular therapy procedures using the Perclose device at the common femoral artery (CFA) between April 2020 and March 2025. Preoperative computed tomography (CT) measured CFA diameter, puncture depth, and a calcification score (0-4). Technical failure was defined as failure to achieve hemostasis with the intended deployment. Independent predictors were identified using patient-clustered generalized estimating equations.Results  Among 300 patients (485 limbs), technical failure occurred in 52 limbs (10.7%), of which 75.0% required surgical repair. Independent predictors were a higher calcification score (odds ratio [OR] 2.13 per 1-point increase; 95% confidence interval [CI] 1.37-3.31), an emergency procedure (OR 3.77; 1.55-9.14), a smaller CFA diameter (OR 1.16 per 1-mm decrease; 1.01-1.32), and a shallower puncture depth (OR 1.22 per 5-mm decrease; 1.01-1.47); the area under the curve was 0.72.Conclusions  Technical failure was driven by unfavorable CFA anatomy-heavy calcification, a small diameter, and a shallow puncture depth-and by emergency procedures. Preprocedural imaging assessment is essential; alternative closure or surgical cutdown should be considered in unfavorable or emergent cases.