Plethysmometry: an automated cuff-based method for estimating ranges of systolic, diastolic and mean arterial pressures

  1. 1Christian Medical College, Vellore

Abstract

BackgroundConventional cuff based methods report single systolic and diastolic pressure values and depend either on Korotkoff sounds or on empirical oscillometric algorithms.

MethodsWe developed a patented cuff based method, termed plethysmometry, that simultaneously records cuff pressure and finger photoplethysmograms from the cuffed and contralateral arms. The contralateral photoplethysmogram serves as an internal physiological reference for pulse amplitude. Following an initial pressure ramp, 16 cuff pressure plateaus are applied. Absolute and normalized pulse amplitudes in the cuffed arm are related to cuff pressure to estimate lower and upper systolic and diastolic pressures. Mean arterial pressure is estimated independently from cuff pressure oscillation amplitude. The estimations are done by automated algorithms developed in MATLAB.

ResultsThe method reports lower and upper systolic, diastolic and mean arterial pressure ranges (during the recording period), together with their average values. Average pulse pressure is also reported. Graphs are provided for reference to check the accuracy of reported pressures.

ConclusionPlethysmometry provides an objective digital record of the cuff pressure and pulse traces and estimates systolic and diastolic pressure ranges from the pulse response of the individual subject to graded cuff pressure. It also reports mean arterial pressure. Comparison of plethysmometric systolic and diastolic pressures with sphygmomanometrically derived pressures and diagnostic agreement are reported in a companion paper.

Related

  1. Association of the Triglyceride–Glucose Index with Systolic and Diastolic Blood Pressure Among Patients With Type 2 Diabetes Mellitus
    Mitha K V, Smruthi V, Pibini Paul
    CureusPeer-reviewedCC BYFull textGeneral medicineEurope PMC
  2. Pak1 overexpression is associated with mild concentric cardiac remodeling and favorable systolic and diastolic functional indices in transgenic mice
    Johnathan Yarbro, Tamayo Kobayashi, Youhua Zhang +4
    Frontiers in cardiovascular medicinePeer-reviewedCC BYFull textCardiovascular medicineEurope PMCCrossref
  3. A Leak before the Beat: Diastolic Atrioventricular Regurgitation in Conduction Disease.
    Thomas Holman, Mikhail Torosoff
    CASE (Philadelphia, Pa.)Peer-reviewedCC BY-NC-NDFull textCardiovascular medicineEurope PMCPMC
  4. Diastolic Musical Murmur From Late Commissural Tear of an Evolut R Valve
    Tomohiko Taniguchi, Toshiaki Toyota, Jiro Esaki +1
    Circulation JournalPeer-reviewedCC BY-NC-NDFull textCardiovascular medicineJ-STAGE
  5. Ubiquitous Sirt1 overexpression is associated with preserved cardiac diastolic function and mitochondrial respiratory capacity in mice on a high-fat diet
    Metabolism openPeer-reviewedCC BYFull textGeneral medicineEurope PMC
BibTeX
@misc{gangadharan2026plethysmometry,
  title = {{Plethysmometry: an automated cuff-based method for estimating ranges of systolic, diastolic and mean arterial pressures}},
  author = {N. Gangadharan and B. Baskaran and A. Manavalan and M. Praveenkumar and M. Anandhakumar and B. Jebaraj and S. Devasahayam and S. Subramani},
  howpublished = {medRxiv},
  year = {2026},
  month = oct,
  doi = {10.64898/2026.10.01.26364471},
  url = {https://www.medrxiv.org/content/10.64898/2026.10.01.26364471v1},
  note = {CC BY-NC-ND},
}
RIS
TY  - UNPB
TI  - Plethysmometry: an automated cuff-based method for estimating ranges of systolic, diastolic and mean arterial pressures
AU  - N. Gangadharan
AU  - B. Baskaran
AU  - A. Manavalan
AU  - M. Praveenkumar
AU  - M. Anandhakumar
AU  - B. Jebaraj
AU  - S. Devasahayam
AU  - S. Subramani
PY  - 2026
DA  - 2026/10/01
PB  - medRxiv
DO  - 10.64898/2026.10.01.26364471
UR  - https://www.medrxiv.org/content/10.64898/2026.10.01.26364471v1
AB  - BackgroundConventional cuff based methods report single systolic and diastolic pressure values and depend either on Korotkoff sounds or on empirical oscillometric algorithms.  MethodsWe developed a patented cuff based method, termed plethysmometry, that simultaneously records cuff pressure and finger photoplethysmograms from the cuffed and contralateral arms. The contralateral photoplethysmogram serves as an internal physiological reference for pulse amplitude. Following an initial pressure ramp, 16 cuff pressure plateaus are applied. Absolute and normalized pulse amplitudes in the cuffed arm are related to cuff pressure to estimate lower and upper systolic and diastolic pressures. Mean arterial pressure is estimated independently from cuff pressure oscillation amplitude. The estimations are done by automated algorithms developed in MATLAB.  ResultsThe method reports lower and upper systolic, diastolic and mean arterial pressure ranges (during the recording period), together with their average values. Average pulse pressure is also reported. Graphs are provided for reference to check the accuracy of reported pressures.  ConclusionPlethysmometry provides an objective digital record of the cuff pressure and pulse traces and estimates systolic and diastolic pressure ranges from the pulse response of the individual subject to graded cuff pressure. It also reports mean arterial pressure. Comparison of plethysmometric systolic and diastolic pressures with sphygmomanometrically derived pressures and diagnostic agreement are reported in a companion paper.
ER  -