Simulation of Shielding Performance of Ta-10W Alloy and Lead in CT Lateral Scattered Radiation Fields
- 1China Southwest Architectural Design and Research Institute Corp. Ltd. General Contracting Engineering Co., Ltd., Chengdu 610000, China
Abstract
Shielding performances of Ta-10W alloy and Pb in CT lateral-scattered radiation fields were compared using NIST XCOM data and MCNP6 models of a tungsten-target source,a PMMA phantom,and flat shielding plates. The air-kerma ratio \begin{document}$B_k was used to characterize transmission. As the tube voltage increased from 80 to 140 kVp,the mean energy of the 90° scattered spectrum increased from 47.01 to 57.95 keV,and the air−kerma contribution from 67.4–88.0 keV increased from 2.40% to 19.14%. For the sameB_k $\end{document} as 2.5 mm Pb,the equivalent Ta-10W thicknesses were 2.01,1.50,1.37,and 1.33 mm at 80,100,120,and 140 kVp,respectively,corresponding to thickness reductions of 19.56%–46.98%. At 100–140 kVp,the areal density was reduced by 10.67%–21.08%. The mass attenuation coefficient of Ta-10W at 80 keV was about 3.15 times that of Pb. The advantage is attributed to enhanced attenuation above the Ta/W K edges and below the Pb K edge,providing a numerical reference for thin local shielding in CT rooms.
Related
- Effect of Sm2O3 Addition on Multifunctional Performance in Si-Zr-B-Based Glass Frits: Mechanical and Radiation Shielding Approach
- Physical, mechanical, and radiation shielding performance of granodiorite for construction applications
- Impact of GCR and SPE on space radiation shielding: OLTARIS simulation and quantum optimization of material selection using quantum algorithms
- A Nano-Bismuth–Silicone Polymer Composite as Shielding for X-Rays in Extreme-Environment Applications: Experimental and Simulation Studies
- Advances in elastomeric composites for radiation shielding: a comprehensive review
BibTeX
@article{teng2026simulation,
title = {{Simulation of Shielding Performance of Ta-10W Alloy and Lead in CT Lateral Scattered Radiation Fields}},
author = {Na TENG},
journal = {Journal of Isotopes},
year = {2026},
month = oct,
doi = {10.7538/tws.2026.youxian.059},
url = {https://doi.org/10.7538/tws.2026.youxian.059},
note = {CC BY-NC-ND},
}
RIS
TY - JOUR
TI - Simulation of Shielding Performance of Ta-10W Alloy and Lead in CT Lateral Scattered Radiation Fields
AU - Na TENG
PY - 2026
DA - 2026/10/26
JO - Journal of Isotopes
DO - 10.7538/tws.2026.youxian.059
UR - https://doi.org/10.7538/tws.2026.youxian.059
AB - Shielding performances of Ta-10W alloy and Pb in CT lateral-scattered radiation fields were compared using NIST XCOM data and MCNP6 models of a tungsten-target source,a PMMA phantom,and flat shielding plates. The air-kerma ratio \begin{document}$B_k was used to characterize transmission. As the tube voltage increased from 80 to 140 kVp,the mean energy of the 90° scattered spectrum increased from 47.01 to 57.95 keV,and the air−kerma contribution from 67.4–88.0 keV increased from 2.40% to 19.14%. For the sameB_k $\end{document} as 2.5 mm Pb,the equivalent Ta-10W thicknesses were 2.01,1.50,1.37,and 1.33 mm at 80,100,120,and 140 kVp,respectively,corresponding to thickness reductions of 19.56%–46.98%. At 100–140 kVp,the areal density was reduced by 10.67%–21.08%. The mass attenuation coefficient of Ta-10W at 80 keV was about 3.15 times that of Pb. The advantage is attributed to enhanced attenuation above the Ta/W K edges and below the Pb K edge,providing a numerical reference for thin local shielding in CT rooms.
SN - 1000-7512
ER -