Guide-sheath-assisted bailout technique for intra-procedural single-leaflet device attachment using the PASCAL system: a case report.

  1. 1Department of Cardiology, Chikamori Hospital, 1-1-16 Okawasuji, Kochi 780-0052, Japan.

Abstract

Single-leaflet device attachment (SLDA) is a rare but serious complication of mitral valve transcatheter edge-to-edge repair (MTEER). When SLDA occurs after full device release, management options are limited. In contrast, intra-procedural SLDA recognized before full implant release may allow a specific bailout strategy. A 77-year-old man with severe degenerative mitral regurgitation (MR) underwent MTEER using the PASCAL Ace system because of high surgical risk. Initial device implantation reduced MR from severe to mild. However, after removal of both clasp lines during the release process, SLDA occurred at the posterior leaflet, resulting in recurrent severe MR. Although standard retrieval was no longer feasible, the implant remained connected to the steerable catheter. Under fluoroscopic and transoesophageal echocardiographic guidance, a guide sheath was advanced coaxially towards the implant. Controlled insertion of the guide sheath between the clasp and paddle, followed by gradual paddle opening, enabled leaflet disengagement and successful device retrieval without leaflet injury. A second implantation achieved a sufficient reduction in mitral regurgitation. This case demonstrates a novel bailout technique for intra-procedural SLDA. Successful retrieval requires precise coaxial device alignment, careful catheter manipulation, continuous monitoring of leaflet tension, and meticulous guidance using fluoroscopy and transoesophageal echocardiography (TEE). Importantly, prompt recognition of SLDA by TEE is a prerequisite for this technique. In conclusion, the guide-sheath-assisted retrieval technique can represent a potential bailout strategy for intra-procedural SLDA before full implant release during MTEER.

Related

  1. Mechanical septal debulking to enable left bundle branch pacing through dense septal scar: A bailout technique.
    Martín Negreira-Caamaño, Agustín Pastor-Fuentes, Joaquin Jesús Alonso-Martín
    HeartRhythm case reportsPeer-reviewedCC BYFull textCardiovascular medicineEurope PMC
  2. Quantifying procedural burden during stent-assisted coil embolization for unruptured intracranial aneurysms: development of the aneurysm endovascular procedural complexity score and internal validation of a multivariable complication model
    Yuxiang Huang, E Zhu, Zijun Huang +4
    Frontiers in NeurologyPeer-reviewedCC BY 4.0NeurologyCrossref
  3. Contrast-enhanced ultrasound-assisted catheter drainage for paravertebral tuberculous abscesses: a retrospective study of procedural safety
    Dan Li, Peijun Chen, Wenzhi Zhang
    Frontiers in medicinePeer-reviewedCC BYFull textSurgery & anaesthesiaEurope PMCPMC
  4. Modified balloon-assisted shoehorn technique for transcatheter aortic valve implantation in functionally horizontal anatomy: a case report
    European heart journal. Case reportsPeer-reviewedCC BYFull textCardiovascular medicineEurope PMC
  5. Safe Ultrasonography‐Assisted Elbow Arthroscopy Technique
    Masato Nakai, Kazuma Miyatake, Takahiro Fujisawa +4
    Arthroscopy techniquesPeer-reviewedCC BYFull textSurgery & anaesthesiaEurope PMC
BibTeX
@article{nukariya2026guidesheathassisted,
  title = {{Guide-sheath-assisted bailout technique for intra-procedural single-leaflet device attachment using the PASCAL system: a case report.}},
  author = {Masato Nukariya and Yuki Nishimura and Yoko Nakaoka and Kazuya Kawai and Hiroki Sugane},
  journal = {European heart journal. Case reports},
  year = {2026},
  month = oct,
  doi = {10.1093/ehjcr/ytag800},
  url = {https://europepmc.org/article/MED/42997296},
  note = {CC BY},
}
RIS
TY  - JOUR
TI  - Guide-sheath-assisted bailout technique for intra-procedural single-leaflet device attachment using the PASCAL system: a case report.
AU  - Masato Nukariya
AU  - Yuki Nishimura
AU  - Yoko Nakaoka
AU  - Kazuya Kawai
AU  - Hiroki Sugane
PY  - 2026
DA  - 2026/10/24
JO  - European heart journal. Case reports
DO  - 10.1093/ehjcr/ytag800
UR  - https://europepmc.org/article/MED/42997296
AB  - Single-leaflet device attachment (SLDA) is a rare but serious complication of mitral valve transcatheter edge-to-edge repair (MTEER). When SLDA occurs after full device release, management options are limited. In contrast, intra-procedural SLDA recognized before full implant release may allow a specific bailout strategy. A 77-year-old man with severe degenerative mitral regurgitation (MR) underwent MTEER using the PASCAL Ace system because of high surgical risk. Initial device implantation reduced MR from severe to mild. However, after removal of both clasp lines during the release process, SLDA occurred at the posterior leaflet, resulting in recurrent severe MR. Although standard retrieval was no longer feasible, the implant remained connected to the steerable catheter. Under fluoroscopic and transoesophageal echocardiographic guidance, a guide sheath was advanced coaxially towards the implant. Controlled insertion of the guide sheath between the clasp and paddle, followed by gradual paddle opening, enabled leaflet disengagement and successful device retrieval without leaflet injury. A second implantation achieved a sufficient reduction in mitral regurgitation. This case demonstrates a novel bailout technique for intra-procedural SLDA. Successful retrieval requires precise coaxial device alignment, careful catheter manipulation, continuous monitoring of leaflet tension, and meticulous guidance using fluoroscopy and transoesophageal echocardiography (TEE). Importantly, prompt recognition of SLDA by TEE is a prerequisite for this technique. In conclusion, the guide-sheath-assisted retrieval technique can represent a potential bailout strategy for intra-procedural SLDA before full implant release during MTEER.
SN  - 2514-2119
ER  -