Comparative evaluation of handheld vs. high-end ultrasound systems for TIPSS hemodynamic assessment: first impressions
5 affiliations
- 1Department of Surgery, University Hospital Regensburg, Regensburg, 93053 Germany
- 2ÜBAG Dr. Vehling-Kaiser MVZ GmbH, Landshut, Germany
- 3Interdisciplinary Ultrasound Center, Department of Radiology, University Hospital Regensburg, Regensburg, Germany
- 4Department of Medical Statistics, University Medical Center Göttingen, Göttingen, Germany
- 5Department of Internal Medicine III, University Hospital Regensburg, Regensburg, Germany
Abstract
Background: Transjugular intrahepatic portosystemic shunt (TIPSS) procedures require reliable post-interventional hemodynamic monitoring to detect shunt dysfunction, for which Doppler ultrasonography is the preferred modality. Owing to the rapid development of handheld ultrasound devices, their diagnostic equivalence to high-end systems warrants investigation.
Materials and methods: A pilot group of 23 patients (n = 13 males, 65%; n = 7 females; 35%; n = 3 unknown) with intrahepatic TIPSS were retrospectively analyzed. Blood flow measurements were obtained and compared using a high-end ultrasound system (Resona R9, Mindray) and modern handheld ultrasound device (Vscan Air™ CL, GE). The TIPSS was evaluated using B-mode and color Doppler sonography. Image quality was assessed using a 5-point scale and flow parameters using statistical analysis. The image quality and the difference between the two devices were analyzed using correlation coefficients and multivariable linear regression models including TIPSS flow values, skin-to-TIPSS distance, and TIPSS dwell time.
Results: Both devices produced clinically interpretable images. However, the high-end device yielded superior image quality in 16 of 23 cases (p < 0.001). Significant correlations were observed between the difference in mean TIPSS flow measurements obtained by the two devices (Δ mean TIPSS flow) and corresponding average flow values measured by the two devices (Kendall’s τ = 0.656; p < 0.001). Significant associations with Δ TIPSS mean flow were identified for skin-to-TIPSS distance (τ = 0.39; p = 0.014) and TIPSS dwell time (τ = −0.39; p = 0.017). In multivariable analysis, both mean TIPSS flow measured by the high-end system and skin-to-TIPSS distance remained significantly associated with Δ TIPSS mean flow. The TIPSS mean flow measured by the high-end system correlated with TIPSS dwell time (τ = 0.38; p = 0.021). Bland–Altman analysis revealed that handheld measurements tended to overestimate low flows and underestimate high flows.
Conclusion: Handheld ultrasound devices show promising potential for TIPSS surveillance; however, they could lead to distorted flow deviations. Contemporary handheld devices may nevertheless be useful in selected point-of-care settings but cannot yet replace high-end systems for precise hemodynamic monitoring. Larger prospective studies are required.
Trial registration: No clinical trial registration was performed, as this study is a single-center retrospective observational analysis based on existing clinical data. The study did not involve prospective patient enrollment, international procedures, or assignment to treatment groups. Therefore, registration in a clinical trial registry was not applicable.
Related
- Diagnostic discordance between referral impressions, clinical assessment, and final electrodiagnostic findings.
- Comparative Evaluation of Microwave- and Ultrasound-Assisted Extraction of Antioxidants from Parsley (Petroselinum crispum (Mill.) Fuss) and Coriander (Coriandrum sativum L.) Processing By-Products.
- Patient Comfort and Preference During Digital Versus Conventional Impressions for Fixed Prosthodontic Rehabilitation: A Comparative Study.
- Before-and-after evaluation of an IDEA-based standardized ultrasound image documentation protocol for deep endometriosis ultrasound assessment.
- Multi-site observational evaluation of Breast AI™-supported ultrasound risk stratification in breast cancer assessment pathways
BibTeX
@article{dropco2026comparative,
title = {{Comparative evaluation of handheld vs. high-end ultrasound systems for TIPSS hemodynamic assessment: first impressions}},
author = {Ivor Dropco and Ursula Vehling-Kaiser and Florian Kaiser and Eleène Kaiser and Sylvia Thurn and Fabian Kück and Wolfgang Herr and Hans Jürgen Schlitt and Christian Stroszczynski and Ernst Michael Jung and Ulrich Kaiser},
journal = {BMC Medical Imaging},
year = {2026},
month = nov,
doi = {10.1186/s12880-026-03011-w},
url = {https://doi.org/10.1186/s12880-026-03011-w},
note = {CC BY 4.0},
}
RIS
TY - JOUR
TI - Comparative evaluation of handheld vs. high-end ultrasound systems for TIPSS hemodynamic assessment: first impressions
AU - Ivor Dropco
AU - Ursula Vehling-Kaiser
AU - Florian Kaiser
AU - Eleène Kaiser
AU - Sylvia Thurn
AU - Fabian Kück
AU - Wolfgang Herr
AU - Hans Jürgen Schlitt
AU - Christian Stroszczynski
AU - Ernst Michael Jung
AU - Ulrich Kaiser
PY - 2026
DA - 2026/11/11
JO - BMC Medical Imaging
DO - 10.1186/s12880-026-03011-w
UR - https://doi.org/10.1186/s12880-026-03011-w
AB - Background: Transjugular intrahepatic portosystemic shunt (TIPSS) procedures require reliable post-interventional hemodynamic monitoring to detect shunt dysfunction, for which Doppler ultrasonography is the preferred modality. Owing to the rapid development of handheld ultrasound devices, their diagnostic equivalence to high-end systems warrants investigation. Materials and methods: A pilot group of 23 patients (n = 13 males, 65%; n = 7 females; 35%; n = 3 unknown) with intrahepatic TIPSS were retrospectively analyzed. Blood flow measurements were obtained and compared using a high-end ultrasound system (Resona R9, Mindray) and modern handheld ultrasound device (Vscan Air™ CL, GE). The TIPSS was evaluated using B-mode and color Doppler sonography. Image quality was assessed using a 5-point scale and flow parameters using statistical analysis. The image quality and the difference between the two devices were analyzed using correlation coefficients and multivariable linear regression models including TIPSS flow values, skin-to-TIPSS distance, and TIPSS dwell time. Results: Both devices produced clinically interpretable images. However, the high-end device yielded superior image quality in 16 of 23 cases (p < 0.001). Significant correlations were observed between the difference in mean TIPSS flow measurements obtained by the two devices (Δ mean TIPSS flow) and corresponding average flow values measured by the two devices (Kendall’s τ = 0.656; p < 0.001). Significant associations with Δ TIPSS mean flow were identified for skin-to-TIPSS distance (τ = 0.39; p = 0.014) and TIPSS dwell time (τ = −0.39; p = 0.017). In multivariable analysis, both mean TIPSS flow measured by the high-end system and skin-to-TIPSS distance remained significantly associated with Δ TIPSS mean flow. The TIPSS mean flow measured by the high-end system correlated with TIPSS dwell time (τ = 0.38; p = 0.021). Bland–Altman analysis revealed that handheld measurements tended to overestimate low flows and underestimate high flows. Conclusion: Handheld ultrasound devices show promising potential for TIPSS surveillance; however, they could lead to distorted flow deviations. Contemporary handheld devices may nevertheless be useful in selected point-of-care settings but cannot yet replace high-end systems for precise hemodynamic monitoring. Larger prospective studies are required. Trial registration: No clinical trial registration was performed, as this study is a single-center retrospective observational analysis based on existing clinical data. The study did not involve prospective patient enrollment, international procedures, or assignment to treatment groups. Therefore, registration in a clinical trial registry was not applicable.
SN - 1471-2342
ER -