Radiographic imaging in pediatric suspected foreign body ingestion: potential for dose reduction through collimation and exposure strategies
- 1Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, Oberdürrbacher Str. 6, Wurzburg, D-97080 Germany
- 2Department of Radiation Oncology, University Hospital Würzburg, Wurzburg, Germany
- 3Department of Pediatrics, University Hospital Würzburg, Wurzburg, Germany
Abstract
Background: Foreign body ingestion constitutes a common emergency in children. Radiography is essential for confirming presence, type, and location of ingested objects. International guidelines recommend broad anatomical coverage from the neck to the pelvis; however, the impact of radiographic exposure parameters and collimation on radiation dose remains underexplored.
Objective: To evaluate the effects of different scan protocols on radiation dose in pediatric foreign body ingestion.
Materials and methods: Two-hundred-twenty-one radiographs from 183 pediatric patients (median age 2.8 years [interquartile range 1.5–5.5 years]; 112 males, 71 females) performed between September 2019 and April 2025 were retrospectively analyzed. Radiographs were categorized as neck–upper abdomen, extended collimation including pelvis, or abdomen only. Notably, the first two protocols used chest radiograph exposure parameters. Data included patient demographics, foreign body type and location, radiographic protocol, dose-area product, and estimated effective dose.
Results: A foreign body was detected in 55.7% of radiographs, most of them in the stomach (59.4%). Coins were most common (31.7%), followed by button batteries (13.8%) and magnets (12.2%). Dose-area products were comparable between standard neck-to-upper-abdomen and extended neck-to-pelvis collimation but were markedly higher for abdomen-only imaging (P<0.001). Effective dose increased with extended collimation (P<0.05) and was highest for abdomen-only radiographs (0.053 mSv [0.034–0.086 mSv], P<0.001). No differences were observed in patient age or repeat imaging rates.
Conclusion: In suspected pediatric foreign body ingestion, collimation and exposure parameters substantially influence radiation dose. Low-dose chest parameters combined with restrictive collimation minimize radiation exposure, whereas abdominal radiographs carry higher dose, attributable predominantly to the abdominal exposure parameters employed.
Graphical abstract:
Related
- Trends in Pediatric Opioid Ingestion and Associated Factors in Washington State, 2014-2023
- Clinical and radiographic improvement in suspected pulmonary lymphangitic carcinomatosis-associated hypoxemic respiratory failure following iruplinalkib-based treatment in ALK-positive lung adenocarcinoma after sequential treatment with ensartinib and lorlatinib: a two-patient case report
- Radiographic Evaluation of Scoliosis in Pediatric, Adolescent, and Young Adult Patients: Associations Between Curve Severity, Localization, and Skeletal Maturity
- Commentary: Accuracy of Clinical Examination in Suspected Pediatric Scaphoid Fractures—A Systematic Review
- Suspected Mesenteric Malperfusion in Acute Type A Aortic Dissection During Pregnancy: A Case Report and Review of Reperfusion‐First Strategies
BibTeX
@article{sturm2026radiographic,
title = {{Radiographic imaging in pediatric suspected foreign body ingestion: potential for dose reduction through collimation and exposure strategies}},
author = {Max-Johann Sturm and Clemens Benoit and Jan-Peter Grunz and Stefan Weick and Franziska Müller-Reichart and Anke Dick and Oliver Andres and Thorsten Alexander Bley and Henner Huflage},
journal = {Pediatric Radiology},
year = {2026},
month = aug,
doi = {10.1007/s00247-026-06751-7},
url = {https://doi.org/10.1007/s00247-026-06751-7},
note = {CC BY 4.0},
}
RIS
TY - JOUR
TI - Radiographic imaging in pediatric suspected foreign body ingestion: potential for dose reduction through collimation and exposure strategies
AU - Max-Johann Sturm
AU - Clemens Benoit
AU - Jan-Peter Grunz
AU - Stefan Weick
AU - Franziska Müller-Reichart
AU - Anke Dick
AU - Oliver Andres
AU - Thorsten Alexander Bley
AU - Henner Huflage
PY - 2026
DA - 2026/08/20
JO - Pediatric Radiology
DO - 10.1007/s00247-026-06751-7
UR - https://doi.org/10.1007/s00247-026-06751-7
AB - Background: Foreign body ingestion constitutes a common emergency in children. Radiography is essential for confirming presence, type, and location of ingested objects. International guidelines recommend broad anatomical coverage from the neck to the pelvis; however, the impact of radiographic exposure parameters and collimation on radiation dose remains underexplored. Objective: To evaluate the effects of different scan protocols on radiation dose in pediatric foreign body ingestion. Materials and methods: Two-hundred-twenty-one radiographs from 183 pediatric patients (median age 2.8 years [interquartile range 1.5–5.5 years]; 112 males, 71 females) performed between September 2019 and April 2025 were retrospectively analyzed. Radiographs were categorized as neck–upper abdomen, extended collimation including pelvis, or abdomen only. Notably, the first two protocols used chest radiograph exposure parameters. Data included patient demographics, foreign body type and location, radiographic protocol, dose-area product, and estimated effective dose. Results: A foreign body was detected in 55.7% of radiographs, most of them in the stomach (59.4%). Coins were most common (31.7%), followed by button batteries (13.8%) and magnets (12.2%). Dose-area products were comparable between standard neck-to-upper-abdomen and extended neck-to-pelvis collimation but were markedly higher for abdomen-only imaging (P<0.001). Effective dose increased with extended collimation (P<0.05) and was highest for abdomen-only radiographs (0.053 mSv [0.034–0.086 mSv], P<0.001). No differences were observed in patient age or repeat imaging rates. Conclusion: In suspected pediatric foreign body ingestion, collimation and exposure parameters substantially influence radiation dose. Low-dose chest parameters combined with restrictive collimation minimize radiation exposure, whereas abdominal radiographs carry higher dose, attributable predominantly to the abdominal exposure parameters employed. Graphical abstract:
SN - 0301-0449
SN - 1432-1998
ER -