Skip to content
DraconiticResearch
Search Sources API
279,231 studies31 sourcesUpdated 5 min ago
Peer-reviewedCC BY 4.0Journal of Applied Clinical Medical Physics

Assessment of simulated respiratory gating based on external‐to‐internal motion prediction using a hybrid, adaptive deep‐learning network

  1. 1Department of Medical Physics Memorial Sloan Kettering Cancer Center New York New York USA
  2. 2Department of Anesthesiology Weill Cornell Medicine New York New York USA
  3. 3Department of Radiology Kentucky University College of Medicine Lexington Kentucky USA

Abstract

Purpose: To assess the accuracy, safety, and efficiency of respiratory‐gated radiotherapy (RGRT) based on internal respiratory waveforms predicted by a hybrid long short‐term memory (LSTM) deep‐learning network, which adapts to breathing irregularities through the application of time‐domain cross‐correlation (TCC) to check and correct any residual phase/time shift in the LSTM predictions.

Methods: The adaptive LSTM‐TCC model was built in two steps: (1) subject‐specific LSTM models were trained to predict internal diaphragm motion from external bellows motion using each of 10 subjects’ concurrent waveforms (≥20 Hz, 3–10 min) acquired from four‐dimensional magnetic resonance imaging (4DMRI) protocol scans and (2) near real‐time TCC was applied to the LSTM predictions to check and correct any residual phase/time shifts to safeguard against new breathing irregularities beyond the training data in the mid‐term (3–10 min) and long‐term (20–30 min) datasets. The testing datasets comprised both the mid‐term and long‐term concurrent motion waveforms, and the external‐to‐internal motion predictions and gating quality assessments for each subject were evaluated and validated with the actual internal motion as the ground truth. The hybrid LSTM‐TCC models were compared with LSTM, TCC, and native external‐to‐internal correlation models. The temporal gating accuracy (beam‐off and beam‐on) and beam‐gating quality (missing‐target rate: %harm, and beam‐on efficiency: %effic) of the hybrid LSTM‐TCC model were evaluated and compared with LSTM, TCC, and native predictions.

Results: The hybrid LSTM‐TCC models produced the highest long‐term sustained correlations (C > 0.8, within 30 min), yielding the lowest %harm = 5 ± 6% and the highest %effic = 67 ± 15%, compared with the LSTM (%harm = 8 ± 8%, %effic = 65 ± 16%), the TCC (%harm = 29 ± 31%, %effic = 43 ± 19%), and the native (%harm = 37 ± 23%, %effic = 38 ± 14%). The hybrid LSTM‐TCC models produced a high, sustained performance (C = 0.9 ± 0.1) without remodeling, regardless of new breathing irregularities, and the average temporal error of the predicted gating windows was ∆t < 0.3 s.

Conclusion: This study assesses the simulated respiratory‐gating qualities of four methods, demonstrating that the hybrid LSTM‐TCC approach outperforms the native waveforms, the TCC method, and the LSTM models, yielding the highest correlation, gating accuracy, and %effic, but the least %harm. The hybrid LSTM‐TCC model can safeguard LSTM predictions and gating accuracy against major breathing irregularities, demonstrating a strong potential for future clinical applications.

Related

  1. The Source of Disturbance Matters: External, Internal, and Control-Generated Noise in Adaptive Regulation
    Veronique Ziegler
    arXivPreprintarXiv licenceFull textStatistical & nonlinear physicsarXiv
  2. Complete characterization of internal-external entanglement trade-offs in two-mode Gaussian states
    Yasusada Nambu
    arXivPreprintCC BY 4.0Full textGravitation & general relativityQuantum physics & quantum informationarXiv
  3. External forces and moments prediction during obstacle crossing: Evaluation and perspectives for military field applications
    Aurélien Schuster, Charles Pontonnier, Georges Dumont
    Multidisciplinary Biomechanics JournalPeer-reviewedCC BY 4.0Full textGeneral physicsHAL
  4. Enhancing Extubation Failure Prediction with LLM-Derived Features from Respiratory Therapy Clinical Notes
    Izzy Chaiken, Aditya Khowal, Neha A. Sathe +2
    arXivPreprintCC BY 4.0Full textStatistical & nonlinear physicsarXiv
  5. Dynamic control of carbon-ion radiation dose distribution amid respiratory motion challenges: A review of prediction research using geant4 and deep learning physics-constrained models.
    Jianhua Ding, Peng Li, Chai Hong Yeong
    Journal of applied clinical medical physicsPeer-reviewedCC BYFull textApplied physics & instrumentationEurope PMC
BibTeX
@article{milewski2026assessment,
  title = {{Assessment of simulated respiratory gating based on external‐to‐internal motion prediction using a hybrid, adaptive deep‐learning network}},
  author = {Andrew Milewski and Xingyu Nie and Guang Li},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  month = nov,
  doi = {10.1002/acm2.70905},
  url = {https://doi.org/10.1002/acm2.70905},
  note = {CC BY 4.0},
}
RIS
TY  - JOUR
TI  - Assessment of simulated respiratory gating based on external‐to‐internal motion prediction using a hybrid, adaptive deep‐learning network
AU  - Andrew Milewski
AU  - Xingyu Nie
AU  - Guang Li
PY  - 2026
DA  - 2026/11/17
JO  - Journal of Applied Clinical Medical Physics
DO  - 10.1002/acm2.70905
UR  - https://doi.org/10.1002/acm2.70905
AB  - Purpose: To assess the accuracy, safety, and efficiency of respiratory‐gated radiotherapy (RGRT) based on internal respiratory waveforms predicted by a hybrid long short‐term memory (LSTM) deep‐learning network, which adapts to breathing irregularities through the application of time‐domain cross‐correlation (TCC) to check and correct any residual phase/time shift in the LSTM predictions.  Methods: The adaptive LSTM‐TCC model was built in two steps: (1) subject‐specific LSTM models were trained to predict internal diaphragm motion from external bellows motion using each of 10 subjects’ concurrent waveforms (≥20 Hz, 3–10 min) acquired from four‐dimensional magnetic resonance imaging (4DMRI) protocol scans and (2) near real‐time TCC was applied to the LSTM predictions to check and correct any residual phase/time shifts to safeguard against new breathing irregularities beyond the training data in the mid‐term (3–10 min) and long‐term (20–30 min) datasets. The testing datasets comprised both the mid‐term and long‐term concurrent motion waveforms, and the external‐to‐internal motion predictions and gating quality assessments for each subject were evaluated and validated with the actual internal motion as the ground truth. The hybrid LSTM‐TCC models were compared with LSTM, TCC, and native external‐to‐internal correlation models. The temporal gating accuracy (beam‐off and beam‐on) and beam‐gating quality (missing‐target rate: %harm, and beam‐on efficiency: %effic) of the hybrid LSTM‐TCC model were evaluated and compared with LSTM, TCC, and native predictions.  Results: The hybrid LSTM‐TCC models produced the highest long‐term sustained correlations (C > 0.8, within 30 min), yielding the lowest %harm = 5 ± 6% and the highest %effic = 67 ± 15%, compared with the LSTM (%harm = 8 ± 8%, %effic = 65 ± 16%), the TCC (%harm = 29 ± 31%, %effic = 43 ± 19%), and the native (%harm = 37 ± 23%, %effic = 38 ± 14%). The hybrid LSTM‐TCC models produced a high, sustained performance (C = 0.9 ± 0.1) without remodeling, regardless of new breathing irregularities, and the average temporal error of the predicted gating windows was ∆t < 0.3 s.  Conclusion: This study assesses the simulated respiratory‐gating qualities of four methods, demonstrating that the hybrid LSTM‐TCC approach outperforms the native waveforms, the TCC method, and the LSTM models, yielding the highest correlation, gating accuracy, and %effic, but the least %harm. The hybrid LSTM‐TCC model can safeguard LSTM predictions and gating accuracy against major breathing irregularities, demonstrating a strong potential for future clinical applications.
SN  - 1526-9914
ER  -