Pressure behavior in a roller pump-driven low-flow VACC mock circuit with inactive de-airing pump: A descriptive bench characterization under reduced patient-side effective filling pressure conditions
6 affiliations
- 1Department of Medical Technology, Faculty of Fukuoka Medical Technology, Teikyo University, Omuta, Fukuoka, Japan
- 2Laboratory of Veterinary Surgery, Joint Faculty of Veterinary Medicine, Yamaguchi University, Yamaguchi, Yamaguchi, Japan
- 3School of Clinical Engineering, Faculty of Health and Medical Care, Saitama Medical University, Hidaka, Saitama, Japan
- 4Department of Clinical Engineering, National Hospital Organization Sendai Medical Center, Sendai, Japan
- 5Department of Clinical Engineering, Saitama Children’s Medical Center, Saitama, Saitama, Japan
- 6Graduate Degree Program of Comprehensive Applied Data Science, Teikyo University, Hachioji, Tokyo, Japan
Abstract
This exploratory bench study characterized pressure behavior in a roller pump-driven low-flow valve-type semi-closed extracorporeal circulation (VACC) mock circuit under experimentally reduced patient-side effective filling pressure conditions with the de-airing pump inactive. Pressures were measured at four sites: circuit-reservoir pressure (Pc), patient-side mock-reservoir outlet pressure (Pv), roller pump inlet pressure (Pi), and roller pump outlet pressure (Po). Pump flow-rate settings were 50–300 mL/min. Circuit-reservoir negative-pressure setpoints were 0, −10, −20, −30, and −40 mmHg, and patient-side mock-reservoir negative-pressure setpoints were 0, −5, and −10 mmHg. A 32 wt% aqueous glycerol solution was used. Pc generally reflected the circuit-reservoir negative-pressure setpoint, with a mean difference of −0.18 mmHg (95% CI, −0.37 to 0.00). The 95% CIs summarize variability among technical replicates within this bench system and should not be interpreted as population-inferential estimates. Pv decreased stepwise according to the patient-side setpoint. Within the common evaluation range, mean differences versus 0 mmHg were −5.7 mmHg (95% CI, −6.0 to −5.3) for the −5 mmHg condition and −9.2 mmHg (95% CI, −9.5 to −8.9) for the −10 mmHg condition. Pi shifted in the negative direction as the pump flow-rate setting increased, whereas Po increased monotonically under the fixed occluder condition. Delivered flow was independently verified volumetrically before the main experiments at 100, 200, and 300 mL/min but was not continuously measured during the pressure experiments. Therefore, Po was interpreted primarily in relation to the pump flow-rate setting rather than as a directly measured hydraulic pressure–flow relationship. These data provide baseline pressure-characterization information for future mock-circuit studies and do not establish clinical safety, efficacy, or a stable operating range.
Related
- Fatigue Behavior and Multiscale Characterization of 100% RAP Warm-Mix Recycled Asphalt Mixtures Under Multifactorial Loading Conditions
- Microencapsulation of Flaxseed Oil Using Chitosan–Gum Arabic Biopolymers: Formulation, Characterization, Stability, and Release Behavior Under Simulated Gastric and Intestinal Conditions
- Operating Conditions Redefine MOF Design: Interpretable Windows Expose Cross-Process Conflict and Pressure-Driven Rank Reversal
- High-Pressure Carbon Dioxide Adsorption and Dynamic Breathing Behavior in a Flexible Zn2L2 Metallosupramolecular Mesocate
- Experimental assessment of efficiency degradation and corrosion behavior of centrifugal pumps under laboratory-controlled high-salinity conditions
BibTeX
@article{wauchi2026pressure,
title = {{Pressure behavior in a roller pump-driven low-flow VACC mock circuit with inactive de-airing pump: A descriptive bench characterization under reduced patient-side effective filling pressure conditions}},
author = {Keizo Wauchi and Takahiro Okumura and Keisuke Matsuda and Hiroki Sato and Masahiro Miyata and Kenji Tani},
journal = {PLOS One},
year = {2026},
month = nov,
doi = {10.1371/journal.pone.0363658},
url = {https://doi.org/10.1371/journal.pone.0363658},
note = {CC BY 4.0},
}
RIS
TY - JOUR
TI - Pressure behavior in a roller pump-driven low-flow VACC mock circuit with inactive de-airing pump: A descriptive bench characterization under reduced patient-side effective filling pressure conditions
AU - Keizo Wauchi
AU - Takahiro Okumura
AU - Keisuke Matsuda
AU - Hiroki Sato
AU - Masahiro Miyata
AU - Kenji Tani
PY - 2026
DA - 2026/11/25
JO - PLOS One
DO - 10.1371/journal.pone.0363658
UR - https://doi.org/10.1371/journal.pone.0363658
AB - This exploratory bench study characterized pressure behavior in a roller pump-driven low-flow valve-type semi-closed extracorporeal circulation (VACC) mock circuit under experimentally reduced patient-side effective filling pressure conditions with the de-airing pump inactive. Pressures were measured at four sites: circuit-reservoir pressure (Pc), patient-side mock-reservoir outlet pressure (Pv), roller pump inlet pressure (Pi), and roller pump outlet pressure (Po). Pump flow-rate settings were 50–300 mL/min. Circuit-reservoir negative-pressure setpoints were 0, −10, −20, −30, and −40 mmHg, and patient-side mock-reservoir negative-pressure setpoints were 0, −5, and −10 mmHg. A 32 wt% aqueous glycerol solution was used. Pc generally reflected the circuit-reservoir negative-pressure setpoint, with a mean difference of −0.18 mmHg (95% CI, −0.37 to 0.00). The 95% CIs summarize variability among technical replicates within this bench system and should not be interpreted as population-inferential estimates. Pv decreased stepwise according to the patient-side setpoint. Within the common evaluation range, mean differences versus 0 mmHg were −5.7 mmHg (95% CI, −6.0 to −5.3) for the −5 mmHg condition and −9.2 mmHg (95% CI, −9.5 to −8.9) for the −10 mmHg condition. Pi shifted in the negative direction as the pump flow-rate setting increased, whereas Po increased monotonically under the fixed occluder condition. Delivered flow was independently verified volumetrically before the main experiments at 100, 200, and 300 mL/min but was not continuously measured during the pressure experiments. Therefore, Po was interpreted primarily in relation to the pump flow-rate setting rather than as a directly measured hydraulic pressure–flow relationship. These data provide baseline pressure-characterization information for future mock-circuit studies and do not establish clinical safety, efficacy, or a stable operating range.
SN - 1932-6203
ER -