Pressure behavior in a roller pump-driven low-flow VACC mock circuit with inactive de-airing pump: A descriptive bench characterization under reduced patient-side effective filling pressure conditions

6 affiliations
  1. 1Department of Medical Technology, Faculty of Fukuoka Medical Technology, Teikyo University, Omuta, Fukuoka, Japan
  2. 2Laboratory of Veterinary Surgery, Joint Faculty of Veterinary Medicine, Yamaguchi University, Yamaguchi, Yamaguchi, Japan
  3. 3School of Clinical Engineering, Faculty of Health and Medical Care, Saitama Medical University, Hidaka, Saitama, Japan
  4. 4Department of Clinical Engineering, National Hospital Organization Sendai Medical Center, Sendai, Japan
  5. 5Department of Clinical Engineering, Saitama Children’s Medical Center, Saitama, Saitama, Japan
  6. 6Graduate Degree Program of Comprehensive Applied Data Science, Teikyo University, Hachioji, Tokyo, Japan

Abstract

This exploratory bench study characterized pressure behavior in a roller pump-driven low-flow valve-type semi-closed extracorporeal circulation (VACC) mock circuit under experimentally reduced patient-side effective filling pressure conditions with the de-airing pump inactive. Pressures were measured at four sites: circuit-reservoir pressure (Pc), patient-side mock-reservoir outlet pressure (Pv), roller pump inlet pressure (Pi), and roller pump outlet pressure (Po). Pump flow-rate settings were 50–300 mL/min. Circuit-reservoir negative-pressure setpoints were 0, −10, −20, −30, and −40 mmHg, and patient-side mock-reservoir negative-pressure setpoints were 0, −5, and −10 mmHg. A 32 wt% aqueous glycerol solution was used. Pc generally reflected the circuit-reservoir negative-pressure setpoint, with a mean difference of −0.18 mmHg (95% CI, −0.37 to 0.00). The 95% CIs summarize variability among technical replicates within this bench system and should not be interpreted as population-inferential estimates. Pv decreased stepwise according to the patient-side setpoint. Within the common evaluation range, mean differences versus 0 mmHg were −5.7 mmHg (95% CI, −6.0 to −5.3) for the −5 mmHg condition and −9.2 mmHg (95% CI, −9.5 to −8.9) for the −10 mmHg condition. Pi shifted in the negative direction as the pump flow-rate setting increased, whereas Po increased monotonically under the fixed occluder condition. Delivered flow was independently verified volumetrically before the main experiments at 100, 200, and 300 mL/min but was not continuously measured during the pressure experiments. Therefore, Po was interpreted primarily in relation to the pump flow-rate setting rather than as a directly measured hydraulic pressure–flow relationship. These data provide baseline pressure-characterization information for future mock-circuit studies and do not establish clinical safety, efficacy, or a stable operating range.

Related

  1. Fatigue Behavior and Multiscale Characterization of 100% RAP Warm-Mix Recycled Asphalt Mixtures Under Multifactorial Loading Conditions
    Xin Jin, Zhixing Pan, Yanhai Yang +3
    PolymersPeer-reviewedCC BY 4.0Full textMaterials & polymersCrossrefEurope PMC
  2. Microencapsulation of Flaxseed Oil Using Chitosan–Gum Arabic Biopolymers: Formulation, Characterization, Stability, and Release Behavior Under Simulated Gastric and Intestinal Conditions
    Rahmani Fathia, Louadj Amel, Omar Larbi Bouamrane +7
    PolymersPeer-reviewedCC BY 4.0Full textMaterials & polymersCrossrefEurope PMC
  3. Operating Conditions Redefine MOF Design: Interpretable Windows Expose Cross-Process Conflict and Pressure-Driven Rank Reversal
    Hosein Alimardani, Mehrdad Asgari
    ChemRxivPreprintLicence not statedGeneral chemistryChemRxiv
  4. High-Pressure Carbon Dioxide Adsorption and Dynamic Breathing Behavior in a Flexible Zn2L2 Metallosupramolecular Mesocate
    Chana Panyanon, Raúl Díaz-Torres, Pacharapon Jearanaiwiwat +4
    ACS OmegaPeer-reviewedCC BY-NC-ND 4.0Full textGeneral chemistryCrossrefEurope PMCEurope PMCPMC
  5. Experimental assessment of efficiency degradation and corrosion behavior of centrifugal pumps under laboratory-controlled high-salinity conditions
    Nsini Ignatius Udo
    Journal of Engineering and Applied SciencePeer-reviewedCC BYFull text7 citedMaterials & polymersOpenAlex
BibTeX
@article{wauchi2026pressure,
  title = {{Pressure behavior in a roller pump-driven low-flow VACC mock circuit with inactive de-airing pump: A descriptive bench characterization under reduced patient-side effective filling pressure conditions}},
  author = {Keizo Wauchi and Takahiro Okumura and Keisuke Matsuda and Hiroki Sato and Masahiro Miyata and Kenji Tani},
  journal = {PLOS One},
  year = {2026},
  month = nov,
  doi = {10.1371/journal.pone.0363658},
  url = {https://doi.org/10.1371/journal.pone.0363658},
  note = {CC BY 4.0},
}
RIS
TY  - JOUR
TI  - Pressure behavior in a roller pump-driven low-flow VACC mock circuit with inactive de-airing pump: A descriptive bench characterization under reduced patient-side effective filling pressure conditions
AU  - Keizo Wauchi
AU  - Takahiro Okumura
AU  - Keisuke Matsuda
AU  - Hiroki Sato
AU  - Masahiro Miyata
AU  - Kenji Tani
PY  - 2026
DA  - 2026/11/25
JO  - PLOS One
DO  - 10.1371/journal.pone.0363658
UR  - https://doi.org/10.1371/journal.pone.0363658
AB  - This exploratory bench study characterized pressure behavior in a roller pump-driven low-flow valve-type semi-closed extracorporeal circulation (VACC) mock circuit under experimentally reduced patient-side effective filling pressure conditions with the de-airing pump inactive. Pressures were measured at four sites: circuit-reservoir pressure (Pc), patient-side mock-reservoir outlet pressure (Pv), roller pump inlet pressure (Pi), and roller pump outlet pressure (Po). Pump flow-rate settings were 50–300 mL/min. Circuit-reservoir negative-pressure setpoints were 0, −10, −20, −30, and −40 mmHg, and patient-side mock-reservoir negative-pressure setpoints were 0, −5, and −10 mmHg. A 32 wt% aqueous glycerol solution was used. Pc generally reflected the circuit-reservoir negative-pressure setpoint, with a mean difference of −0.18 mmHg (95% CI, −0.37 to 0.00). The 95% CIs summarize variability among technical replicates within this bench system and should not be interpreted as population-inferential estimates. Pv decreased stepwise according to the patient-side setpoint. Within the common evaluation range, mean differences versus 0 mmHg were −5.7 mmHg (95% CI, −6.0 to −5.3) for the −5 mmHg condition and −9.2 mmHg (95% CI, −9.5 to −8.9) for the −10 mmHg condition. Pi shifted in the negative direction as the pump flow-rate setting increased, whereas Po increased monotonically under the fixed occluder condition. Delivered flow was independently verified volumetrically before the main experiments at 100, 200, and 300 mL/min but was not continuously measured during the pressure experiments. Therefore, Po was interpreted primarily in relation to the pump flow-rate setting rather than as a directly measured hydraulic pressure–flow relationship. These data provide baseline pressure-characterization information for future mock-circuit studies and do not establish clinical safety, efficacy, or a stable operating range.