Hemolysis in neonatal whole-blood gas testing: prevalence and association with potassium.
- 1Department of Neonatology, Hospital General Universitario Gregorio Marañón, Madrid, Spain. [email protected].
- 2Department of Neonatology, Hospital General Universitario Gregorio Marañón, Madrid, Spain.
Abstract
Hemolysis is a frequent preanalytical problem in neonatal blood sampling and may lead to falsely elevated whole-blood potassium measurements. Integrated hemolysis detection may identify affected samples at the point of care. We retrospectively analyzed neonatal whole-blood samples from 18 May to 9 September 2026 using the GEM Premier 7000 with integrated hemolysis index (HI) measurement. Significant hemolysis was defined as HI ≥ 2 (> 50 mg/dL plasma-free hemoglobin) and elevated potassium as K⁺ > 5.2 mmol/L. We assessed hemolysis prevalence, collection container, sample source, and associations with potassium and pH. Repeated measurements were accounted for using patient-clustered generalized estimating equations. Among 2520 samples from 334 patients, 257 (10.2%) were hemolysis-positive. Hemolysis was more frequent in capillary- than syringe-collected specimens (12.6 vs 6.1%; p < 0.001) and differed by source (capillary 14.0%, venous 7.1%, arterial 2.0%; p < 0.001). Elevated potassium occurred in 44.0% of hemolyzed versus 14.2% of non-hemolyzed samples (RR 3.09, 95% CI 2.60-3.67; p < 0.001) and increased with higher HI (ρ = 0.23; p < 0.001). At HI ≥ 3, 63.1% had elevated potassium versus 16.0% below this threshold (RR 3.93, 95% CI 3.20-4.83). Hemolysis remained independently associated with elevated potassium (adjusted OR 4.62, 95% CI 3.39-6.29; p < 0.001). The pH-potassium association differed by hemolysis status (interaction p = 0.015).Conclusions
Hemolysis was strongly associated with elevated potassium, with a graded relationship across HI categories. Hemolysis also varied by collection characteristics and sample source. Integrating HI with potassium, pH, and sample source may improve neonatal potassium interpretation and preanalytical quality monitoring.What is known
• Hemolysis is a common preanalytical problem in neonatal blood sampling and can cause falsely elevated potassium concentrations. • Integrated hemolysis detection in blood-gas analyzers may improve recognition of hemolysis-related potassium interference.What is new
• In neonatal whole-blood samples, hemolysis was independently associated with elevated potassium and showed a graded relationship with increasing hemolysis severity. • Hemolysis prevalence differed according to collection characteristics and sample source, with the highest prevalence in capillary specimens; the relationship between pH and potassium also differed according to hemolysis status.
Related
- The prevalence of neonatal sepsis caused by gram-negative bacteria and their antibiotic resistance patterns in Pakistan: a systematic review and meta-analysis
- A prospective study on the prevalence of neonatal alloimmune thrombocytopenia with maternal and neonatal human platelet antigen genotypic profiling.
- Prevalence of Hepatitis C Infection and Treatment Initiation Following Dried Blood Spot Testing in New South Wales, Australia
- Unexpected neonatal ABO discrepancy revealing ABO*cisAB.05 during routine cord blood testing
- Association of sodium intake with incident asthma and the modifying role of potassium: insights from a prospective study.
BibTeX
@article{merinohernandez2026hemolysis,
title = {{Hemolysis in neonatal whole-blood gas testing: prevalence and association with potassium.}},
author = {Amaia Merino-Hernández and Carmen Gómez-Gérez and Helena Caraça-Valente Barrera and Cristina Ramos-Navarro and Noelia González-Pacheco and Lorena Suárez-Fernández and Manuel Sánchez-Luna},
journal = {European journal of pediatrics},
year = {2026},
month = nov,
doi = {10.1007/s00431-026-07658-z},
url = {https://europepmc.org/article/MED/43077136},
note = {CC BY-NC-ND},
}
RIS
TY - JOUR
TI - Hemolysis in neonatal whole-blood gas testing: prevalence and association with potassium.
AU - Amaia Merino-Hernández
AU - Carmen Gómez-Gérez
AU - Helena Caraça-Valente Barrera
AU - Cristina Ramos-Navarro
AU - Noelia González-Pacheco
AU - Lorena Suárez-Fernández
AU - Manuel Sánchez-Luna
PY - 2026
DA - 2026/11/24
JO - European journal of pediatrics
DO - 10.1007/s00431-026-07658-z
UR - https://europepmc.org/article/MED/43077136
AB - Hemolysis is a frequent preanalytical problem in neonatal blood sampling and may lead to falsely elevated whole-blood potassium measurements. Integrated hemolysis detection may identify affected samples at the point of care. We retrospectively analyzed neonatal whole-blood samples from 18 May to 9 September 2026 using the GEM Premier 7000 with integrated hemolysis index (HI) measurement. Significant hemolysis was defined as HI ≥ 2 (> 50 mg/dL plasma-free hemoglobin) and elevated potassium as K⁺ > 5.2 mmol/L. We assessed hemolysis prevalence, collection container, sample source, and associations with potassium and pH. Repeated measurements were accounted for using patient-clustered generalized estimating equations. Among 2520 samples from 334 patients, 257 (10.2%) were hemolysis-positive. Hemolysis was more frequent in capillary- than syringe-collected specimens (12.6 vs 6.1%; p < 0.001) and differed by source (capillary 14.0%, venous 7.1%, arterial 2.0%; p < 0.001). Elevated potassium occurred in 44.0% of hemolyzed versus 14.2% of non-hemolyzed samples (RR 3.09, 95% CI 2.60-3.67; p < 0.001) and increased with higher HI (ρ = 0.23; p < 0.001). At HI ≥ 3, 63.1% had elevated potassium versus 16.0% below this threshold (RR 3.93, 95% CI 3.20-4.83). Hemolysis remained independently associated with elevated potassium (adjusted OR 4.62, 95% CI 3.39-6.29; p < 0.001). The pH-potassium association differed by hemolysis status (interaction p = 0.015).Conclusions Hemolysis was strongly associated with elevated potassium, with a graded relationship across HI categories. Hemolysis also varied by collection characteristics and sample source. Integrating HI with potassium, pH, and sample source may improve neonatal potassium interpretation and preanalytical quality monitoring.What is known • Hemolysis is a common preanalytical problem in neonatal blood sampling and can cause falsely elevated potassium concentrations. • Integrated hemolysis detection in blood-gas analyzers may improve recognition of hemolysis-related potassium interference.What is new • In neonatal whole-blood samples, hemolysis was independently associated with elevated potassium and showed a graded relationship with increasing hemolysis severity. • Hemolysis prevalence differed according to collection characteristics and sample source, with the highest prevalence in capillary specimens; the relationship between pH and potassium also differed according to hemolysis status.
SN - 0340-6199
SN - 1432-1076
ER -