Development and Temporal Validation of a Nomogram to Predict Actionable Arrhythmias During Implantable Cardiac Monitoring in Patients with Unexplained Syncope.
4 affiliations
- 1Department of Cardiology, Beijing Hospital, National Center for Gerontology, National Clinical Research Center for Gerontology, The Key Laboratory of Geriatrics of NHC, Institute of Geriatric Medicine, Chinese Academy of Medical Sciences & Peking Union Medical College, Beijing 100730, China.
- 2Department of Cardiology, Beijing Anzhen Hospital, Capital Medical University, Beijing 100029, China.
- 3Department of Cardiology, Cardiovascular Center, Beijing Friendship Hospital, Capital Medical University, Beijing 100050, China.
- 4Department of Cardiology, Fuwai Shenzhen Hospital, Chinese Academy of Medical Sciences, Shenzhen 518057, China.
Abstract
Prediction tools for actionable arrhythmias after implantable cardiac monitor (ICM) implantation in patients with unexplained syncope remain limited. This dual-center retrospective study included 302 patients divided chronologically into development (n = 91) and temporal validation (n = 211) cohorts. The endpoint was the first ICM-detected arrhythmia prompting permanent pacemaker or defibrillator implantation, catheter ablation, initiation of Vaughan Williams Class I or III antiarrhythmic drugs, or oral anticoagulation. A multivariable Cox model was developed and internally validated using 1000 bootstrap resamples, repeating variable selection and model fitting. During a median follow-up of 23.4 months, 78 patients reached the endpoint, including 32 in the development cohort and 46 in the temporal validation cohort. The nomogram included age, a history of recurrent syncope, and first-degree atrioventricular block. The optimism-corrected C-index was 0.720 in development. In temporal validation, the C-index was 0.707 (95% confidence interval, 0.623-0.790), and the 12- and 24-month time-dependent areas under the receiver operating characteristic curve were 0.733 and 0.741, respectively. The nomogram showed moderate discrimination in temporal validation at the same two centers. The model underestimated risk at 12 months, whereas mean predicted risk was closer to observed risk at 24 months.
Related
- Unexplained Syncope in Patients With Congenital Long QT Syndrome ― Role of Implantable Loop Recorders in Diagnostic Evaluation ―
- Predictive Value of the Electrical Risk Score for Ventricular Arrhythmias in Heart Failure Patients with Primary Prevention Implantable Cardioverter Defibrillators
- EP10.02 Identification & pre-clinical validation of actionable immunotherapy targets in pediatric CNS tumors
- Machine-learning prediction of high implantable collamer lens vault using angle-to-angle anatomy: a patient-grouped internal validation study
- Multi-angle validation of the iFalcon V700 rebound tonometer against implantable pressure sensor in ex-vivo rabbit eyes
BibTeX
@article{shen2026development,
title = {{Development and Temporal Validation of a Nomogram to Predict Actionable Arrhythmias During Implantable Cardiac Monitoring in Patients with Unexplained Syncope.}},
author = {Shuhui Shen and Jia Wang and Jiaxin Yang and Haiyan Yang and Donghao Liu and Yuxin Ma and Zeyu Niu and Lifeng Liang and Rui Wang and Zhijun Sun and Tong Zou},
journal = {Journal of cardiovascular development and disease},
year = {2026},
month = nov,
doi = {10.3390/jcdd13110595},
url = {https://europepmc.org/article/MED/43083102},
note = {CC BY},
}
RIS
TY - JOUR
TI - Development and Temporal Validation of a Nomogram to Predict Actionable Arrhythmias During Implantable Cardiac Monitoring in Patients with Unexplained Syncope.
AU - Shuhui Shen
AU - Jia Wang
AU - Jiaxin Yang
AU - Haiyan Yang
AU - Donghao Liu
AU - Yuxin Ma
AU - Zeyu Niu
AU - Lifeng Liang
AU - Rui Wang
AU - Zhijun Sun
AU - Tong Zou
PY - 2026
DA - 2026/11/14
JO - Journal of cardiovascular development and disease
DO - 10.3390/jcdd13110595
UR - https://europepmc.org/article/MED/43083102
AB - Prediction tools for actionable arrhythmias after implantable cardiac monitor (ICM) implantation in patients with unexplained syncope remain limited. This dual-center retrospective study included 302 patients divided chronologically into development (n = 91) and temporal validation (n = 211) cohorts. The endpoint was the first ICM-detected arrhythmia prompting permanent pacemaker or defibrillator implantation, catheter ablation, initiation of Vaughan Williams Class I or III antiarrhythmic drugs, or oral anticoagulation. A multivariable Cox model was developed and internally validated using 1000 bootstrap resamples, repeating variable selection and model fitting. During a median follow-up of 23.4 months, 78 patients reached the endpoint, including 32 in the development cohort and 46 in the temporal validation cohort. The nomogram included age, a history of recurrent syncope, and first-degree atrioventricular block. The optimism-corrected C-index was 0.720 in development. In temporal validation, the C-index was 0.707 (95% confidence interval, 0.623-0.790), and the 12- and 24-month time-dependent areas under the receiver operating characteristic curve were 0.733 and 0.741, respectively. The nomogram showed moderate discrimination in temporal validation at the same two centers. The model underestimated risk at 12 months, whereas mean predicted risk was closer to observed risk at 24 months.
SN - 2308-3425
ER -