Quantitative Imaging of Scandium Isotopes with a Long Axial Field-of-View PET/CT

Abstract

Objective. 43Sc and 44Sc are positron-emitting isotopes of scandium that have attracted growing interest for theranostic applications in conjunction with the therapeutic isotope 47Sc. Both isotopes are non-pure positron emitters, and the associated prompt-gamma emission together with the high positron energy introduce complexities in achieving precise quantitative imaging. Approach. We evaluated the quantitative imaging performance of 44Sc and a 43/44Sc mixture on a long axial field-of-view (LAFOV) PET/CT using a NEMA IEC body phantom. We assessed contrast recovery coefficients, image noise and bias for different numbers of iterations and for absolute versus relative scatter correction, and implemented an isotope-specific point-spread-function (PSF) correction based on FLUKA-simulated positron ranges. Main results. The quantitative imaging performance of both isotopes is comparable to 18F, and the vendor's prompt-gamma correction effectively suppresses the additional random coincidences and lung error caused by prompt-gamma emission. The isotope-specific PSF correction further improves contrast recovery. While the bias obtained for 44Sc meets the EARL acceptance criterion for non-standard radionuclides, the bias for the 43/44Sc mixture exceeds it, which we attribute to the mixture-specific cross-calibration of the dose calibrators rather than to the PET imaging. Significance. Our results support the use of LAFOV PET/CT for quantitative imaging of scandium isotopes and provide practical reconstruction recommendations for clinical protocols.

Related

  1. Axial memory effect theory for imaging through thin scattering media
    Shaurya Aarav, Jason Fleischer
    Optics ExpressPeer-reviewedLicence not statedOptics & photonicsCrossref
  2. Low-latency axial depth estimation of 2.82 ms using light field microscopy
    Jiaqi li, Yu Cao, Hiromasa Oku
    Optics ExpressPeer-reviewedLicence not statedOptics & photonicsCrossref
  3. Enhancement of axial magnetic field generation during relativistic self-channeling of laser radiation propagating along thin films
    Vitalia A. Kuleshova, Artem V. Korzhimanov
    arXivPreprintCC BY 4.0Full textPlasma physicsarXiv
  4. Can 225Ac be imaged on a long axial field-of-view PET/CT with standard clinical settings?
    Lorenzo Mercolli
    arXivPreprintCC BY 4.0Full textBiological & medical physicsarXiv
  5. Quantitative phase nano-imaging with a laboratory source
    Luca Fardin, C Armstrong, Alberto Astolfo +8
    Optics ExpressPeer-reviewedLicence not statedApplied physics & instrumentationCrossref
BibTeX
@misc{mercolli2026quantitative,
  title = {{Quantitative Imaging of Scandium Isotopes with a Long Axial Field-of-View PET/CT}},
  author = {Lorenzo Mercolli and Mohammadreza Teimoorisichani and Alenka E. Gaschen and Pascal V. Grundler and Anzhelika N. Moiseeva and Saverio Braccini and Nicholas P. van der Meulen},
  howpublished = {arXiv},
  year = {2026},
  month = oct,
  doi = {10.48550/arxiv.2610.28884},
  eprint = {2610.28884},
  archivePrefix = {arXiv},
  url = {https://arxiv.org/abs/2610.28884},
  note = {CC BY 4.0},
}
RIS
TY  - UNPB
TI  - Quantitative Imaging of Scandium Isotopes with a Long Axial Field-of-View PET/CT
AU  - Lorenzo Mercolli
AU  - Mohammadreza Teimoorisichani
AU  - Alenka E. Gaschen
AU  - Pascal V. Grundler
AU  - Anzhelika N. Moiseeva
AU  - Saverio Braccini
AU  - Nicholas P. van der Meulen
PY  - 2026
DA  - 2026/10/20
PB  - arXiv
DO  - 10.48550/arxiv.2610.28884
UR  - https://arxiv.org/abs/2610.28884
AB  - Objective. ^43Sc and ^44Sc are positron-emitting isotopes of scandium that have attracted growing interest for theranostic applications in conjunction with the therapeutic isotope ^47Sc. Both isotopes are non-pure positron emitters, and the associated prompt-gamma emission together with the high positron energy introduce complexities in achieving precise quantitative imaging. Approach. We evaluated the quantitative imaging performance of ^44Sc and a ^43/44Sc mixture on a long axial field-of-view (LAFOV) PET/CT using a NEMA IEC body phantom. We assessed contrast recovery coefficients, image noise and bias for different numbers of iterations and for absolute versus relative scatter correction, and implemented an isotope-specific point-spread-function (PSF) correction based on FLUKA-simulated positron ranges. Main results. The quantitative imaging performance of both isotopes is comparable to ^18F, and the vendor's prompt-gamma correction effectively suppresses the additional random coincidences and lung error caused by prompt-gamma emission. The isotope-specific PSF correction further improves contrast recovery. While the bias obtained for ^44Sc meets the EARL acceptance criterion for non-standard radionuclides, the bias for the ^43/44Sc mixture exceeds it, which we attribute to the mixture-specific cross-calibration of the dose calibrators rather than to the PET imaging. Significance. Our results support the use of LAFOV PET/CT for quantitative imaging of scandium isotopes and provide practical reconstruction recommendations for clinical protocols.
ER  -