Reducing the Risk of Gradient Induced EEG Hardware Heating During Simultaneous EEG-fMRI

  1. 1Sir Peter Mansfield Imaging Centre, School of Physics and Astronomy, University of Nottingham

Abstract

Simultaneous electroencephalography (EEG) and functional MRI (fMRI) has the potential to provide a more complete characterisation of brain function. Acquiring these data simultaneously requires the EEG hardware to be MR safe, since the EEG cap, amplifiers and powerpacks must be located within the MR environment. Placing them in this environment exposes them to both static and varying (radiofrequency (RF) and gradient) magnetic fields, presenting a potential heating risk. Previous studies have typically positioned the amplifiers and powerpacks outside of the bore, meaning that this risk is largely limited to the EEG cap. However, current guidelines recommend positioning the amplifiers within the bore. This reduces the distance between the EEG electrodes and amplifier, improving EEG signal quality and reducing risks of RF heating at the electrodes, but presents a potential heating risk for the amplifiers and powerpacks from the varying magnetic field gradients. In this study, we demonstrate that placing the amplifiers and powerpacks in the bore presents a significant heating risk, both at 7T and 3T. At 7T, we show that EPI sequences with a slew rate of at least 88 T/m/s resulted in significant heating of the EEG amplifiers. This heating effect could be negated by reducing the slew rate to 61 T/m/s. At 3T, we replicate the EEG amplifier heating and show that an alternative approach to reducing the gradient slew rates experience by the hardware is to increase the distance between the EEG amplifier/powerpacks and the isocentre of the magnet. Increasing the distance between the EEG hardware and MR isocentre by 30 cm mitigates the heating risk to the amplifiers and powerpacks, regardless of sequence parameters (e.g., slew rate). Therefore, whilst reducing the distance between the electrodes and amplifiers may improve EEG signal quality and reduce the potential for electrode RF heating, we show that risk of heating of hardware by gradient switching is greatly increased. This is particularly important to consider with the faster gradient slew rates available on modern MRI scanners, desired to increase spatial resolution. Experimenters must make careful decisions about hardware placement to ensure that the EEG-fMRI setup they use is safe, taking account of the varying risks that are posed by different approaches.

Related

  1. Sleep-related Brain Activity in an Arousal-reducing Agent–induced Mouse Model Evaluated with fMRI
    Hanabi Tamada, Junichi Hata, Kanako Muta +5
    Magnetic Resonance in Medical SciencesPeer-reviewedCC BYFull textNeuroscienceJ-STAGE
  2. Light-induced carotenogenesis enhances growth and high hydrostatic pressure tolerance in a deep-sea bacterium by reducing intracellular ROS
    Jin Lin, Xue-Gong Li, Wei-Jia Zhang +3
    Applied and Environmental MicrobiologyPeer-reviewedCC BYFull textEcologyHAL
  3. Ultrafast firing effects in Y-doped zirconia nanoparticles: Heating rate matters even under “Conventional” conditions
    Mattia Biesuz, Etienne Martin, Levent Karacasulu +4
    Journal of the European Ceramic SocietyPeer-reviewedCC BY 4.0Full textGeneral biologyGeneral chemistryHAL
  4. QuImputer: an LD-informed QUBO/Ising formulation of genotype imputation with quantum-hardware execution
    M. Kim, T.-H. Lee
    bioRxivPreprintAll rights reservedFull textBioinformatics & computational biologybioRxiv
  5. WinPopLDdecay: a hardware-adaptive and reproducible reimplementation of PopLDdecay with a graphical interface
    Boyang Zhao, Chenyang Zhang, Zhao-Kui Gong +5
    GigaSciencePeer-reviewedCC BY 4.0Full textBioinformatics & computational biologyCrossref
BibTeX
@misc{ingram2026reducing,
  title = {{Reducing the Risk of Gradient Induced EEG Hardware Heating During Simultaneous EEG-fMRI}},
  author = {B. T. Ingram and R. A. Waggoner and M. Uji and K. Ueno and C. Suzuki and F. Sato and X. Li and S. D. Mayhew and A. P. Bagshaw and M. Tamaki and K. J. Mullinger},
  howpublished = {bioRxiv},
  year = {2026},
  month = oct,
  doi = {10.64898/2026.10.30.761889},
  url = {https://www.biorxiv.org/content/10.64898/2026.10.30.761889v1},
  note = {CC BY},
}
RIS
TY  - UNPB
TI  - Reducing the Risk of Gradient Induced EEG Hardware Heating During Simultaneous EEG-fMRI
AU  - B. T. Ingram
AU  - R. A. Waggoner
AU  - M. Uji
AU  - K. Ueno
AU  - C. Suzuki
AU  - F. Sato
AU  - X. Li
AU  - S. D. Mayhew
AU  - A. P. Bagshaw
AU  - M. Tamaki
AU  - K. J. Mullinger
PY  - 2026
DA  - 2026/10/30
PB  - bioRxiv
DO  - 10.64898/2026.10.30.761889
UR  - https://www.biorxiv.org/content/10.64898/2026.10.30.761889v1
AB  - Simultaneous electroencephalography (EEG) and functional MRI (fMRI) has the potential to provide a more complete characterisation of brain function. Acquiring these data simultaneously requires the EEG hardware to be MR safe, since the EEG cap, amplifiers and powerpacks must be located within the MR environment. Placing them in this environment exposes them to both static and varying (radiofrequency (RF) and gradient) magnetic fields, presenting a potential heating risk. Previous studies have typically positioned the amplifiers and powerpacks outside of the bore, meaning that this risk is largely limited to the EEG cap. However, current guidelines recommend positioning the amplifiers within the bore. This reduces the distance between the EEG electrodes and amplifier, improving EEG signal quality and reducing risks of RF heating at the electrodes, but presents a potential heating risk for the amplifiers and powerpacks from the varying magnetic field gradients. In this study, we demonstrate that placing the amplifiers and powerpacks in the bore presents a significant heating risk, both at 7T and 3T. At 7T, we show that EPI sequences with a slew rate of at least 88 T/m/s resulted in significant heating of the EEG amplifiers. This heating effect could be negated by reducing the slew rate to 61 T/m/s. At 3T, we replicate the EEG amplifier heating and show that an alternative approach to reducing the gradient slew rates experience by the hardware is to increase the distance between the EEG amplifier/powerpacks and the isocentre of the magnet. Increasing the distance between the EEG hardware and MR isocentre by 30 cm mitigates the heating risk to the amplifiers and powerpacks, regardless of sequence parameters (e.g., slew rate). Therefore, whilst reducing the distance between the electrodes and amplifiers may improve EEG signal quality and reduce the potential for electrode RF heating, we show that risk of heating of hardware by gradient switching is greatly increased. This is particularly important to consider with the faster gradient slew rates available on modern MRI scanners, desired to increase spatial resolution. Experimenters must make careful decisions about hardware placement to ensure that the EEG-fMRI setup they use is safe, taking account of the varying risks that are posed by different approaches.
ER  -