Thermomechanical Comparison of PET-G and TC-85 Orthodontic Aligner Materials
- 1Department of Orthodontics, Faculty of Dentistry, Yozgat Bozok University, 66100 Yozgat, Türkiye
- 2Department of Orthodontics, Gülhane Faculty of Dental Medicine, University of Health Sciences, 06010 Ankara, Türkiye
- 3Department of Orthodontics, Faculty of Dentistry, Çanakkale Onsekiz Mart University, 17100 Canakkale, Türkiye
Abstract
This study combined thermomechanical characterisation with finite element analysis to compare polyethylene terephthalate glycol (PET-G) and Tera Harz TC-85 orthodontic aligner materials. Tensile properties were evaluated at 23, 37, 55 and 80 °C, with five observations per material and temperature. Identical full-arch geometries with a nominal thickness of 0.75 mm were subjected to 600 N compression at 300 K, heating under maintained load to 352 K and cooling to 310 K. At 23 °C, tensile moduli were 1396.4 ± 44.9 MPa for PET-G and 675.4 ± 21.2 MPa for TC-85. Dynamic mechanical analysis (DMA) loss-tangent maxima occurred at sampled temperatures of 80 and 45 °C. Tensile and DMA moduli diverged at elevated temperatures. Initial maximum effective strains were 0.165 and 0.196, and medium-to-fine mesh changes were 1.78% and 2.61%. TC-85 showed a greater reduction in maximum strain following heating to 78.85 °C. Recovery benchmark values at 60 min were 63.3% for PET-G and 92.5% for TC-85; paired-series metrics quantified curve agreement. PET-G exhibited higher tensile stiffness, whereas TC-85 exhibited greater benchmark recovery and a larger thermal-stage strain reduction. The high-temperature endpoint represents ex vivo activation. Constitutive and experimental limitations constrain extrapolation to clinical appliance performance.
Related
- Bibliometric Review of Bio-Based Building Materials and a Comparison with Prior Research: Toward Probabilistic Approaches for Material Variability
- Electrochemically driven thermomechanical energy harvesting
- Automated in situ thermomechanical analysis of cobalt under EBSD
- Thermomechanical Properties of Carbon and Glass Fiber-Reinforced Epoxy Composites as Function of Drilling Parameters
- Per-Layer UV Exposure-Dependent Cytotoxicity of 3D-Printed Aligner Resins and Estrogen-like Activity Screening
BibTeX
@article{askın2026thermomechanical,
title = {{Thermomechanical Comparison of PET-G and TC-85 Orthodontic Aligner Materials}},
author = {Melis Büşra Aşkın and Serkan Görgülü and Gökhan Serhat Duran},
journal = {Polymers},
year = {2026},
month = oct,
doi = {10.3390/polym18202517},
url = {https://www.mdpi.com/2073-4360/18/20/2517},
note = {CC BY 4.0},
}
RIS
TY - JOUR
TI - Thermomechanical Comparison of PET-G and TC-85 Orthodontic Aligner Materials
AU - Melis Büşra Aşkın
AU - Serkan Görgülü
AU - Gökhan Serhat Duran
PY - 2026
DA - 2026/10/01
JO - Polymers
DO - 10.3390/polym18202517
UR - https://www.mdpi.com/2073-4360/18/20/2517
AB - This study combined thermomechanical characterisation with finite element analysis to compare polyethylene terephthalate glycol (PET-G) and Tera Harz TC-85 orthodontic aligner materials. Tensile properties were evaluated at 23, 37, 55 and 80 °C, with five observations per material and temperature. Identical full-arch geometries with a nominal thickness of 0.75 mm were subjected to 600 N compression at 300 K, heating under maintained load to 352 K and cooling to 310 K. At 23 °C, tensile moduli were 1396.4 ± 44.9 MPa for PET-G and 675.4 ± 21.2 MPa for TC-85. Dynamic mechanical analysis (DMA) loss-tangent maxima occurred at sampled temperatures of 80 and 45 °C. Tensile and DMA moduli diverged at elevated temperatures. Initial maximum effective strains were 0.165 and 0.196, and medium-to-fine mesh changes were 1.78% and 2.61%. TC-85 showed a greater reduction in maximum strain following heating to 78.85 °C. Recovery benchmark values at 60 min were 63.3% for PET-G and 92.5% for TC-85; paired-series metrics quantified curve agreement. PET-G exhibited higher tensile stiffness, whereas TC-85 exhibited greater benchmark recovery and a larger thermal-stage strain reduction. The high-temperature endpoint represents ex vivo activation. Constitutive and experimental limitations constrain extrapolation to clinical appliance performance.
SN - 2073-4360
ER -