Case Report: Portable hypothermic machine perfusion for interregional deceased-donor kidney sharing in Vietnam
4 affiliations
- 1Department of Pediatric Surgery and Abdominal Emergency, Hue Central Hospital, Hue, Vietnam
- 2Department of Urology B, Hue Central Hospital, Hue, Vietnam
- 3Department of Thoracic and Cardiovascular Surgery, Hue Central Hospital, Hue, Vietnam
- 4Department of Urology A, Hue Central Hospital, Hue, Vietnam
Abstract
Long-distance transportation can prolong cold ischemia and complicate the utilization of deceased-donor kidneys, particularly in geographically dispersed transplant systems. We report, to our knowledge, the first clinical implementation in Vietnam of portable hypothermic machine perfusion (HMP) during interregional kidney transport. A kidney from a 61-year-old expanded-criteria deceased donor was allocated to a 42-year-old man with end-stage renal disease who had been receiving hemodialysis for 2 years following failure of a previous kidney allograft. The graft was transported over 1,000 km by ground transfer and scheduled commercial flight and was preserved using a LifePort Kidney Transporter with KPS-1 solution under pressure-controlled HMP at 30 mmHg. Total cold ischemia time was 10 h 50 min. From HMP initiation to arrival, renal vascular resistance decreased from 1.68 to 0.44 mmHg/mL/min, while perfusate flow increased from 16 to 52 mL/min. Despite three separate renal arteries and a second warm ischemia time of 55 min, the graft produced urine immediately after reperfusion. The recipient required no dialysis during the first postoperative week and maintained stable graft function at 6 months, with a serum creatinine of 142 µmol/L and an estimated glomerular filtration rate of approximately 55 mL/min/1.73 m2. This case demonstrates the operational feasibility of portable HMP for long-distance, interregional deceased-donor kidney sharing in Vietnam.
Related
- Factors Associated with Delayed Graft Function in a Universal Hypothermic Machine Perfusion Program: A Single-Center Retrospective Cohort Study
- First Split Liver Transplantation Using VitaSmart Hypothermic Machine Perfusion in the United States.
- Texture Analysis Using Indocyanine Green Fluorescence Measurement During Porcine Hypothermic Oxygenated Machine Perfusion
- Comment on Sun et al. First Experience with Hypothermic Oxygenated Perfusion in Human Uteri: Feasibility and Metabolic Characterization. J. Clin. Med. 2026, 15, 2820
- Hypothermic Oxygenated Perfusion Versus Static Cold Storage in Liver Transplantation: A Meta‐ Analysis of Randomized Trials
BibTeX
@article{pham2026case,
title = {{Case Report: Portable hypothermic machine perfusion for interregional deceased-donor kidney sharing in Vietnam}},
author = {Hiep Nhu Pham and Xuan Thanh Nguyen and Khanh Ngoc Tran and Dung Duc Nguyen and Anh Van Quoc Nguyen and Kha Cong Le Bui and Hung Ngoc Pham},
journal = {Frontiers in Transplantation},
year = {2026},
month = oct,
doi = {10.3389/frtra.2026.1982759},
url = {https://doi.org/10.3389/frtra.2026.1982759},
note = {CC BY 4.0},
}
RIS
TY - JOUR
TI - Case Report: Portable hypothermic machine perfusion for interregional deceased-donor kidney sharing in Vietnam
AU - Hiep Nhu Pham
AU - Xuan Thanh Nguyen
AU - Khanh Ngoc Tran
AU - Dung Duc Nguyen
AU - Anh Van Quoc Nguyen
AU - Kha Cong Le Bui
AU - Hung Ngoc Pham
PY - 2026
DA - 2026/10/21
JO - Frontiers in Transplantation
DO - 10.3389/frtra.2026.1982759
UR - https://doi.org/10.3389/frtra.2026.1982759
AB - Long-distance transportation can prolong cold ischemia and complicate the utilization of deceased-donor kidneys, particularly in geographically dispersed transplant systems. We report, to our knowledge, the first clinical implementation in Vietnam of portable hypothermic machine perfusion (HMP) during interregional kidney transport. A kidney from a 61-year-old expanded-criteria deceased donor was allocated to a 42-year-old man with end-stage renal disease who had been receiving hemodialysis for 2 years following failure of a previous kidney allograft. The graft was transported over 1,000 km by ground transfer and scheduled commercial flight and was preserved using a LifePort Kidney Transporter with KPS-1 solution under pressure-controlled HMP at 30 mmHg. Total cold ischemia time was 10 h 50 min. From HMP initiation to arrival, renal vascular resistance decreased from 1.68 to 0.44 mmHg/mL/min, while perfusate flow increased from 16 to 52 mL/min. Despite three separate renal arteries and a second warm ischemia time of 55 min, the graft produced urine immediately after reperfusion. The recipient required no dialysis during the first postoperative week and maintained stable graft function at 6 months, with a serum creatinine of 142 µmol/L and an estimated glomerular filtration rate of approximately 55 mL/min/1.73 m2. This case demonstrates the operational feasibility of portable HMP for long-distance, interregional deceased-donor kidney sharing in Vietnam.
SN - 2813-2440
ER -