Mechanism-guided endograft explantation eliminates shear-mediated aortic injury in a controlled simulation study.
- 1Texas A&M School of Engineering Medicine, Houston, TX.
- 2Department of Cardiovascular Surgery, Houston Methodist Hospital, Houston, TX.
Abstract
Objective
Endograft explantation is associated with high morbidity and mortality, driven in part by uncontrolled shear-mediated aortic injury. In this study, we evaluated whether a mechanism-based explantation device could reduce this injury and improve the procedural performance.Methods
In a controlled simulation using a multilayered three-dimensional printed aortic model, 25 vascular trainees and surgeons performed endograft explantation using both a conventional syringe technique and a novel device (EndoEx). The outcomes included quantitative intimal injury, procedure time, and validated usability metrics.Results
The EndoEx device reduced intimal injury by approximately 90% compared with the syringe technique (0.94% vs 9.56%; P < .001), representing near-elimination of shear-mediated disruption. The procedure time was moderately longer (+32 seconds; P < .05). EndoEx demonstrated superior usability across all domains, including ease of use, safety perception, and overall cognitive load (P < .05).Conclusions
In this controlled simulation, a mechanism-based explantation strategy (EndoEx) markedly reduced shear-mediated aortic injury while improving operator control. These findings support further in vivo evaluation and establish a mechanistic foundation for a device-based approach to endovascular aneurysm repair explantation that warrants clinical validation.
Related
- Surgical explantation following early transcatheter aortic valve failure: Technical considerations for complex situations.
- Surgical aortic valve replacement-level aortotomy for self-expanding transcatheter aortic valve explantation
- Opioid Withdrawal Following Intrathecal Morphine Pump Explantation Managed with a Structured Oral Opioid Taper: A Case Study
- Case Report: A case of severe chronic infection of an intraperitoneally placed composite mesh treated with complete mesh explantation and primary tissue repair using the anterior component separation technique
- Functional pseudocoarctation after branched arch endograft implantation in a patient with previous type A aortic dissection repair
BibTeX
@article{moin2026mechanismguided,
title = {{Mechanism-guided endograft explantation eliminates shear-mediated aortic injury in a controlled simulation study.}},
author = {Zain Moin and Alayne Morrel and Maham Rahimi},
journal = {Journal of vascular surgery cases and innovative techniques},
year = {2026},
month = sep,
doi = {10.1016/j.jvscit.2026.102489},
url = {https://europepmc.org/article/MED/42904182},
note = {CC BY-NC-ND},
}
RIS
TY - JOUR
TI - Mechanism-guided endograft explantation eliminates shear-mediated aortic injury in a controlled simulation study.
AU - Zain Moin
AU - Alayne Morrel
AU - Maham Rahimi
PY - 2026
DA - 2026/09/03
JO - Journal of vascular surgery cases and innovative techniques
DO - 10.1016/j.jvscit.2026.102489
UR - https://europepmc.org/article/MED/42904182
AB - Objective Endograft explantation is associated with high morbidity and mortality, driven in part by uncontrolled shear-mediated aortic injury. In this study, we evaluated whether a mechanism-based explantation device could reduce this injury and improve the procedural performance.Methods In a controlled simulation using a multilayered three-dimensional printed aortic model, 25 vascular trainees and surgeons performed endograft explantation using both a conventional syringe technique and a novel device (EndoEx). The outcomes included quantitative intimal injury, procedure time, and validated usability metrics.Results The EndoEx device reduced intimal injury by approximately 90% compared with the syringe technique (0.94% vs 9.56%; P < .001), representing near-elimination of shear-mediated disruption. The procedure time was moderately longer (+32 seconds; P < .05). EndoEx demonstrated superior usability across all domains, including ease of use, safety perception, and overall cognitive load (P < .05).Conclusions In this controlled simulation, a mechanism-based explantation strategy (EndoEx) markedly reduced shear-mediated aortic injury while improving operator control. These findings support further in vivo evaluation and establish a mechanistic foundation for a device-based approach to endovascular aneurysm repair explantation that warrants clinical validation.
SN - 2468-4287
ER -