Evaluation of Multi-Needle Mesotherapy Needles: Structural Integrity, Solution Uniformity, and Nociceptive Response

  1. 1Bloomage Biotech Co., Ltd., Jinan, Shandong, People’s Republic of China

Abstract

Background: Mesotherapy using nine-needle systems has gained popularity in aesthetic medicine, yet key performance parameters—structural integrity, tip alignment, solution uniformity, and nociceptive—remain insufficiently characterized.

Methods: Needle integrity and tip alignment were assessed by scanning electron microscopy (SEM) and industrial computed tomography (CT) before and after 140 porcine-skin insertions. Inter-needle solution uniformity was quantified using a custom 3D-printed collection device. Nociceptive responses were evaluated by electromyographic (EMG) recording of the flexor reflex in three New Zealand White rabbits (n = 3). In each rabbit, the test sample was evaluated on the right foot and the control sample on the left foot. For both the test and control samples, either unused needles or needles that had undergone 140 consecutive punctures in porcine skin were used for the flexor-reflex test to simulate repeated clinical use and assess its effect on the nociceptive response.

Results: Both devices maintained structural continuity after 140 insertions, with no fracture. SEM revealed that control samples exhibited tip bending, blunting, edge irregularity, and increased surface roughness, whereas test samples retained a sharp, symmetric tip with minimal morphological change. The test samples showed significantly lower exit errors than the control group both before insertion (±0.0173–0.0240 vs ±0.0315–0.0432, P≤0.01) and after insertion (±0.0155–0.0242 vs ±0.0322–0.0431, P≤0.01). In comparison, the alignment of the test samples was more consistent (the detection error of the instrument was controlled within ±0.05 mm). Solution uniformity analysis showed a narrower deviation range for the test sample, indicating improved flow consistency. EMG showed that the test sample induced lower flexor reflex amplitudes than the control, both before (608–879 μV vs 462–1458 μV) and after 140 insertions (68–425 μV vs 234–855 μV), indicating reduced nociceptive stimulation.

Conclusion: While both nine-needle devices maintained structural integrity after repeated use, the test sample demonstrated favorable solution uniformity and lower nociceptive response. Optimizing needle design and manufacturing consistency may improve patient comfort and treatment reliability in mesotherapy.

Related

  1. Spatiotemporally-guided defunctionalization of nociceptive nerves with capsaicin enhances structural repair in a murine Achilles tendon rupture model
    Chengyu Zhuang, Rong Zhou, Yifan Qu +5
    Scientific ReportsPeer-reviewedCC BYFull textSurgery & anaesthesiaDOAJ
  2. Structural Integrity of Mechanical Heart Valve Rings During Valve-in-Valve Interventions
    Paulina Anna Karlsson, Christian Butter, Anja Haase-Fielitz +1
    Structural heart : the journal of the Heart TeamPeer-reviewedCC BYFull textCardiovascular medicineEurope PMC
  3. Volumetric Integrity of Frontotemporal and Perisylvian Regions for Predicting Behavioral and Language Impairment in Behavioral FTD: A Structural MRI Study
    A. H. Zare, I. Kiani, L. Zeinalzadeh +3
    medRxivPreprintAll rights reservedFull textNeurologymedRxiv
  4. Efficacy of Hair Mesotherapy With Biomimetic Peptides and Micronutrients in Patients With Androgenetic Alopecia: A Retrospective Observational Study.
    Özlem Köse, Murat Borlu
    Journal of cosmetic dermatologyPeer-reviewedCC BYFull textGeneral medicineEurope PMC
  5. Short-Term Clinical Outcomes of Mesotherapy vs. Saline Injection in Patients with Lateral Epicondylitis: A Retrospective Comparative Study
    Arzu Dinç Yavaş, Sümeyye Akçay
    Journal of Clinical MedicinePeer-reviewedCC BY 4.0Surgery & anaesthesiaCrossref
BibTeX
@article{guan2026evaluation,
  title = {{Evaluation of Multi-Needle Mesotherapy Needles: Structural Integrity, Solution Uniformity, and Nociceptive Response}},
  author = {Qinyu Guan and Xin Li and Qinshuang Ou and Ruiyan Wang and Yang Su and Zeyu Fu and Xiaoling Qi},
  journal = {Medical Devices (Auckland, N.Z.)},
  year = {2026},
  month = nov,
  doi = {10.2147/mder.s641180},
  url = {https://doi.org/10.2147/mder.s641180},
  note = {CC BY-NC 4.0},
}
RIS
TY  - JOUR
TI  - Evaluation of Multi-Needle Mesotherapy Needles: Structural Integrity, Solution Uniformity, and Nociceptive Response
AU  - Qinyu Guan
AU  - Xin Li
AU  - Qinshuang Ou
AU  - Ruiyan Wang
AU  - Yang Su
AU  - Zeyu Fu
AU  - Xiaoling Qi
PY  - 2026
DA  - 2026/11/11
JO  - Medical Devices (Auckland, N.Z.)
DO  - 10.2147/mder.s641180
UR  - https://doi.org/10.2147/mder.s641180
AB  - Background: Mesotherapy using nine-needle systems has gained popularity in aesthetic medicine, yet key performance parameters—structural integrity, tip alignment, solution uniformity, and nociceptive—remain insufficiently characterized.  Methods: Needle integrity and tip alignment were assessed by scanning electron microscopy (SEM) and industrial computed tomography (CT) before and after 140 porcine-skin insertions. Inter-needle solution uniformity was quantified using a custom 3D-printed collection device. Nociceptive responses were evaluated by electromyographic (EMG) recording of the flexor reflex in three New Zealand White rabbits (n = 3). In each rabbit, the test sample was evaluated on the right foot and the control sample on the left foot. For both the test and control samples, either unused needles or needles that had undergone 140 consecutive punctures in porcine skin were used for the flexor-reflex test to simulate repeated clinical use and assess its effect on the nociceptive response.  Results: Both devices maintained structural continuity after 140 insertions, with no fracture. SEM revealed that control samples exhibited tip bending, blunting, edge irregularity, and increased surface roughness, whereas test samples retained a sharp, symmetric tip with minimal morphological change. The test samples showed significantly lower exit errors than the control group both before insertion (±0.0173–0.0240 vs ±0.0315–0.0432, P≤0.01) and after insertion (±0.0155–0.0242 vs ±0.0322–0.0431, P≤0.01). In comparison, the alignment of the test samples was more consistent (the detection error of the instrument was controlled within ±0.05 mm). Solution uniformity analysis showed a narrower deviation range for the test sample, indicating improved flow consistency. EMG showed that the test sample induced lower flexor reflex amplitudes than the control, both before (608–879 μV vs 462–1458 μV) and after 140 insertions (68–425 μV vs 234–855 μV), indicating reduced nociceptive stimulation.  Conclusion: While both nine-needle devices maintained structural integrity after repeated use, the test sample demonstrated favorable solution uniformity and lower nociceptive response. Optimizing needle design and manufacturing consistency may improve patient comfort and treatment reliability in mesotherapy.
SN  - 1179-1470
ER  -