Comparison of Cumulative Dissipated Energy Between Two Phacoemulsification Systems Using a Modified Flip-and-Chop Technique With Multidirectional Jerking Movement From a Phacoemulsification Handpiece.

  1. 1Ophthalmology, John A. Burns School of Medicine, University of Hawaii, Honolulu, USA.
  2. 2Ophthalmology, Zhongli University Eye Center, Taoyuan, TWN.
  3. 3Medicine, John A. Burns School of Medicine, University of Hawaii, Honolulu, USA.

Abstract

Introduction The four-dimensional Unity phacoemulsification tip is designed to optimize tip motion and reduce time and energy consumption in cataract surgery. This study highlights a modified flip-and-chop technique with an added multidirectional jerking motion to the phacoemulsification handpiece intraocularly (referred to as the "five-dimensional flip-and-chop") and compares the cumulative dissipated energy (CDE) between the four-dimensional Unity system (Alcon Inc., Geneva, Switzerland) and the torsional Centurion system (Alcon Inc., Geneva, Switzerland) under this technique. Methods This is a retrospective, sequential, non-randomized comparison of 40 consecutive cases of Centurion to 40 consecutive cases of Unity performed by a single surgeon at a single ambulatory surgical suite. The initial cataract density between the groups was compared using optical coherence tomography (OCT) image quality (IQ) values from a single Maestro 2 device (Topcon, Tokyo, Japan). Results The two groups had comparable mean ages, sex distributions, and IQs. The mean CDE was significantly lower in the Unity group (5.39; SD=2.8) than in the Centurion group (8.23; SD=5.9), with p=0.0075. Conclusion Utilizing the modified "five-dimensional flip-and-chop" technique, the four-dimensional Unity system produced significantly lower CDE than the torsional Centurion system. The addition of this technique to the phacoemulsification handpiece may further improve the efficiency of phacoemulsification.

Related

  1. Swaasa an AI-Based Novel Technique to Assess Lung Function in Yoga Practitioners in Comparison to Non-Yoga Practitioners: A Cross-Sectional study
    S. Rao, R. Pal, A. Rawat +5
    medRxivPreprintCC BY-NC-NDFull textGeneral medicinemedRxiv
  2. Comparison of Minimally Invasive Colectomy Between Intra‐ Versus Extra‐Corporeal Anastomosis Technique (COMMIT Study): Short‐Term Analysis
    Takuya Takami, Takehito Yamamoto, Yoshiro Itatani +7
    Annals of Gastroenterological SurgeryPeer-reviewedCC BY-NC-ND 4.0Full textOncologyPMC
  3. Aspiration-line pressure stability in Venturi and peristaltic pump configurations during phacoemulsification: an in vitro study
    Tomoyuki Kunishige, Hisaharu Suzuki, Fumiki Okamoto
    Scientific ReportsPeer-reviewedCC BY-NC-ND 4.0Full textGeneral medicineCrossref
  4. Unilateral ‘cheetah-like’ endotheliopathy: A rare endothelial phenotype regressing after phacoemulsification
    Sylvain Poinard, Oliver Dorado-Cortez, Zhiguo He +4
    American journal of ophthalmology case reportsPeer-reviewedCC BYFull textGeneral medicineEurope PMC
  5. Balanced versus hybrid phacoemulsification tips in grade 3 nuclear cataracts: a prospective randomized clinical study
    Giacomo Visioli, Oscar Matteo Gagliardi, Chiara Ciccarè +7
    Scientific ReportsPeer-reviewedCC BY-NC-ND 4.0Full textGeneral medicineCrossref
BibTeX
@article{chen2026comparison,
  title = {{Comparison of Cumulative Dissipated Energy Between Two Phacoemulsification Systems Using a Modified Flip-and-Chop Technique With Multidirectional Jerking Movement From a Phacoemulsification Handpiece.}},
  author = {Ming Chen and Hung-Yuan Lin and Emily J Kang},
  journal = {Cureus},
  year = {2026},
  month = oct,
  doi = {10.7759/cureus.118645},
  url = {https://europepmc.org/article/MED/42948744},
  note = {CC BY},
}
RIS
TY  - JOUR
TI  - Comparison of Cumulative Dissipated Energy Between Two Phacoemulsification Systems Using a Modified Flip-and-Chop Technique With Multidirectional Jerking Movement From a Phacoemulsification Handpiece.
AU  - Ming Chen
AU  - Hung-Yuan Lin
AU  - Emily J Kang
PY  - 2026
DA  - 2026/10/26
JO  - Cureus
DO  - 10.7759/cureus.118645
UR  - https://europepmc.org/article/MED/42948744
AB  - Introduction The four-dimensional Unity phacoemulsification tip is designed to optimize tip motion and reduce time and energy consumption in cataract surgery. This study highlights a modified flip-and-chop technique with an added multidirectional jerking motion to the phacoemulsification handpiece intraocularly (referred to as the "five-dimensional flip-and-chop") and compares the cumulative dissipated energy (CDE) between the four-dimensional Unity system (Alcon Inc., Geneva, Switzerland) and the torsional Centurion system (Alcon Inc., Geneva, Switzerland) under this technique. Methods This is a retrospective, sequential, non-randomized comparison of 40 consecutive cases of Centurion to 40 consecutive cases of Unity performed by a single surgeon at a single ambulatory surgical suite. The initial cataract density between the groups was compared using optical coherence tomography (OCT) image quality (IQ) values from a single Maestro 2 device (Topcon, Tokyo, Japan). Results The two groups had comparable mean ages, sex distributions, and IQs. The mean CDE was significantly lower in the Unity group (5.39; SD=2.8) than in the Centurion group (8.23; SD=5.9), with p=0.0075. Conclusion Utilizing the modified "five-dimensional flip-and-chop" technique, the four-dimensional Unity system produced significantly lower CDE than the torsional Centurion system. The addition of this technique to the phacoemulsification handpiece may further improve the efficiency of phacoemulsification.
SN  - 2168-8184
ER  -