Pulse duration and pulse repetition frequency shape cavitation dynamics and volumetric liquefaction by histotripsy in a fibrous tissue-mimicking phantom
Abstract
Histotripsy is a focused ultrasound technique that mechanically liquefies tissue through cavitation bubble activity and has shown promise for noninvasive tumor ablation. However, treating fibrous, mechanically robust tissues remains challenging, and the influence of pulse parameters on cavitation dynamics and liquefaction efficiency in these tissues is not well understood. Here, we systematically investigated the combined effects of pulse duration and pulse repetition frequency (PRF) on cavitation behavior and lesion formation in a fibrous tissue-mimicking double-network hydrogel. Pulse duration and PRF were varied over four orders of magnitude while maintaining a constant 1% duty cycle. High-speed photographic imaging, shear wave elastography (SWE), B-mode ultrasound, and open-top light-sheet (OTLS) microscopy were used to relate cavitation behavior to phantom damage and liquefaction. The 1-cycle condition produced the largest bubble cloud area at longer exposure times, while the 1 and 10-cycle conditions exhibited higher frame-to-frame correlation and greater spatial persistence. In contrast, longer-pulse, low-PRF exposures (≥ 100 cycles) produced more localized cavitation and contiguous lesion formation. Among the tested parameters, 1,000 -cycle pulses at 10 Hz yielded the highest volumetric liquefaction rates. In contrast, short-pulse, high-PRF conditions generated substantial total damage but little to no fully liquefied volume. These findings show that pulse duration and PRF shape cavitation behavior and liquefaction in this fibrous tissue-mimicking model.
Related
- On-chip high-repetition-rate passively mode-locked frequency comb source based on a microring composite cavity
- Maturation of Auditory Stream Segregation in School-Age Children: Effects of Signal Repetition and Frequency Predictability
- Volumetric Organization in Electromagnetic Motion: Distance, Frequency, and Common-Center Geometry
- A Portable Optical Frequency Comb with Low Repetition-Rate Drift Based on "Optical Cubes"
- Relativistic Mass Transport and Geometric Coupling: High-Frequency 2¹⁵ Hz Quartz Reflection for Volumetric Light-Speed Displacement
BibTeX
@article{kumar2026pulse,
title = {{Pulse duration and pulse repetition frequency shape cavitation dynamics and volumetric liquefaction by histotripsy in a fibrous tissue-mimicking phantom}},
author = {Yashwanth Nanda Kumar and Yak-Nam Wang and Wayne Kreider and Qinghua Han and Jonathan T. C. Liu and Eli Vlaisavljevich and George R. Schade and Adam D. Maxwell},
journal = {Scientific Reports},
year = {2026},
month = oct,
doi = {10.1038/s41598-026-72148-z},
url = {https://www.nature.com/articles/s41598-026-72148-z},
note = {CC BY 4.0},
}
RIS
TY - JOUR
TI - Pulse duration and pulse repetition frequency shape cavitation dynamics and volumetric liquefaction by histotripsy in a fibrous tissue-mimicking phantom
AU - Yashwanth Nanda Kumar
AU - Yak-Nam Wang
AU - Wayne Kreider
AU - Qinghua Han
AU - Jonathan T. C. Liu
AU - Eli Vlaisavljevich
AU - George R. Schade
AU - Adam D. Maxwell
PY - 2026
DA - 2026/10/01
JO - Scientific Reports
DO - 10.1038/s41598-026-72148-z
UR - https://www.nature.com/articles/s41598-026-72148-z
AB - Histotripsy is a focused ultrasound technique that mechanically liquefies tissue through cavitation bubble activity and has shown promise for noninvasive tumor ablation. However, treating fibrous, mechanically robust tissues remains challenging, and the influence of pulse parameters on cavitation dynamics and liquefaction efficiency in these tissues is not well understood. Here, we systematically investigated the combined effects of pulse duration and pulse repetition frequency (PRF) on cavitation behavior and lesion formation in a fibrous tissue-mimicking double-network hydrogel. Pulse duration and PRF were varied over four orders of magnitude while maintaining a constant 1% duty cycle. High-speed photographic imaging, shear wave elastography (SWE), B-mode ultrasound, and open-top light-sheet (OTLS) microscopy were used to relate cavitation behavior to phantom damage and liquefaction. The 1-cycle condition produced the largest bubble cloud area at longer exposure times, while the 1 and 10-cycle conditions exhibited higher frame-to-frame correlation and greater spatial persistence. In contrast, longer-pulse, low-PRF exposures (≥ 100 cycles) produced more localized cavitation and contiguous lesion formation. Among the tested parameters, 1,000 -cycle pulses at 10 Hz yielded the highest volumetric liquefaction rates. In contrast, short-pulse, high-PRF conditions generated substantial total damage but little to no fully liquefied volume. These findings show that pulse duration and PRF shape cavitation behavior and liquefaction in this fibrous tissue-mimicking model.
SN - 2045-2322
ER -