Creation of a Durable Conduit with a Novel Self-Expandable Metal Stent for Stone Extraction in Endoscopic Retrograde Cholangiopancreatography: A Pilot Study
- 1Division of Gastroenterology, Department of Internal Medicine, Chonnam National University Hospital, Chonnam National University Medical School, Gwangju, Korea
Abstract
Background/Aims: Despite technical advancements, various conditions such as bleeding diathesis and altered anatomy challenge the creation of a durable conduit for biliary stone removal in endoscopic retrograde cholangiopancreatography (ERCP). A fully covered self-expandable metal stent (SEMS) has been used as a robust conduit for drainage in a wide spectrum of pancreatobiliary diseases. However, its primary use for stone removal has not been fully evaluated. This study aimed to evaluate the efficacy and safety of a novel fully covered SEMS that is 3 cm in length and has a wide distal flare as a durable conduit for stone extraction through the papilla.
Methods: Twenty patients with common bile duct stones were enrolled prospectively and underwent ERCP with the novel SEMS. Primary outcomes were technical and therapeutic success rates. Secondary outcomes were the device removal success rate and any adverse events.
Results: The technical and therapeutic success rates were 95% (19/20) and 90% (18/20), respectively. The novel SEMS was removed successfully in all patients (100%). Mild post-ERCP pancreatitis developed in 15% of the patients (3/20), and post-ERCP cholangitis occurred in 5% (1/20). There were no cases of moderate to severe pancreatitis, perforation, bleeding, or procedure-related mortality.
Conclusions: The novel SEMS offers a durable conduit for safe and effective biliary stone extraction without major adverse events. Further studies in larger cohorts are warranted to confirm these findings (CRiS identifier: KCT0006185).
Related
- Impact of Stone Size on the Technical Success of Endoscopic Retrograde Cholangiopancreatography for Common Bile Duct Stone Extraction in Older Adults: A Single-Center Retrospective Study at a Secondary Care Hospital in Mexico
- A novel four-layer sheet-assisted technique for patient repositioning during endoscopic retrograde cholangiopancreatography
- Successful gel-immersion cannulation of hepaticojejunostomy during enteroscopy-assisted endoscopic retrograde cholangiopancreatography
- Remimazolam-based spontaneous breathing sedation for endoscopic retrograde cholangiopancreatography: a propensity score-matched retrospective cohort study
- Breaking the compatibility barrier: cholangioscopy-guided laser lithotripsy during enteroscopy-assisted endoscopic retrograde cholangiopancreatography in Roux-en-Y anatomy.
BibTeX
@article{cho2026creation,
title = {{Creation of a Durable Conduit with a Novel Self-Expandable Metal Stent for Stone Extraction in Endoscopic Retrograde Cholangiopancreatography: A Pilot Study}},
author = {Eunae Cho and Chang Hwan Park},
journal = {Gut and Liver},
year = {2026},
month = nov,
doi = {10.5009/gnl250355},
url = {https://doi.org/10.5009/gnl250355},
note = {CC BY-NC 4.0},
}
RIS
TY - JOUR
TI - Creation of a Durable Conduit with a Novel Self-Expandable Metal Stent for Stone Extraction in Endoscopic Retrograde Cholangiopancreatography: A Pilot Study
AU - Eunae Cho
AU - Chang Hwan Park
PY - 2026
DA - 2026/11/15
JO - Gut and Liver
DO - 10.5009/gnl250355
UR - https://doi.org/10.5009/gnl250355
AB - Background/Aims: Despite technical advancements, various conditions such as bleeding diathesis and altered anatomy challenge the creation of a durable conduit for biliary stone removal in endoscopic retrograde cholangiopancreatography (ERCP). A fully covered self-expandable metal stent (SEMS) has been used as a robust conduit for drainage in a wide spectrum of pancreatobiliary diseases. However, its primary use for stone removal has not been fully evaluated. This study aimed to evaluate the efficacy and safety of a novel fully covered SEMS that is 3 cm in length and has a wide distal flare as a durable conduit for stone extraction through the papilla. Methods: Twenty patients with common bile duct stones were enrolled prospectively and underwent ERCP with the novel SEMS. Primary outcomes were technical and therapeutic success rates. Secondary outcomes were the device removal success rate and any adverse events. Results: The technical and therapeutic success rates were 95% (19/20) and 90% (18/20), respectively. The novel SEMS was removed successfully in all patients (100%). Mild post-ERCP pancreatitis developed in 15% of the patients (3/20), and post-ERCP cholangitis occurred in 5% (1/20). There were no cases of moderate to severe pancreatitis, perforation, bleeding, or procedure-related mortality. Conclusions: The novel SEMS offers a durable conduit for safe and effective biliary stone extraction without major adverse events. Further studies in larger cohorts are warranted to confirm these findings (CRiS identifier: KCT0006185).
SN - 1976-2283
SN - 2005-1212
ER -