Semi-Automated Measurement of Central Retinal Artery Blood Flow Using Doppler Ultrasound Coupled With Electrocardiography

  1. 1Department of Ophthalmology, Icahn School of Medicine at Mount Sinai, New York, NY, USA
  2. 2Maine College of Engineering and Computing, The University of Maine, Orono, ME, USA
  3. 3Department of Cardiology, Icahn School of Medicine at Mount Sinai, New York, NY, USA

Abstract

Purpose: To design, implement, and test a semi-automated method for measuring central retinal artery (CRA) blood flow using color Doppler imaging (CDI) waveforms synchronized with electrocardiography (ECG).

Methods: Twenty-four patients (48 eyes; 16/8 male/female; mean age 64.5 ± 15.8 years) underwent CDI-ECG imaging. Peak systolic velocity (PSV), end-diastolic velocity (EDV) and time from the ECG r-wave to the CRA PSV (r_PSV) were measured manually and with a semi-automated method.

Results: PSV showed excellent correlation and agreement between manual and semi-automated methods (right eye [OD] r = 0.95, intraclass correlation coefficient [ICC] = 0.945; left eye [OS] r = 0.94, ICC = 0.928). EDV had weaker results (OD r = 0.67, ICC = 0.441; OS r = 0.86, ICC = 0.585). The r_PSV had moderate correlations (OD r = 0.75; OS r = 0.42) and agreement (ICC = 0.702 and 0.387). Bland-Altman analysis showed small PSV bias (OD = 0.327; OS 0.419 cm/s), EDV overestimation (bias OD −0.556; OS −0.886 cm/s), and minimal r_PSV bias (OD = −0.012; OS = −0.010 s). Mixed-effects modeling found a significant method and eye effect only for EDV.

Conclusions: Semi-automated PSV measurements show excellent agreement with manual analysis and represent the most robust and clinically applicable output of the method. In contrast, EDV and r_PSV demonstrate lower agreement and should be considered exploratory, likely reflecting sensitivity to waveform morphology and low-magnitude signal characteristics.

Translational Relevance: A semi-automated CDI-ECG waveform analysis could allow for reproducible assessment of ocular blood flow and enable large-scale assessment of data examining the heart-eye relationship.

Related

  1. A comparative study of the state-of-the-art artificial intelligence models for automated plaque measurement in a carotid ultrasound Japanese cohort
    Nikhil Singh, Pankaj K. Jain, Rubeena Vohra +22
    Cardiovascular Diagnosis and TherapyPeer-reviewedCC BY-NC-ND 4.0Full textCardiovascular medicinePMC
  2. Automated Segmentation of Retinal Pigment Epithelium Layers in High-Resolution OCT
    Giulia Ritter, Ben Isselmann, Marlene Saßmannshausen +6
    Ophthalmology sciencePeer-reviewedCC BYFull textGeneral medicineEurope PMC
  3. Central retinal artery occlusion preceding hemispheric ischemic stroke treated with mechanical thrombectomy : A case report
    XiaoYan Wen, Baoyi Guo, Xue Zeng +3
    MedicinePeer-reviewedCC BYFull textGeneral medicineEurope PMC
  4. Effect of Intravitreal Anti-VEGF Injections on the Vitreomacular Interface in Treatment-Naïve Eyes With Central Retinal Vein Occlusion
    Rachel D’Cunha, Angela Hemesath, Yuxi Zheng +6
    Journal of Vitreoretinal DiseasesPeer-reviewedLicence not statedFull text1 citedGeneral medicinePMC
  5. A Semi-Automated Interactive Annotation Framework for Intraoperative Neurosurgical Ultrasound: A Proof-of-Concept Study
    R. Bertani, S. Bahadir, M. Y. Ferreira +3
    medRxivPreprintAll rights reservedFull textMedical imaging & radiologymedRxiv
BibTeX
@article{wood2026semiautomated,
  title = {{Semi-Automated Measurement of Central Retinal Artery Blood Flow Using Doppler Ultrasound Coupled With Electrocardiography}},
  author = {Keren Wood and Alon Harris and Temitope Olayinka and Alice Verticchio Vercellin and Yoav Niv-Granot and Stamatios Lerakis and Chen Gurevitz and Brent Siesky and Yash Lahoti and Giovanna Guidoboni},
  journal = {Translational Vision Science \& Technology},
  year = {2026},
  month = oct,
  doi = {10.1167/tvst.15.10.43},
  url = {https://doi.org/10.1167/tvst.15.10.43},
  note = {CC BY-NC-ND 4.0},
}
RIS
TY  - JOUR
TI  - Semi-Automated Measurement of Central Retinal Artery Blood Flow Using Doppler Ultrasound Coupled With Electrocardiography
AU  - Keren Wood
AU  - Alon Harris
AU  - Temitope Olayinka
AU  - Alice Verticchio Vercellin
AU  - Yoav Niv-Granot
AU  - Stamatios Lerakis
AU  - Chen Gurevitz
AU  - Brent Siesky
AU  - Yash Lahoti
AU  - Giovanna Guidoboni
PY  - 2026
DA  - 2026/10/27
JO  - Translational Vision Science & Technology
DO  - 10.1167/tvst.15.10.43
UR  - https://doi.org/10.1167/tvst.15.10.43
AB  - Purpose: To design, implement, and test a semi-automated method for measuring central retinal artery (CRA) blood flow using color Doppler imaging (CDI) waveforms synchronized with electrocardiography (ECG).  Methods: Twenty-four patients (48 eyes; 16/8 male/female; mean age 64.5 ± 15.8 years) underwent CDI-ECG imaging. Peak systolic velocity (PSV), end-diastolic velocity (EDV) and time from the ECG r-wave to the CRA PSV (r_PSV) were measured manually and with a semi-automated method.  Results: PSV showed excellent correlation and agreement between manual and semi-automated methods (right eye [OD] r = 0.95, intraclass correlation coefficient [ICC] = 0.945; left eye [OS] r = 0.94, ICC = 0.928). EDV had weaker results (OD r = 0.67, ICC = 0.441; OS r = 0.86, ICC = 0.585). The r_PSV had moderate correlations (OD r = 0.75; OS r = 0.42) and agreement (ICC = 0.702 and 0.387). Bland-Altman analysis showed small PSV bias (OD = 0.327; OS 0.419 cm/s), EDV overestimation (bias OD −0.556; OS −0.886 cm/s), and minimal r_PSV bias (OD = −0.012; OS = −0.010 s). Mixed-effects modeling found a significant method and eye effect only for EDV.  Conclusions: Semi-automated PSV measurements show excellent agreement with manual analysis and represent the most robust and clinically applicable output of the method. In contrast, EDV and r_PSV demonstrate lower agreement and should be considered exploratory, likely reflecting sensitivity to waveform morphology and low-magnitude signal characteristics.  Translational Relevance: A semi-automated CDI-ECG waveform analysis could allow for reproducible assessment of ocular blood flow and enable large-scale assessment of data examining the heart-eye relationship.
SN  - 2164-2591
ER  -