Skip to content
DraconiticResearch
Search Sources API
203,234 studies31 sourcesUpdated 28 min ago
Peer-reviewedCC BYJournal of applied clinical medical physics

Linac performance benchmarking: Feasibility of a web-based tool for comparison of local machine performance check (MPC) results with the global community.

1 more authors
15 affiliations
  1. 1Department of Radiation Oncology, Calvary Mater Hospital, Newcastle, New South Wales, Australia.
  2. 2School of Science, University of Newcastle, Newcastle, New South Wales, Australia.
  3. 3Centre for Medical and Radiation Physics, University of Wollongong, Wollongong, New South Wales, Australia.
  4. 4Medical Physics Department, Guys and St Thomas' Hospital, London, UK.
  5. 5School of Biomedical Engineering & Imaging Sciences, King's College London, London, UK.
  6. 6Central Coast Cancer Centre, Gosford Hospital, Gosford, New South Wales, Australia.
  7. 7Medical Physics and Radiation Engineering, Canberra Hospital, Garran, Australian Capital Territory, Australia.
  8. 8Department of Radiation Oncology, King Khalid University Medical City, Abha, Saudi Arabia.
  9. 9Central West Cancer Care Centre, Orange Health Service, Orange, New South Wales, Australia.
  10. 10North West Cancer Centre, Tamworth Rural Referral Hospital, Tamworth, New South Wales, Australia.
  11. 11Radiotherapy and Radiation Dosimetry Group, National Physical Laboratory, Teddington, UK.
  12. 12Department of Radiotherapy Physics, University College London Hospital, London, UK.
  13. 13Department of Medical Physics and Bioengineering, University College London, London, UK.
  14. 14National Radiotherapy Trials Quality Assurance Group, University College London Hospital, London, UK.
  15. 15Institute of Medical Physics, University of Sydney, Sydney, New South Wales, Australia.

Abstract

Background

Varian machine performance check (MPC) is a manufacturer integrated quality control system included as standard for TrueBeam and Halcyon linear accelerators (linacs) to assess performance using a broad range of geometric and beam tests. MPC is a highly standardized method of performance testing, is generally free of inter-user variability and is quick and easy to perform allowing daily acquisitions. Therefore, previous studies have highlighted the potential of comparative analysis of MPC data in investigating differences in performance of individual linacs and identifying outliers. Statistical process control (SPC) is an established approach commonly used to assess stability and behavior of sequential data and identify standard operating ranges which has previously been applied to radiotherapy quality assurance data and MPC data.Purpose

This study aimed to investigate the feasibility of a web-based platform using SPC and aggregated MPC data to allow users to compare their machine performance against that of a cohort of others in the radiotherapy community.Methods

A previously developed web-based platform for radiotherapy data sharing was expanded to include the MPC data collection and comparison tool. Pilot centers exported data from their MPC applications and uploaded it through center-specific portals. Within the application, SPC analysis was used to calculate upper control limits (UCL), lower control limits (LCL) and central lines (CL) for the data sets of each individual linac as well as cohort averages. Interactive plots of longitudinal data for individual machines were displayed for users and overlaid with SPC limits for the specific machine and the cohort averages. Additional analyses were conducted externally from the app where the average and range of SPC control limits for the cohort were compared against MPC tolerances.Results

MPC data for 57 648 measurements taken between April 2016 and May 2026 were collected across seven centers from 21 TrueBeam linacs (15 with Millennium multi-leaf collimator (MLC), six with HD120 MLCs) and one Halcyon. Variation in performance for specific tests was identified for individual machines compared against the cohort as well as for individual MLC leaves for specific machines. Across the cohort of TrueBeam linacs, average control ranges for all geometric tests were within MPC thresholds. However, the range of values for UCL and LCL exceeded tolerances for relative gantry angle. Machines installed before 2021 were found to have slightly larger variability in mean MLC offset with an average control range width of 0.20 mm and 0.18 mm for bank A and B, respectively, compared to machines installed after 2021 of 0.10 mm and 0.13 mm. Analysis of data for individual linacs segmented per year showed a change in mean MLC offset occurring around 2021.Conclusions

The usefulness and feasibility of a tool to compare MPC data against the radiotherapy community was demonstrated. Inter-center comparison of data allowed trends to be identified which would otherwise not be observable for individual centers. Further development and distribution of the MPC data sharing and comparison tool may provide improved confidence and understanding of machine performance and assist low-resourced clinics in delivering safe and accurate radiotherapy treatments.

Related

  1. Performance vs Portability in Heterogeneous HPC Environments: Why Pre-execution Benchmarking is Required
    Mindaugas Macernis
    arXivPreprintarXiv licenceFull textComputational physicsarXiv
  2. SI-Traceable Calibration and Performance Benchmarking of a Terahertz Photomixer Transmitter-Receiver System Using a Rydberg Atomic Sensor
    Alisher Duspayev, Kaitlin R. Moore, Georg Raithel +1
    arXivPreprintCC BY-NC-ND 4.0Full textAtomic, molecular & optical physicsApplied physics & instrumentationarXiv
  3. Comparison of spectral performance of three dual-energy CT scanners equipped with a deep-learning image reconstruction algorithm and one photon counting CT scanner: A phantom study
    Joël Greffier, Cécile Salvat, Maxime Pastor +5
    Diagnostic and Interventional ImagingPeer-reviewedCC BY-SA 4.0Full text11 citedGeneral physicsHAL
  4. Performance comparison of tethered and self-propelled models of fish locomotion using unsteady thin airfoil theory
    Anshul Nayak, Emad Masroor, Hodjat Pendar
    arXivPreprintCC BY 4.0Full textCondensed matterFluid dynamicsarXiv
  5. Dosimetric feasibility of lattice radiotherapy for bulky tumors: a paired VMAT vs. helical tomotherapy comparison - the STRYKE trial
    Mottareale R, Serra M, Pezzulla D +18
    Frontiers in OncologyPeer-reviewedCC BY 4.0Full textApplied physics & instrumentationCrossref
BibTeX
@article{may2026linac,
  title = {{Linac performance benchmarking: Feasibility of a web-based tool for comparison of local machine performance check (MPC) results with the global community.}},
  author = {Lauren May and Michael P Barnes and Michael Pearson and Benjamin Zwan and José Antonio Baeza-Ortega and Benjamin Cooper and Mohammad Hussein and Ajeet Mishra and Shiaw Juen Eugene Tan and Jackson Zifodya and Peter B Greer and Catharine H Clark and Joerg Lehmann},
  journal = {Journal of applied clinical medical physics},
  year = {2026},
  month = oct,
  doi = {10.1002/acm2.70892},
  url = {https://europepmc.org/article/MED/42949303},
  note = {CC BY},
}
RIS
TY  - JOUR
TI  - Linac performance benchmarking: Feasibility of a web-based tool for comparison of local machine performance check (MPC) results with the global community.
AU  - Lauren May
AU  - Michael P Barnes
AU  - Michael Pearson
AU  - Benjamin Zwan
AU  - José Antonio Baeza-Ortega
AU  - Benjamin Cooper
AU  - Mohammad Hussein
AU  - Ajeet Mishra
AU  - Shiaw Juen Eugene Tan
AU  - Jackson Zifodya
AU  - Peter B Greer
AU  - Catharine H Clark
AU  - Joerg Lehmann
PY  - 2026
DA  - 2026/10/28
JO  - Journal of applied clinical medical physics
DO  - 10.1002/acm2.70892
UR  - https://europepmc.org/article/MED/42949303
AB  - Background  Varian machine performance check (MPC) is a manufacturer integrated quality control system included as standard for TrueBeam and Halcyon linear accelerators (linacs) to assess performance using a broad range of geometric and beam tests. MPC is a highly standardized method of performance testing, is generally free of inter-user variability and is quick and easy to perform allowing daily acquisitions. Therefore, previous studies have highlighted the potential of comparative analysis of MPC data in investigating differences in performance of individual linacs and identifying outliers. Statistical process control (SPC) is an established approach commonly used to assess stability and behavior of sequential data and identify standard operating ranges which has previously been applied to radiotherapy quality assurance data and MPC data.Purpose  This study aimed to investigate the feasibility of a web-based platform using SPC and aggregated MPC data to allow users to compare their machine performance against that of a cohort of others in the radiotherapy community.Methods  A previously developed web-based platform for radiotherapy data sharing was expanded to include the MPC data collection and comparison tool. Pilot centers exported data from their MPC applications and uploaded it through center-specific portals. Within the application, SPC analysis was used to calculate upper control limits (UCL), lower control limits (LCL) and central lines (CL) for the data sets of each individual linac as well as cohort averages. Interactive plots of longitudinal data for individual machines were displayed for users and overlaid with SPC limits for the specific machine and the cohort averages. Additional analyses were conducted externally from the app where the average and range of SPC control limits for the cohort were compared against MPC tolerances.Results  MPC data for 57 648 measurements taken between April 2016 and May 2026 were collected across seven centers from 21 TrueBeam linacs (15 with Millennium multi-leaf collimator (MLC), six with HD120 MLCs) and one Halcyon. Variation in performance for specific tests was identified for individual machines compared against the cohort as well as for individual MLC leaves for specific machines. Across the cohort of TrueBeam linacs, average control ranges for all geometric tests were within MPC thresholds. However, the range of values for UCL and LCL exceeded tolerances for relative gantry angle. Machines installed before 2021 were found to have slightly larger variability in mean MLC offset with an average control range width of 0.20 mm and 0.18 mm for bank A and B, respectively, compared to machines installed after 2021 of 0.10 mm and 0.13 mm. Analysis of data for individual linacs segmented per year showed a change in mean MLC offset occurring around 2021.Conclusions  The usefulness and feasibility of a tool to compare MPC data against the radiotherapy community was demonstrated. Inter-center comparison of data allowed trends to be identified which would otherwise not be observable for individual centers. Further development and distribution of the MPC data sharing and comparison tool may provide improved confidence and understanding of machine performance and assist low-resourced clinics in delivering safe and accurate radiotherapy treatments.
SN  - 1526-9914
ER  -