Transradial Intracranial Aneurysm Embolization Using a 7F Sheath-Compatible 0.087-Inch Guide Catheter: A Single-Center Retrospective Study
- 1Department of Radiology, Mayo Clinic, Rochester, MN, USA
- 2University of North Dakota School of Medicine & Health Sciences, Grand Forks, ND, USA
- 3Department of Neurosurgery, Mayo Clinic, Rochester, MN, USA
Abstract
Purpose: Transradial intracranial aneurysm embolization is well established, but experience with large-bore 7F sheath-based guide platforms remains comparatively limited. We evaluated the technical feasibility and periprocedural outcomes of a 7F sheath-compatible 0.087-inch guide catheter for transradial intracranial aneurysm embolization.
Materials and Methods: In this single-arm, noncomparative retrospective study, we reviewed consecutive patients who underwent attempted transradial intracranial aneurysm embolization using the NeuGlide as the primary guide catheter at a single center between January 2025 and May 2026. Preaccess radial ultrasonography was performed in all patients, and radial diameter ≥1.7 mm was required for 7F access. The primary outcome was completion of the intended treatment through radial access without femoral conversion.
Results: Target-vessel catheterization was achieved in 71/74 procedures (95.9%), and radial procedural completion in 69/74 (93.2%; 95% confidence interval, 85.1–97.1%). Three procedures required femoral conversion, and 2 were terminated without aneurysm treatment; overall treatment completion was 72/74 (97.3%). Guide instability occurred in 2 procedures: 1 repositioning-only event and 1 event resulting in femoral conversion; no guide exchanges occurred. Seven patients had 9 access-site component events, all managed conservatively; all 6 hematomas were low-grade (EASY I–II). Radial artery occlusion was identified in 1/15 patients formally assessed.
Conclusion: In this radial-first cohort, the 7F sheath-compatible 0.087-inch NeuGlide guide catheter achieved a 93.2% radial procedural completion rate across multiple intracranial aneurysm treatment strategies. These findings support the NeuGlide as a technically feasible and versatile large-bore transradial platform in appropriately selected patients.
Related
- Quantifying procedural burden during stent-assisted coil embolization for unruptured intracranial aneurysms: development of the aneurysm endovascular procedural complexity score and internal validation of a multivariable complication model
- “Pave and Cover” strategy for Rotarex-assisted endovascular repair of thrombosed popliteal artery aneurysm to mitigate distal embolization
- “Pave and Cover” strategy for Rotarex-assisted endovascular repair of thrombosed popliteal artery aneurysm to mitigate distal embolization
- Hybrid Occipital Artery–AICA Bypass Followed by Endovascular Embolization for a Radiation-Induced Dissecting AICA Aneurysm in a Pediatric Patient
- Vascular Tissue-Type Plasminogen Activator Promotes Intracranial Aneurysm Formation
BibTeX
@article{hajj2026transradial,
title = {{Transradial Intracranial Aneurysm Embolization Using a 7F Sheath-Compatible 0.087-Inch Guide Catheter: A Single-Center Retrospective Study}},
author = {Gerard El Hajj and Dani Douri and Jacob Teigen and Charles Garrity and Waleed Brinjikji and Giuseppe Lanzino and Harry Cloft and Lorenzo Rinaldo},
journal = {Neurointervention},
year = {2026},
month = oct,
doi = {10.5469/neuroint.2026.01032},
url = {https://doi.org/10.5469/neuroint.2026.01032},
note = {CC BY-NC 4.0},
}
RIS
TY - JOUR
TI - Transradial Intracranial Aneurysm Embolization Using a 7F Sheath-Compatible 0.087-Inch Guide Catheter: A Single-Center Retrospective Study
AU - Gerard El Hajj
AU - Dani Douri
AU - Jacob Teigen
AU - Charles Garrity
AU - Waleed Brinjikji
AU - Giuseppe Lanzino
AU - Harry Cloft
AU - Lorenzo Rinaldo
PY - 2026
DA - 2026/10/26
JO - Neurointervention
DO - 10.5469/neuroint.2026.01032
UR - https://doi.org/10.5469/neuroint.2026.01032
AB - Purpose: Transradial intracranial aneurysm embolization is well established, but experience with large-bore 7F sheath-based guide platforms remains comparatively limited. We evaluated the technical feasibility and periprocedural outcomes of a 7F sheath-compatible 0.087-inch guide catheter for transradial intracranial aneurysm embolization. Materials and Methods: In this single-arm, noncomparative retrospective study, we reviewed consecutive patients who underwent attempted transradial intracranial aneurysm embolization using the NeuGlide as the primary guide catheter at a single center between January 2025 and May 2026. Preaccess radial ultrasonography was performed in all patients, and radial diameter ≥1.7 mm was required for 7F access. The primary outcome was completion of the intended treatment through radial access without femoral conversion. Results: Target-vessel catheterization was achieved in 71/74 procedures (95.9%), and radial procedural completion in 69/74 (93.2%; 95% confidence interval, 85.1–97.1%). Three procedures required femoral conversion, and 2 were terminated without aneurysm treatment; overall treatment completion was 72/74 (97.3%). Guide instability occurred in 2 procedures: 1 repositioning-only event and 1 event resulting in femoral conversion; no guide exchanges occurred. Seven patients had 9 access-site component events, all managed conservatively; all 6 hematomas were low-grade (EASY I–II). Radial artery occlusion was identified in 1/15 patients formally assessed. Conclusion: In this radial-first cohort, the 7F sheath-compatible 0.087-inch NeuGlide guide catheter achieved a 93.2% radial procedural completion rate across multiple intracranial aneurysm treatment strategies. These findings support the NeuGlide as a technically feasible and versatile large-bore transradial platform in appropriately selected patients.
SN - 2093-9043
SN - 2233-6273
ER -