Skip to content
DraconiticResearch
Search Sources API
203,234 studies31 sourcesUpdated 14 min ago
Peer-reviewedCC BY-NCMedical physics

Spot size reduction and minibeam creation in a proton PBS nozzle using multipole permanent magnets.

  1. 1Department of Radiation Medicine, Loma Linda University, Loma Linda, California, USA.

Abstract

Background

Smaller spot sizes in pencil beam scanning (PBS) nozzles are generally desired to allow more conformal treatments and to spare normal tissue and organs at risk. The production of small diameter proton beams, especially proton minibeams, is challenging because long drift spaces in standard PBS nozzles exacerbate effects of MCS and finite beam divergence.Purpose

To investigate the use of multipole magnets in a proton PBS nozzle to create pencil beams of reduced spot size, including the creation of proton minibeams.Methods

Monte Carlo simulations were performed in a shortened He-filled nozzle incorporating a doublet of octupole magnets and a single quadrupole magnet modeled as Halbach cylinders. Beam diameter reduction was realized by the combination of octupole tail folding and quadrupole magnetic focusing. The octupole doublet was positioned at nozzle entrance upstream of the scanning magnets and the quadrupole was (optionally) positioned at the nozzle exit. The nozzle was configured for four modes of operation, roughly described as standard pencil beam (PB mode), reduced-diameter pencil beam (PBR), symmetric minibeam (MB), and planar minibeam (MBP). Depending on nozzle mode, symmetric beam spots were produced by a continuous rotation (i.e., spinning) of the octupole doublet and the quadrupole magnet about the beam axis. Particle energies investigated covered nearly the full range typically used in clinical applications (70, 100, 150, and 225 MeV). A symmetric Gaussian beam model was used with σ = 3 mm (FWHM = 7 mm) and σ' (divergence) chosen from literature. Pencil beam and reduced diameter pencil beam dose distributions in a water phantom were compared. Minibeam composite dose distributions were created by laterally shifting single beamlet distributions in one or two dimensions, and PVDR and beamlet FWHM were determined. Modulated beams in each nozzle mode were also created.Results

FWHM, FW20M, and penumbra were reduced for all PBR pencil beams compared to unshaped PB beams: at nozzle exit FWHM and FW20M were 19%-38% lower and penumbra P80/20 were 14%-59% lower, and at Bragg depth the differential full widths and penumbra were generally present; with full widths up to 21% smaller, and penumbras up to 15% smaller. The creation of single-layer planar and symmetric minibeams in MBP and MB modes showed minimum beamlet FWHM near water phantom entrance (e.g., 10 mm WED) as low as 0.7-1.4 mm and <2 mm up to a range of 3-4.5 cm of the entrance channel. PVDR was the largest for MBP beams and was as high as 910, whereas PVDR for MB beams was as high as 144. Modulated PBR penumbra were smaller compared to similar PB beams. Modulated MB and MBP minibeams produced clinically relevant PVDR. That study results were achieved ∼1 m downstream of nozzle entrance and beyond is noteworthy.Conclusions

The proposed PBS nozzle could potentially be used to produce pencil beams and modulated beams with smaller spot sizes and penumbras, including proton minibeams, suitable for clinical or preclinical applications.

Related

  1. Youth-led co-creation of a non-communicable disease risk reduction intervention in urban informal settlements of Karachi, Pakistan: a participatory action research
    Romaina Iqbal, Sabahat Naz, Maira Mubashir +3
    BMJ Public HealthPeer-reviewedCC BYFull textEpidemiology & public healthDOAJ
  2. Minibeam radiation therapy supports CAR T therapy in a rat model of CNS lymphoma
    Lorea Iturri, Sarah Potiron, Liangchao Sun +9
    iSciencePeer-reviewedCC BY-NC-ND 4.0Full textNeurologyCell PressCrossrefEurope PMC
  3. Analysis of health impacts associated with adopting permanent standard time or permanent daylight saving time to inform policy in Canada.
    Raymond W Lam, Ralph Mistlberger, Patrick Baylis +1
    CMAJ : Canadian Medical Association journal = journal de l'Association medicale canadiennePeer-reviewedCC BY-NC-NDFull text1 citedPhysiologyEurope PMC
  4. Feasibility and Safety of Transthoracic Echocardiography–Guided Transvenous Permanent Pacemaker Implantation Without Fluoroscopy
    JACC. AdvancesPeer-reviewedCC BYFull textCardiovascular medicineEurope PMC
  5. Clinical and Radiographic Success of Pulpotomy in Permanent Teeth With Irreversible Pulpitis: A Systematic Review
    Omar A Almohsen
    CureusPeer-reviewedCC BYFull textGeneral medicineEurope PMC
BibTeX
@article{mcauley2026spot,
  title = {{Spot size reduction and minibeam creation in a proton PBS nozzle using multipole permanent magnets.}},
  author = {Grant A McAuley},
  journal = {Medical physics},
  year = {2026},
  month = oct,
  doi = {10.1002/mp.70716},
  url = {https://europepmc.org/article/MED/42952162},
  note = {CC BY-NC},
}
RIS
TY  - JOUR
TI  - Spot size reduction and minibeam creation in a proton PBS nozzle using multipole permanent magnets.
AU  - Grant A McAuley
PY  - 2026
DA  - 2026/10/28
JO  - Medical physics
DO  - 10.1002/mp.70716
UR  - https://europepmc.org/article/MED/42952162
AB  - Background  Smaller spot sizes in pencil beam scanning (PBS) nozzles are generally desired to allow more conformal treatments and to spare normal tissue and organs at risk. The production of small diameter proton beams, especially proton minibeams, is challenging because long drift spaces in standard PBS nozzles exacerbate effects of MCS and finite beam divergence.Purpose  To investigate the use of multipole magnets in a proton PBS nozzle to create pencil beams of reduced spot size, including the creation of proton minibeams.Methods  Monte Carlo simulations were performed in a shortened He-filled nozzle incorporating a doublet of octupole magnets and a single quadrupole magnet modeled as Halbach cylinders. Beam diameter reduction was realized by the combination of octupole tail folding and quadrupole magnetic focusing. The octupole doublet was positioned at nozzle entrance upstream of the scanning magnets and the quadrupole was (optionally) positioned at the nozzle exit. The nozzle was configured for four modes of operation, roughly described as standard pencil beam (PB mode), reduced-diameter pencil beam (PBR), symmetric minibeam (MB), and planar minibeam (MBP). Depending on nozzle mode, symmetric beam spots were produced by a continuous rotation (i.e., spinning) of the octupole doublet and the quadrupole magnet about the beam axis. Particle energies investigated covered nearly the full range typically used in clinical applications (70, 100, 150, and 225 MeV). A symmetric Gaussian beam model was used with σ = 3 mm (FWHM = 7 mm) and σ' (divergence) chosen from literature. Pencil beam and reduced diameter pencil beam dose distributions in a water phantom were compared. Minibeam composite dose distributions were created by laterally shifting single beamlet distributions in one or two dimensions, and PVDR and beamlet FWHM were determined. Modulated beams in each nozzle mode were also created.Results  FWHM, FW20M, and penumbra were reduced for all PBR pencil beams compared to unshaped PB beams: at nozzle exit FWHM and FW20M were 19%-38% lower and penumbra P80/20 were 14%-59% lower, and at Bragg depth the differential full widths and penumbra were generally present; with full widths up to 21% smaller, and penumbras up to 15% smaller. The creation of single-layer planar and symmetric minibeams in MBP and MB modes showed minimum beamlet FWHM near water phantom entrance (e.g., 10 mm WED) as low as 0.7-1.4 mm and <2 mm up to a range of 3-4.5 cm of the entrance channel. PVDR was the largest for MBP beams and was as high as 910, whereas PVDR for MB beams was as high as 144. Modulated PBR penumbra were smaller compared to similar PB beams. Modulated MB and MBP minibeams produced clinically relevant PVDR. That study results were achieved ∼1 m downstream of nozzle entrance and beyond is noteworthy.Conclusions  The proposed PBS nozzle could potentially be used to produce pencil beams and modulated beams with smaller spot sizes and penumbras, including proton minibeams, suitable for clinical or preclinical applications.
SN  - 0094-2405
SN  - 2473-4209
ER  -