Validation of a high throughput fluorescent Capillary Electrophoresis Sodium Dodecyl Sulfate method for monoclonal antibody size heterogeneity assessment

  1. 1Agilent Technologies, Inc

Abstract

Fluorescent capillary gel electrophoresis (CGE) with sodium dodecyl sulfate (CE SDS) provides a powerful, high sensitivity alternative to ultraviolet (UV) based detection for characterizing therapeutic monoclonal antibodies (mAb). Regulatory and standards organizations, such as the United States Pharmacopeia (USP), include only UV based CE-SDS methods, hindering adoption of alternative detection methods. There is growing opportunity to expand beyond exclusively UV based CE SDS methods. In this study, we present an analytical validation of a light-emitting diode (LED) fluorescence-based parallel CE SDS method for both non reduced and reduced analysis of therapeutic antibodies. Using the NISTmAb reference material as a model system, size heterogeneity critical quality attributes (CQAs) including monomeric purity (non-reduced conditions), percent glycosylation (reduced conditions), and percent thioether (reduced conditions) were assessed. The fluorescence method demonstrated high specificity and precision with relative standard deviation (RSD) values <1% for monomeric purity and glycosylation, and <3% for thioether, as well as robust performance across variations in injection voltage, electrophoresis voltage, labeling temperature, and Labeling Buffer concentration. Ruggedness testing across users and reagent lots confirmed reproducibility, and accuracy assessments showed strong agreement with reported values from the National Institute of Standards (NIST) and traditional UV detection measurements. Linearity studies yielded coefficient of determination values R2>0.99 for both non reduced and reduced analyses. These results highlight the high sensitivity, stable baseline performance, and suitability of LED fluorescence-based parallel CE SDS as a validated, higher throughput alternative to traditional UV based methods for mAb quality control (QC).

Related

  1. CardioShift: Resolving cell-to-cell functional heterogeneity in hiPSC-derived cardiomyocytes from high-throughput single-cell imaging
    T. C. F. Spanjersberg, C. N. van der Wilt, J.-C. Chang +9
    bioRxivPreprintCC BYFull textCell biologybioRxiv
  2. High-throughput, single-cell-encoded colony imaging reveals infection-linked heterogeneity
    S. G. Suer, T. Doruk, N. Hasanzada +2
    bioRxivPreprintCC BYFull textMicrobiologybioRxiv
  3. Destructive harvest validation of high-throughput measurements show that water use efficiency is unaffected by moderate drought in tobacco
    S. S. Stutz, R. Edquilang, C. J. Bernacchi +1
    bioRxivPreprintCC BY-NC-NDFull textPlant biologybioRxiv
  4. Morphologic intratumoral heterogeneity from routine whole-slide histopathology is prognostic for survival in primary central nervous system lymphoma: development in the LOC Network and international external validation
    L. Rincon de la Rosa, N. Barillot, I. Hernandez-Verdin +39
    bioRxivPreprintCC BY-NC-NDFull textBioinformatics & computational biologybioRxiv
  5. TCA precipitation and ethanol/HCl single-step purification evaluation: One-dimensional gel electrophoresis, bradford assays, spectrofluorometry and Raman spectroscopy data on HSA, Rnase, lysozyme - Mascots and Skyline data
    Balkis Eddhif, Nadia Guignard, Yann Batonneau +4
    Data in BriefPeer-reviewedCC BYFull text10 citedGeneral biochemistrySystems & synthetic biologyHAL
BibTeX
@misc{luttgeharm2026validation,
  title = {{Validation of a high throughput fluorescent Capillary Electrophoresis Sodium Dodecyl Sulfate method for monoclonal antibody size heterogeneity assessment}},
  author = {K. D. Luttgeharm and M. Grover and S.-Y. Huang and W. A. Pike},
  howpublished = {bioRxiv},
  year = {2026},
  month = jul,
  doi = {10.64898/2026.07.15.738750},
  url = {https://www.biorxiv.org/content/10.64898/2026.07.15.738750v2},
  note = {CC BY-ND},
}
RIS
TY  - UNPB
TI  - Validation of a high throughput fluorescent Capillary Electrophoresis Sodium Dodecyl Sulfate method for monoclonal antibody size heterogeneity assessment
AU  - K. D. Luttgeharm
AU  - M. Grover
AU  - S.-Y. Huang
AU  - W. A. Pike
PY  - 2026
DA  - 2026/07/15
PB  - bioRxiv
DO  - 10.64898/2026.07.15.738750
UR  - https://www.biorxiv.org/content/10.64898/2026.07.15.738750v2
AB  - Fluorescent capillary gel electrophoresis (CGE) with sodium dodecyl sulfate (CE SDS) provides a powerful, high sensitivity alternative to ultraviolet (UV) based detection for characterizing therapeutic monoclonal antibodies (mAb). Regulatory and standards organizations, such as the United States Pharmacopeia (USP), include only UV based CE-SDS methods, hindering adoption of alternative detection methods. There is growing opportunity to expand beyond exclusively UV based CE SDS methods. In this study, we present an analytical validation of a light-emitting diode (LED) fluorescence-based parallel CE SDS method for both non reduced and reduced analysis of therapeutic antibodies. Using the NISTmAb reference material as a model system, size heterogeneity critical quality attributes (CQAs) including monomeric purity (non-reduced conditions), percent glycosylation (reduced conditions), and percent thioether (reduced conditions) were assessed. The fluorescence method demonstrated high specificity and precision with relative standard deviation (RSD) values <1% for monomeric purity and glycosylation, and <3% for thioether, as well as robust performance across variations in injection voltage, electrophoresis voltage, labeling temperature, and Labeling Buffer concentration. Ruggedness testing across users and reagent lots confirmed reproducibility, and accuracy assessments showed strong agreement with reported values from the National Institute of Standards (NIST) and traditional UV detection measurements. Linearity studies yielded coefficient of determination values R2>0.99 for both non reduced and reduced analyses. These results highlight the high sensitivity, stable baseline performance, and suitability of LED fluorescence-based parallel CE SDS as a validated, higher throughput alternative to traditional UV based methods for mAb quality control (QC).
ER  -