Evaluation of computer-aided detection products in community and facility tuberculosis screening: 2025 policy update
- 1FIND, Geneva, Switzerland.
- 2World Health Organization, Geneva, Switzerland.
Abstract
Background
Since the World Health Organization (WHO) issued its initial recommendation on computer-aided detection (CAD) use in 2021, the number of commercially available products has expanded substantially, necessitating an updated independent evaluation to inform global policy.Methods
We conducted a standardised, head-to-head evaluation of eight commercially available CAD products for adult tuberculosis (TB) screening using two large, well-characterised, geographically diverse chest radiography (CXR) libraries representing community-based and facility-based screening use cases. CAD products were selected through a WHO public call for expressions of interest and evaluated independently on a secure validation platform with offline installation. Diagnostic accuracy was assessed using receiver operating characteristic (ROC) analysis, partial area under the curve (AUC) in clinically relevant specificity ranges (40-60%), and comparisons with expert radiologist performance and WHO target product profile thresholds.Results
Diagnostic performance varied substantially across products. In community screening, AUCs ranged from 0.72 to 0.84, and in facility screening from 0.77 to 0.86. Six CAD products (qXR, Insight CXR, CAD4TB, InferRead DR Chest, Genki Edge and DrAid) demonstrated optimal diagnostic accuracy as they had overlapping confidence intervals with the expert radiologists' estimates and met WHO performance considerations for screening use, whereas two products (Radify Chest and JVIEWER-X) underperformed. At a fixed sensitivity of 90%, specificities ranged from 26-56% in community screening and 30-49% in facility screening.Conclusions
These findings demonstrate that CAD performance varies markedly across products, underscoring the need for standardised, product-specific evaluation and careful threshold selection. This work constituted evidence informing the WHO policy update on CAD use for TB screening.
Related
- Comparative cost analysis of community-based tuberculosis screening: Computer-Aided Detection (CAD) alone versus CAD with point-of-care C-reactive protein testing in South Africa and Lesotho
- Development and Validation of a Point-of-Care Triage Scorecard to Enhance Tuberculosis Case Detection During Active Community Screening in Yogyakarta, Indonesia
- High burden of cavitary tuberculosis during community-based active case-finding: evaluating the roles of computer-aided radiographic tools and PET-CT
- qPCR-based molecular testing of community-collected tongue swabs for tuberculosis screening among household contacts: a cross-sectional diagnostic performance, yield, and implementation study
- Projected population level impact and cost-effectiveness of clinic and community-based tuberculosis screening approaches
BibTeX
@article{kohli2026evaluation,
title = {{Evaluation of computer-aided detection products in community and facility tuberculosis screening: 2025 policy update}},
author = {Mikashmi Kohli and Aurelien Mace and Cecily Miller and Sam Linsen and Dennis Falzon and Matthew Arentz and Rigveda Kadam and Sandra Vivian Kik and Morten Ruhwald and Technical Advisory Group},
journal = {ERJ open research},
year = {2026},
month = nov,
doi = {10.1183/23120541.00215-2026},
url = {https://europepmc.org/article/PMC/PMC13794295},
note = {CC BY-NC},
}
RIS
TY - JOUR
TI - Evaluation of computer-aided detection products in community and facility tuberculosis screening: 2025 policy update
AU - Mikashmi Kohli
AU - Aurelien Mace
AU - Cecily Miller
AU - Sam Linsen
AU - Dennis Falzon
AU - Matthew Arentz
AU - Rigveda Kadam
AU - Sandra Vivian Kik
AU - Morten Ruhwald
AU - Technical Advisory Group
PY - 2026
DA - 2026/11/01
JO - ERJ open research
DO - 10.1183/23120541.00215-2026
UR - https://europepmc.org/article/PMC/PMC13794295
AB - Background Since the World Health Organization (WHO) issued its initial recommendation on computer-aided detection (CAD) use in 2021, the number of commercially available products has expanded substantially, necessitating an updated independent evaluation to inform global policy.Methods We conducted a standardised, head-to-head evaluation of eight commercially available CAD products for adult tuberculosis (TB) screening using two large, well-characterised, geographically diverse chest radiography (CXR) libraries representing community-based and facility-based screening use cases. CAD products were selected through a WHO public call for expressions of interest and evaluated independently on a secure validation platform with offline installation. Diagnostic accuracy was assessed using receiver operating characteristic (ROC) analysis, partial area under the curve (AUC) in clinically relevant specificity ranges (40-60%), and comparisons with expert radiologist performance and WHO target product profile thresholds.Results Diagnostic performance varied substantially across products. In community screening, AUCs ranged from 0.72 to 0.84, and in facility screening from 0.77 to 0.86. Six CAD products (qXR, Insight CXR, CAD4TB, InferRead DR Chest, Genki Edge and DrAid) demonstrated optimal diagnostic accuracy as they had overlapping confidence intervals with the expert radiologists' estimates and met WHO performance considerations for screening use, whereas two products (Radify Chest and JVIEWER-X) underperformed. At a fixed sensitivity of 90%, specificities ranged from 26-56% in community screening and 30-49% in facility screening.Conclusions These findings demonstrate that CAD performance varies markedly across products, underscoring the need for standardised, product-specific evaluation and careful threshold selection. This work constituted evidence informing the WHO policy update on CAD use for TB screening.
SN - 2312-0541
ER -