Safety and feasibility of a balloon-based pulsed field ablation system for the catheter ablation of paroxysmal atrial fibrillation
- 1TEDA International Cardiovascular Hospital, Tianjin, China
Abstract
Aims: Pulsed field ablation (PFA) is a ablation modality that is increasingly used for pulmonary vein isolation (PVI) in atrial fibrillation (AF) patients. This study assessed the safety and effectiveness of a balloon-based PFA system at 6 months after treatment.
Methods and results: This study is a prospective, single-center, premarket study. A total of 56 patients with drug-refractory paroxysmal atrial fibrillation (PAF) were enrolled between November 2024 and September 2025 and underwent PVI with a balloon-based PFA system. The study endpoints were the rate of primary serious adverse events, acute procedural effectiveness and 6-month freedom from recurrence. Acute effectiveness was achieved in 100% (231/231) of the treated PVs (100% of patients, 56/56), with 33 ± 5.5 PFA applications/patient. The procedure, fluoroscopy, and transpired ablation times were 89.0 ± 30.7, 24.2 ± 8.3, and 20.0 ± 12.0 min, respectively. One (1.8%; 1/56) primary serious adverse event occurred: one patient developed intraoperative gastrointestinal bleeding due to TEE-induced mechanical injury. One patient developed a cerebral infarction on postoperative Day 28. The rate of freedom from documented atrial arrhythmias was 92.9% through 6 months after the index procedure.
Conclusion: The primary results of this study demonstrate the safety and effectiveness of the balloon-based PFA system for performing PVI in patients with PAF.
Supplementary Information: The online version contains supplementary material available at https://doi.org/10.1007/s10840-026-02417-4.
Related
- Differential Recurrence Risk in Paroxysmal Versus Persistent Atrial Fibrillation Post-Catheter Ablation: HFpEF as an Independent Predictor.
- Unipolar electrogram-guided vs. ablation index-guided high-power short-duration pulmonary vein isolation in patients with paroxysmal atrial fibrillation: a single-center randomized controlled trial
- Clinical validation of a novel catheter–tissue contact algorithm in atrial fibrillation ablation using pulsed field ablation
- Predictive Value of Left Atrial Strain Combined with Low-Voltage Substrate for Recurrence After Radiofrequency Catheter Ablation in Persistent Atrial Fibrillation
- Radiofrequency ablation with a conventional solid tip catheter in contact with an implantable cardioverter-defibrillator coil causing inadvertent ventricular fibrillation.
BibTeX
@article{wang2026safety,
title = {{Safety and feasibility of a balloon-based pulsed field ablation system for the catheter ablation of paroxysmal atrial fibrillation}},
author = {Peize Wang and Kun Zhang and Jinlong Bai and Xiuhong Dong and Shuai Yang and Lichao Tian and Zhaoqian Zhang and Zhi Li and Yuhao Miao and Jingjing Liu},
journal = {Journal of Interventional Cardiac Electrophysiology},
year = {2026},
month = aug,
doi = {10.1007/s10840-026-02417-4},
url = {https://doi.org/10.1007/s10840-026-02417-4},
note = {CC BY-NC-ND 4.0},
}
RIS
TY - JOUR
TI - Safety and feasibility of a balloon-based pulsed field ablation system for the catheter ablation of paroxysmal atrial fibrillation
AU - Peize Wang
AU - Kun Zhang
AU - Jinlong Bai
AU - Xiuhong Dong
AU - Shuai Yang
AU - Lichao Tian
AU - Zhaoqian Zhang
AU - Zhi Li
AU - Yuhao Miao
AU - Jingjing Liu
PY - 2026
DA - 2026/08/19
JO - Journal of Interventional Cardiac Electrophysiology
DO - 10.1007/s10840-026-02417-4
UR - https://doi.org/10.1007/s10840-026-02417-4
AB - Aims: Pulsed field ablation (PFA) is a ablation modality that is increasingly used for pulmonary vein isolation (PVI) in atrial fibrillation (AF) patients. This study assessed the safety and effectiveness of a balloon-based PFA system at 6 months after treatment. Methods and results: This study is a prospective, single-center, premarket study. A total of 56 patients with drug-refractory paroxysmal atrial fibrillation (PAF) were enrolled between November 2024 and September 2025 and underwent PVI with a balloon-based PFA system. The study endpoints were the rate of primary serious adverse events, acute procedural effectiveness and 6-month freedom from recurrence. Acute effectiveness was achieved in 100% (231/231) of the treated PVs (100% of patients, 56/56), with 33 ± 5.5 PFA applications/patient. The procedure, fluoroscopy, and transpired ablation times were 89.0 ± 30.7, 24.2 ± 8.3, and 20.0 ± 12.0 min, respectively. One (1.8%; 1/56) primary serious adverse event occurred: one patient developed intraoperative gastrointestinal bleeding due to TEE-induced mechanical injury. One patient developed a cerebral infarction on postoperative Day 28. The rate of freedom from documented atrial arrhythmias was 92.9% through 6 months after the index procedure. Conclusion: The primary results of this study demonstrate the safety and effectiveness of the balloon-based PFA system for performing PVI in patients with PAF. Supplementary Information: The online version contains supplementary material available at https://doi.org/10.1007/s10840-026-02417-4.
SN - 1383-875X
SN - 1572-8595
ER -