A preliminary assessment of a novel antiviral air filter for continuous end-tidal sampling of carbon dioxide for respiratory monitoring of sedated patients: a pilot study.
4 affiliations
- 1Division of Anaesthesiology and Perioperative Medicine, Singapore General Hospital, Outram Road, Singapore, 169608, Singapore.
- 2Department of Anaesthesiology, Sengkang General Hospital, 110 Sengkang E Way, Singapore, 544886, Singapore.
- 3Division of Anaesthesiology and Perioperative Medicine, Singapore General Hospital, Outram Road, Singapore, 169608, Singapore. [email protected].
- 4Department of Anaesthesiology, Sengkang General Hospital, 110 Sengkang E Way, Singapore, 544886, Singapore. [email protected].
Abstract
Background
Monitoring expired end-tidal carbon dioxide (CO₂) is crucial for sedated patients undergoing various procedures. However, a significant concern is the potential contamination of CO₂ sampling lines and gas analyzers when used on patients with respiratory tract infections. To address this, a novel antiviral filter was developed featuring an antimicrobial compound synthesized from cashew skin and husks. Our study aims to assess the device's clinical feasibility for continuous end-tidal CO₂ monitoring.Methods
We conducted a preliminary assessment on 15 adult patients undergoing minor upper limb surgical procedures with local anesthesia, where continuous end-tidal CO₂ monitoring is essential. In all cases, the antiviral filter was positioned above the patient's simple face mask and remained there throughout the entire procedure. The primary outcome was the incidence end-tidal CO₂ capnogram signal disruptions. The secondary outcome was the occurrence of any patient-reported adverse events during device utilization.Results
The median [range] duration of end-tidal CO₂ monitoring using the antiviral filter was 25 [12-38] minutes. Independent observers found no disruptions in the continuous end-tidal CO₂ capnograms across all patients. Furthermore, no patients reported any discomfort or foul smell from the antiviral filter.Conclusion
This pilot feasibility study demonstrates the antiviral filter's potential as an inexpensive, effective, and reliable device for end-tidal CO₂ monitoring in clinical setting. Our findings also advocate for its role in promoting sustainable healthcare practices.
Related
- Development and preliminary clinical application of a portable intelligent respiratory training and monitoring system in patients with stable COPD
- Impact of antenatal corticosteroid exposure on glycemic control in preterm infants assessed by continuous glucose monitoring: preliminary results from a single center study
- Design assessment and preliminary performance evaluation of a novel device for non-invasive emergency treatment of acute respiratory failure.
- Preliminary diagnostic performance of the MOXO continuous performance test in young adults undergoing ADHD assessment
- Assessment of the duration of continuous EEG monitoring in aneurysmal subarachnoid hemorrhage.
BibTeX
@article{tan2026preliminary,
title = {{A preliminary assessment of a novel antiviral air filter for continuous end-tidal sampling of carbon dioxide for respiratory monitoring of sedated patients: a pilot study.}},
author = {Daryl Jian An Tan and Alex Chin Koon Koh and Yew Weng Chan},
journal = {BMC anesthesiology},
year = {2026},
month = aug,
doi = {10.1186/s12871-026-04209-9},
url = {https://europepmc.org/article/MED/43073772},
note = {CC BY-NC-ND},
}
RIS
TY - JOUR
TI - A preliminary assessment of a novel antiviral air filter for continuous end-tidal sampling of carbon dioxide for respiratory monitoring of sedated patients: a pilot study.
AU - Daryl Jian An Tan
AU - Alex Chin Koon Koh
AU - Yew Weng Chan
PY - 2026
DA - 2026/08/27
JO - BMC anesthesiology
DO - 10.1186/s12871-026-04209-9
UR - https://europepmc.org/article/MED/43073772
AB - Background Monitoring expired end-tidal carbon dioxide (CO₂) is crucial for sedated patients undergoing various procedures. However, a significant concern is the potential contamination of CO₂ sampling lines and gas analyzers when used on patients with respiratory tract infections. To address this, a novel antiviral filter was developed featuring an antimicrobial compound synthesized from cashew skin and husks. Our study aims to assess the device's clinical feasibility for continuous end-tidal CO₂ monitoring.Methods We conducted a preliminary assessment on 15 adult patients undergoing minor upper limb surgical procedures with local anesthesia, where continuous end-tidal CO₂ monitoring is essential. In all cases, the antiviral filter was positioned above the patient's simple face mask and remained there throughout the entire procedure. The primary outcome was the incidence end-tidal CO₂ capnogram signal disruptions. The secondary outcome was the occurrence of any patient-reported adverse events during device utilization.Results The median [range] duration of end-tidal CO₂ monitoring using the antiviral filter was 25 [12-38] minutes. Independent observers found no disruptions in the continuous end-tidal CO₂ capnograms across all patients. Furthermore, no patients reported any discomfort or foul smell from the antiviral filter.Conclusion This pilot feasibility study demonstrates the antiviral filter's potential as an inexpensive, effective, and reliable device for end-tidal CO₂ monitoring in clinical setting. Our findings also advocate for its role in promoting sustainable healthcare practices.
SN - 1471-2253
ER -