Effect of small changes in set oxygen saturation target range during routine use of automated FiO₂ control: a randomized physiological study.

  1. 1Department of Biomedical Technology, Faculty of Biomedical Engineering, Czech Technical University in Prague, Kladno, Czechia.
  2. 2Neonatal Unit, Department of Obstetrics and Gynecology, Second Faculty of Medicine, Motol and Homolka University Hospital, Prague, Czechia.
  3. 3Department of Pathological Physiology, First Faculty of Medicine, Charles University, Prague, Czechia.

Abstract

Background

The study aimed to determine what effect small changes in the set target range of an automated FiO₂ system would have on SpO₂ exposure.Methods

Single center randomized crossover design, with auto SpO₂ targets switched between 3 settings (standard, narrow and higher) every 12 h. VLBW infants needing extended respiratory support were studied. Primary outcomes were time in the intended range (88%-93% < 29 weeks PMA, otherwise 90%-95%) and time at SpO₂ extremes (≤80%, >98% with supplemental oxygen). Additional levels of SpO₂ above and below the target range were also evaluated.Results

Ten infants, birth weight 820 grams (IQR 673-943), were evaluated. At entry they were 14 days old (IQR 13-18) and participated for 9-30 days. Four hundred and thirty-one 12-hour epochs were evaluated. The oxygen demand was relatively low [FiO₂ 0.24 (IQR 0.22-0.30)] and oxygen saturation well controlled. They spent a median of 80% (IQR 68-87) of the time in the intended target range, with differences among the three settings not statistically significant. In contrast the % time with SpO₂ ≤ 80% during the higher setting was reduced and the % time with SpO₂ > 98% with the higher setting was increased. Narrow compared to the standard SpO₂ exposures were not different. The descriptive levels of hyperoxemia and hypoxemia were consistent with the two primary SpO₂ extremes.Conclusions

Narrowing the target range had little impact, however shifting the target range up just one point had a marked impact of decreasing hypoxemia and increasing hyperoxemia.Clinical trial registration

ClinicalTrials.gov, identifier NCT06207994.

Related

  1. Microbiological surveillance after automated reprocessing of gastrointestinal endoscopes in routine clinical practice: A prospective randomized microbiological surveillance study
    Masayuki Kato, Hideki Hayashi, Nobuharu Hironaka +6
    PLOS OnePeer-reviewedCC BY 4.0Full textMicrobiologyEpidemiology & public healthPLOS
  2. A Comprehensive Assessment of Physiological and Neural Changes Underlying Cognitive Performance after Heat Stress: A Randomized Clinical Trial.
    Lu Zhou, Zhang Shi, Ye Yao +8
    Environment & health (Washington, D.C.)Peer-reviewedCC BYFull textPhysiologyEurope PMC
  3. Fully automated diaphragm segmentation on routine 2D-cine and 3D thoracic MRI: multireader validation
    Samia Boussouar, Alexiane Pasquier, Damien Vaurs +16
    European radiology experimentalPeer-reviewedCC BYFull textSystems & synthetic biologyMedical imaging & radiologyEurope PMCHAL
  4. Effects of step exercise on human physiological variables, physical fitness, and muscle oxygen saturation in women with sarcopenia: a pilot study
    Jae-Wan Park, Dong-Hyun Yoo, Sang-Geun Jo +1
    Journal of Exercise RehabilitationPeer-reviewedCC BY-NC 4.0Full textPhysiologyPMCEurope PMC
  5. EP15.04 Longitudinal Changes in Distress, Depression, and Anxiety Among Brain Tumor Patients and Their Caregivers: Insights From Routine Psycho-Oncological Outpatient Care
    M Rutenkröger, R Detchev, M Härter +1
    Neuro-OncologyPeer-reviewedPublisher termsFull textOncologyPMC
BibTeX
@article{dornakova2026effect,
  title = {{Effect of small changes in set oxygen saturation target range during routine use of automated FiO₂ control: a randomized physiological study.}},
  author = {Jana Dornakova and Thomas Bachman and Radim Brabec and Ekaterina Orlova and Jan Janota and Jakub Rafl},
  journal = {Frontiers in pediatrics},
  year = {2026},
  month = nov,
  doi = {10.3389/fped.2026.1971581},
  url = {https://europepmc.org/article/MED/43080582},
  note = {CC BY},
}
RIS
TY  - JOUR
TI  - Effect of small changes in set oxygen saturation target range during routine use of automated FiO₂ control: a randomized physiological study.
AU  - Jana Dornakova
AU  - Thomas Bachman
AU  - Radim Brabec
AU  - Ekaterina Orlova
AU  - Jan Janota
AU  - Jakub Rafl
PY  - 2026
DA  - 2026/11/10
JO  - Frontiers in pediatrics
DO  - 10.3389/fped.2026.1971581
UR  - https://europepmc.org/article/MED/43080582
AB  - Background  The study aimed to determine what effect small changes in the set target range of an automated FiO₂ system would have on SpO₂ exposure.Methods  Single center randomized crossover design, with auto SpO₂ targets switched between 3 settings (standard, narrow and higher) every 12 h. VLBW infants needing extended respiratory support were studied. Primary outcomes were time in the intended range (88%-93% < 29 weeks PMA, otherwise 90%-95%) and time at SpO₂ extremes (≤80%, >98% with supplemental oxygen). Additional levels of SpO₂ above and below the target range were also evaluated.Results  Ten infants, birth weight 820 grams (IQR 673-943), were evaluated. At entry they were 14 days old (IQR 13-18) and participated for 9-30 days. Four hundred and thirty-one 12-hour epochs were evaluated. The oxygen demand was relatively low [FiO₂ 0.24 (IQR 0.22-0.30)] and oxygen saturation well controlled. They spent a median of 80% (IQR 68-87) of the time in the intended target range, with differences among the three settings not statistically significant. In contrast the % time with SpO₂ ≤ 80% during the higher setting was reduced and the % time with SpO₂ > 98% with the higher setting was increased. Narrow compared to the standard SpO₂ exposures were not different. The descriptive levels of hyperoxemia and hypoxemia were consistent with the two primary SpO₂ extremes.Conclusions  Narrowing the target range had little impact, however shifting the target range up just one point had a marked impact of decreasing hypoxemia and increasing hyperoxemia.Clinical trial registration  ClinicalTrials.gov, identifier NCT06207994.
SN  - 2296-2360
ER  -