Diagnostic CT-Based Palliative Radiotherapy Using kV Image Verification: A Prospective Feasibility Study.

4 affiliations
  1. 1Department of Radiation Oncology, BC Cancer, Vancouver, BC V5Z 4E6, Canada.
  2. 2Faculty of Medicine, The University of British Columbia, Vancouver, BC V6T 1Z3, Canada.
  3. 3Department of Radiation Therapy, BC Cancer, Vancouver, BC V5Z 4E6, Canada.
  4. 4Department of Medical Physics, BC Cancer, Vancouver, BC V5Z 4E6, Canada.

Abstract

Patients receiving palliative radiotherapy (PRT) often experience substantial symptom burden and may face challenges attending multiple hospital visits. Planning treatment using an existing diagnostic computed tomography (dCT) scan instead of a dedicated computed tomography (CT) simulation may improve access to timely care. This prospective feasibility study evaluated a dCT-based planning workflow using kilovoltage (kV) image verification for patients with bone metastases, with potential applicability to resource-limited settings. Ten patients receiving conventional PRT (8 Gy in 1 fraction or 20 Gy in 5 fractions) with an eligible dCT acquired within 28 days were enrolled. As part of this feasibility study, all patients also underwent standard CT simulation, which served as the dosimetric comparator and fallback treatment pathway. Radiotherapy plans generated on dCT and CT simulation scans were compared for dosimetric quality, while treatment deliverability, patient satisfaction, and estimated carbon emissions were assessed. Seven of ten patients were successfully treated using the dCT workflow. The three remaining patients were transitioned to standard CT simulation treatments due to an uncorrectable dCT image rotation in the planning software (n = 1), scheduling constraints (n = 1), and a larger-than-anticipated difference in patient separation on the treatment unit (n = 1). Target coverage and organ-at-risk doses were comparable between dCT and CT simulation plans, satisfaction among dCT-treated patients was high, and potential avoidance of CT simulation was estimated to reduce carbon emissions by approximately 10 kg carbon dioxide equivalent (CO₂e) per patient. These findings demonstrate that dCT-based planning with kV image verification is feasible, provides dosimetry comparable to conventional CT simulation, and warrants further evaluation as a strategy to improve access to timely PRT while reducing healthcare resource use and environmental impact.

Related

  1. Retrospective intrafraction motion verification using triggered kilovoltage images during bone metastases stereotactic body radiotherapy
    Quentin Techer, Noé Grandgirard, Saturnin Sandjong +2
    Physics and Imaging in Radiation OncologyPeer-reviewedCC BY 4.0Full textSystems & synthetic biologyMedical imaging & radiologyHAL
  2. Feasibility of Urethra-Sparing Ultra-Hypofractionated Three-Dose-Level Radiotherapy for Prostate Cancer with a Dominant Intraprostatic Lesion Abutting the High-Dose Avoidance Zone.
    Annalisa Muscato, Denis La Fauci, Anna Viola +2
    Precision radiation oncologyPeer-reviewedCC BY-NC-NDFull textApplied physics & instrumentationEurope PMC
  3. Evaluating a Large Language Model for Radiotherapy Patient Scheduling under Realistic Clinical and Operational Constraints: a Feasibility Study
    Eric Zhang, Youfang Lai, Wen Li +4
    arXivPreprintCC BY 4.0Full textBiological & medical physicsarXiv
  4. Dosimetric feasibility of lattice radiotherapy for bulky tumors: a paired VMAT vs. helical tomotherapy comparison - the STRYKE trial
    Mottareale R, Serra M, Pezzulla D +18
    Frontiers in OncologyPeer-reviewedCC BY 4.0Full textApplied physics & instrumentationCrossref
  5. Feasibility of reduced PTV margins in magnetic resonance image-guided online adaptive radiotherapy for bladder cancer: a personalised approach.
    Bethany Williams, Saleh Sandoughdaran, Joan Chick +7
    Clinical and translational radiation oncologyPeer-reviewedCC BYFull textApplied physics & instrumentationEurope PMC
BibTeX
@article{wei2026diagnostic,
  title = {{Diagnostic CT-Based Palliative Radiotherapy Using kV Image Verification: A Prospective Feasibility Study.}},
  author = {Kelly Wei and Jason Hui and Bradford Gill and Sheri Grahame and Alexandra Guytingco and Wei Liu and Mitchell Liu and Eric Berthelet and Srinivas Raman and Elaine Wai and Shilo Lefresne},
  journal = {Current oncology (Toronto, Ont.)},
  year = {2026},
  month = sep,
  doi = {10.3390/curroncol33100577},
  url = {https://europepmc.org/article/MED/42930144},
  note = {CC BY},
}
RIS
TY  - JOUR
TI  - Diagnostic CT-Based Palliative Radiotherapy Using kV Image Verification: A Prospective Feasibility Study.
AU  - Kelly Wei
AU  - Jason Hui
AU  - Bradford Gill
AU  - Sheri Grahame
AU  - Alexandra Guytingco
AU  - Wei Liu
AU  - Mitchell Liu
AU  - Eric Berthelet
AU  - Srinivas Raman
AU  - Elaine Wai
AU  - Shilo Lefresne
PY  - 2026
DA  - 2026/09/25
JO  - Current oncology (Toronto, Ont.)
DO  - 10.3390/curroncol33100577
UR  - https://europepmc.org/article/MED/42930144
AB  - Patients receiving palliative radiotherapy (PRT) often experience substantial symptom burden and may face challenges attending multiple hospital visits. Planning treatment using an existing diagnostic computed tomography (dCT) scan instead of a dedicated computed tomography (CT) simulation may improve access to timely care. This prospective feasibility study evaluated a dCT-based planning workflow using kilovoltage (kV) image verification for patients with bone metastases, with potential applicability to resource-limited settings. Ten patients receiving conventional PRT (8 Gy in 1 fraction or 20 Gy in 5 fractions) with an eligible dCT acquired within 28 days were enrolled. As part of this feasibility study, all patients also underwent standard CT simulation, which served as the dosimetric comparator and fallback treatment pathway. Radiotherapy plans generated on dCT and CT simulation scans were compared for dosimetric quality, while treatment deliverability, patient satisfaction, and estimated carbon emissions were assessed. Seven of ten patients were successfully treated using the dCT workflow. The three remaining patients were transitioned to standard CT simulation treatments due to an uncorrectable dCT image rotation in the planning software (n = 1), scheduling constraints (n = 1), and a larger-than-anticipated difference in patient separation on the treatment unit (n = 1). Target coverage and organ-at-risk doses were comparable between dCT and CT simulation plans, satisfaction among dCT-treated patients was high, and potential avoidance of CT simulation was estimated to reduce carbon emissions by approximately 10 kg carbon dioxide equivalent (CO₂e) per patient. These findings demonstrate that dCT-based planning with kV image verification is feasible, provides dosimetry comparable to conventional CT simulation, and warrants further evaluation as a strategy to improve access to timely PRT while reducing healthcare resource use and environmental impact.
SN  - 1198-0052
SN  - 1718-7729
ER  -