Dilution‐Endpoint Comparison of a Flow Cytometric Platelet Activation Assay With PF4/Heparin ELISA and a Lateral Flow Immunoassay in Suspected HIT

  1. 1Slovenian Institute for Transfusion Medicine Ljubljana Slovenia
  2. 2Biotechnical Faculty, University of Ljubljana Ljubljana Slovenia

Abstract

Introduction: Heparin‐induced thrombocytopenia (HIT) is a rare but potentially life‐threatening, immune‐mediated adverse reaction to heparin. PF4‐dependent immunoassays are sensitive screening tests, but they detect both platelet‐activating and nonactivating antibodies; therefore, positive results require interpretation in clinical context and, when available, confirmation with a functional platelet activation assay.

Methods: We compared dilution endpoints of an in‐house flow cytometric platelet activation assay with an IgG‐specific PF4/polyanion ELISA and a PF4‐dependent lateral flow immunoassay (LFA). Six consecutive ELISA‐positive patient sera and one external ELISA‐positive control serum were tested neat and in serial twofold dilutions (up to 1:2048) in parallel with all three methods. Platelet activation was assessed using platelet‐rich plasma from four healthy donors in the presence of low and high doses of heparin and quantified by CD62P expression on CD61‐positive platelets.

Results: Five patient sera and the external control showed heparin‐dependent platelet activation, whereas one ELISA‐positive serum was nonactivating. All three assays were concordantly positive through dilutions up to 1:32 in platelet‐activating samples. At higher dilutions, the LFA remained positive more often than the ELISA or the flow cytometric assay. Across 72 dilution points from platelet‐activating sera and the external control, the ELISA was positive in 47/72, the flow assay in 51/72, and the LFA in 67/72. In several sera, mild dilution increased the ELISA signal and/or platelet activation, consistent with a prozone‐like stoichiometric effect.

Conclusion: In this pilot dilution‐series study, the in‐house flow cytometric assay showed dilution endpoints broadly comparable to the IgG‐specific ELISA for platelet‐activating anti‐PF4/heparin antibodies. The LFA remained positive at lower apparent antibody levels than either the ELISA or the functional assay, but also gave a positive result in the serum that did not contain platelet‐activating anti‐PF4/heparin antibodies.

Related

  1. Development and Internal Validation of a Predictive Score for Nucleic-Acid Amplification Test (NAAT) Positivity in Suspected Clostridioides difficile Infection (CDI) with Glutamate Dehydrogenase (GDH)-Positive and Toxin A/B Enzyme Immunoassay (EIA)-Negative Tests
    Diagnostics (Basel, Switzerland)Peer-reviewedCC BYFull textGeneral medicineEurope PMC
  2. Method comparison and agreement of a fluorescence immunoassay-based HbA1c test with a routine HPLC method
    Harresh Adikesavalu, Neema Raja Beaulah Budithi, Suja Ramanathan +4
    Diabetology internationalPeer-reviewedLicence not statedFull textEndocrinology & metabolismPMC
  3. Flow Cytometric Detection of Bone Marrow Cell Populations Phenotypically Compatible with Neuroblastoma: An Exploratory Comparison with Conventional Methods
    Karolina Bukowska-Strakova, Marta Surman, Izabela Siemińska +9
    CancersPeer-reviewedCC BY 4.0Full textNeurologyCrossrefEurope PMC
  4. In vitro susceptibility of Cutibacterium acnes to dalbavancin: comparison between gradient test and the agar dilution reference method.
    Sara Svensson, Susanne Jacobsson, Erika Matuschek +1
    The Journal of antimicrobial chemotherapyPeer-reviewedCC BYFull textSurgery & anaesthesiaEurope PMC
  5. Comparison of three interpretation criteria for [⁶⁸Ga]Ga-DOTATATE PET/CT in suspected pheochromocytoma and paraganglioma of the adrenal gland or retroperitoneal space.
    Liangqian Tong, Fengshuang Wu, Dong Huang +5
    EJNMMI researchPeer-reviewedCC BY-NC-NDFull textSurgery & anaesthesiaEurope PMCDOAJPMC
BibTeX
@article{malicev2026dilutionendpoint,
  title = {{Dilution‐Endpoint Comparison of a Flow Cytometric Platelet Activation Assay With PF4/Heparin ELISA and a Lateral Flow Immunoassay in Suspected HIT}},
  author = {Elvira Maličev and Matjaž Urbajs and Katerina Jazbec Gradišar and Ana Kolenc and Primož Rožman},
  journal = {International Journal of Laboratory Hematology},
  year = {2026},
  month = jul,
  doi = {10.1111/ijlh.70210},
  url = {https://doi.org/10.1111/ijlh.70210},
  note = {CC BY-NC-ND 4.0},
}
RIS
TY  - JOUR
TI  - Dilution‐Endpoint Comparison of a Flow Cytometric Platelet Activation Assay With PF4/Heparin ELISA and a Lateral Flow Immunoassay in Suspected HIT
AU  - Elvira Maličev
AU  - Matjaž Urbajs
AU  - Katerina Jazbec Gradišar
AU  - Ana Kolenc
AU  - Primož Rožman
PY  - 2026
DA  - 2026/07/27
JO  - International Journal of Laboratory Hematology
DO  - 10.1111/ijlh.70210
UR  - https://doi.org/10.1111/ijlh.70210
AB  - Introduction: Heparin‐induced thrombocytopenia (HIT) is a rare but potentially life‐threatening, immune‐mediated adverse reaction to heparin. PF4‐dependent immunoassays are sensitive screening tests, but they detect both platelet‐activating and nonactivating antibodies; therefore, positive results require interpretation in clinical context and, when available, confirmation with a functional platelet activation assay.  Methods: We compared dilution endpoints of an in‐house flow cytometric platelet activation assay with an IgG‐specific PF4/polyanion ELISA and a PF4‐dependent lateral flow immunoassay (LFA). Six consecutive ELISA‐positive patient sera and one external ELISA‐positive control serum were tested neat and in serial twofold dilutions (up to 1:2048) in parallel with all three methods. Platelet activation was assessed using platelet‐rich plasma from four healthy donors in the presence of low and high doses of heparin and quantified by CD62P expression on CD61‐positive platelets.  Results: Five patient sera and the external control showed heparin‐dependent platelet activation, whereas one ELISA‐positive serum was nonactivating. All three assays were concordantly positive through dilutions up to 1:32 in platelet‐activating samples. At higher dilutions, the LFA remained positive more often than the ELISA or the flow cytometric assay. Across 72 dilution points from platelet‐activating sera and the external control, the ELISA was positive in 47/72, the flow assay in 51/72, and the LFA in 67/72. In several sera, mild dilution increased the ELISA signal and/or platelet activation, consistent with a prozone‐like stoichiometric effect.  Conclusion: In this pilot dilution‐series study, the in‐house flow cytometric assay showed dilution endpoints broadly comparable to the IgG‐specific ELISA for platelet‐activating anti‐PF4/heparin antibodies. The LFA remained positive at lower apparent antibody levels than either the ELISA or the functional assay, but also gave a positive result in the serum that did not contain platelet‐activating anti‐PF4/heparin antibodies.
SN  - 1751-5521
SN  - 1751-553X
ER  -