Skip to content
DraconiticResearch
Search Sources API
203,234 studies31 sourcesUpdated 53 min ago
Peer-reviewedCC BY 4.0Cureus

Breakage-Related Remanufacture of Mandibular Advancement Devices in Patients With Obstructive Sleep Apnea: A Retrospective Study of 219 Patients

  1. 1Department of Oral and Maxillofacial Surgery, Kanazawa Medical University, Uchinada, JPN
  2. 2Department of Physical and Cognitive Rehabilitation Technology, Kanazawa Medical University Hospital, Uchinada, JPN
  3. 3Department of Pathology Ⅱ, Kanazawa Medical University, Uchinada, JPN

Abstract

Introduction: Mandibular advancement devices (MADs) are used for obstructive sleep apnea (OSA), but the frequency of new-device fabrication following breakage during treatment is uncertain.

Methods: We analyzed 219 patients treated with a custom-made fixed-type MAD at a single center. The endpoint was the first documented new-device fabrication due to device breakage, across successive appliances if necessary, measured from the initial recorded fabrication date. The record term “repair” denoted new-device fabrication, whereas adjustment of an existing device was not an event. Dental-treatment and loss-related replacements did not end patient-level observation. For patients without a documented breakage event, absence of a return visit was assumed to mean no recorded event through March 31, 2025.

Results: Fifty-eight of 219 patients (26.5%) had a first breakage-related remanufacture; 161 were administratively censored. Two patients had an intervening new device for a non-device reason: one later had a breakage event and one did not. Median analytic observation time was 45.9 months (IQR 18.4-77.3; range 0.3-180.7). Estimated event-free survival at one and three years was 91.6% (95% CI 87.0-94.6) and 81.1% (95% CI 74.8-86.0), respectively. Actual last-contact dates and loss to clinical follow-up could not be determined.

Conclusion: Breakage-related new-device fabrication was documented during MAD treatment. These patient-level estimates depend on an unverified assumption that patients who did not return had no recorded event; they do not describe the physical lifespan of a single appliance or treatment efficacy.

Related

  1. Effects of mandibular advancement device versus continuous positive airway pressure on cardiac remodeling in obstructive sleep apnea: a randomized CMR substudy
    Nithin R. Iyer, Yi-Hui Ou, Jennifer A. Bryant +8
    Scientific ReportsPeer-reviewedCC BY-NC-ND 4.0Full textPhysiologyPMC
  2. Maxillomandibular Advancement Guided by the Barcelona Line, a Facial Soft-Tissue Reference, in Patients with Obstructive Sleep Apnea: A Feasibility Study.
    Federico Hernández-Alfaro, Maria Cuevas-Nunez, Francesc Duran-Valles +3
    Craniomaxillofacial trauma & reconstructionPeer-reviewedCC BYFull textPhysiologyEurope PMC
  3. Time to diagnosis and treatment of obstructive sleep apnoea using mandibular jaw movement monitoring versus polysomnography: an open-label, multicentre, randomised, controlled trial
    Jean-Louis Pépin, Renaud Tamisier, Marc Manceau +74
    The Lancet Regional Health - EuropePeer-reviewedCC BYPhysiologyHAL
  4. Association of Mandibular Morphological and Trabecular Bone Characteristics with Long-Term Outcomes Following First Permanent Molar Extraction: A Retrospective Study
    Aybike Bas Ozturk, Gelengul Urvasizoglu, Rama Abdulkader +2
    Diagnostics (Basel, Switzerland)Peer-reviewedCC BYFull textGeneral medicineEurope PMC
  5. Functional Capacity Beyond 350 m in Patients with Chronic Obstructive Pulmonary Disease: A Retrospective Analytical Study
    Stephanie B. Sant’ana, José J. S. Oliveira, Angela A. Carvalho +8
    Medical sciences (Basel, Switzerland)Peer-reviewedCC BYFull textGeneral medicineEurope PMC
BibTeX
@article{kakuda2026breakagerelated,
  title = {{Breakage-Related Remanufacture of Mandibular Advancement Devices in Patients With Obstructive Sleep Apnea: A Retrospective Study of 219 Patients}},
  author = {Shintaro Kakuda and Shuto Ito and Yasuhisa Sawai and Yota Yamauchi and Fukunori Toukai and Satoshi Wada and Takanobu Takata and Eiji Mitate and Hiroyuki Nakano},
  journal = {Cureus},
  year = {2026},
  month = oct,
  doi = {10.7759/cureus.117448},
  url = {https://doi.org/10.7759/cureus.117448},
  note = {CC BY 4.0},
}
RIS
TY  - JOUR
TI  - Breakage-Related Remanufacture of Mandibular Advancement Devices in Patients With Obstructive Sleep Apnea: A Retrospective Study of 219 Patients
AU  - Shintaro Kakuda
AU  - Shuto Ito
AU  - Yasuhisa Sawai
AU  - Yota Yamauchi
AU  - Fukunori Toukai
AU  - Satoshi Wada
AU  - Takanobu Takata
AU  - Eiji Mitate
AU  - Hiroyuki Nakano
PY  - 2026
DA  - 2026/10/04
JO  - Cureus
DO  - 10.7759/cureus.117448
UR  - https://doi.org/10.7759/cureus.117448
AB  - Introduction: Mandibular advancement devices (MADs) are used for obstructive sleep apnea (OSA), but the frequency of new-device fabrication following breakage during treatment is uncertain.  Methods: We analyzed 219 patients treated with a custom-made fixed-type MAD at a single center. The endpoint was the first documented new-device fabrication due to device breakage, across successive appliances if necessary, measured from the initial recorded fabrication date. The record term “repair” denoted new-device fabrication, whereas adjustment of an existing device was not an event. Dental-treatment and loss-related replacements did not end patient-level observation. For patients without a documented breakage event, absence of a return visit was assumed to mean no recorded event through March 31, 2025.  Results: Fifty-eight of 219 patients (26.5%) had a first breakage-related remanufacture; 161 were administratively censored. Two patients had an intervening new device for a non-device reason: one later had a breakage event and one did not. Median analytic observation time was 45.9 months (IQR 18.4-77.3; range 0.3-180.7). Estimated event-free survival at one and three years was 91.6% (95% CI 87.0-94.6) and 81.1% (95% CI 74.8-86.0), respectively. Actual last-contact dates and loss to clinical follow-up could not be determined.  Conclusion: Breakage-related new-device fabrication was documented during MAD treatment. These patient-level estimates depend on an unverified assumption that patients who did not return had no recorded event; they do not describe the physical lifespan of a single appliance or treatment efficacy.
SN  - 2168-8184
ER  -