E-norms at Ten: What a Decade Taught the Method About Deriving Reference Values from Patient Data

  1. 1Tufts University School of Medicine

Abstract

Introduction/AimsThe e-norms method derives reference values from a laboratorys own mixed patient data, without recruiting healthy volunteers. A decade of use exposed three things the original description did not address: a dependence on how finely measurements are recorded, a minimum sample size below which derived limits are systematically too permissive, and an extension of the method to engineered composite features. This work characterizes the first two and describes a corrected plateau rule.

MethodsTwenty-six sensory amplitude datasets (18 to 3,301 measurements), three sensory conduction velocity datasets, nine median-nerve datasets and a 2,000-measurement peroneal file were analyzed. Amplitudes were progressively rounded; velocities were made finer by adding noise of 0.05 m/s. Subsamples of 50 to 750 were drawn from thirteen large datasets on raw, logarithmic and square-root scales. The corrected rule was validated against expert visual plateau selection and against the plateau proportions reported in 2015, and limit stability was estimated by resampling.

ResultsAt n = 50, amplitudes as recorded returned 0.28 of the full-sample limit with 73.1% failing; rounded to whole numbers they returned 1.03 with none failing. Integer velocities returned 1.03 with no failures, but failed in all 300 draws once noise was added. Derived limits were systematically too low below 250 to 300 measurements. The corrected rule returned 1.00 with 0.1% failures and left integer velocities essentially unchanged.

DiscussionThe dependence is intrinsic to locating the plateau from gaps between sorted values. Obtaining 250 normal studies means accumulating 530 to 1,250, depending on a laboratorys normal rate.

Related

  1. Ensemble‐based label adjudication (EBLA): A method for validating continuous physiological measures against observer‐dependent reference standards
    Guillermo Gutierrez, Fernando Papa
    Physiological reportsPeer-reviewedCC BYFull textGeneral medicineEurope PMC
  2. In vitro susceptibility of Cutibacterium acnes to dalbavancin: comparison between gradient test and the agar dilution reference method.
    Sara Svensson, Susanne Jacobsson, Erika Matuschek +1
    The Journal of antimicrobial chemotherapyPeer-reviewedCC BYFull textSurgery & anaesthesiaEurope PMC
  3. “We follow what our elders taught us”: an ethnographic study of kinship, rituals, and reproductive transitions among women in the GaRBH cohort, Rajasthan, India
    Sonia Maurya, Gagandeep Kaur Walia, Subhash Kumar Jha +4
    Frontiers in global women's healthPeer-reviewedCC BYFull textGeneral medicineEurope PMC
  4. Reference method matters: DXA system differences influence agreement with 3D optical imaging body composition estimates
    Sophia Ramirez, Miriam Bustami, Christian Rodriguez +5
    International Journal of Obesity (2005)Peer-reviewedCC BY 4.0Full text1 citedPhysiologyPMC
  5. Harmonization of Lipoprotein(a) Immunoassays with the IFCC- Endorsed Reference Measurement Procedure Validated in Patient Samples and Multiple Routine testing as A Second Step Towards Apolipoprotein Standardization
    Keita Yamashita, Mitsuru Wakita, Noriyasu Niizeki +16
    Journal of Atherosclerosis and ThrombosisPeer-reviewedCC BY-NC-SAFull text1 citedSurgery & anaesthesiaJ-STAGE
BibTeX
@misc{jabre2026enorms,
  title = {{E-norms at Ten: What a Decade Taught the Method About Deriving Reference Values from Patient Data}},
  author = {J. F. Jabre},
  howpublished = {medRxiv},
  year = {2026},
  month = aug,
  doi = {10.64898/2026.08.30.26361424},
  url = {https://www.medrxiv.org/content/10.64898/2026.08.30.26361424v1},
  note = {CC BY-NC-ND},
}
RIS
TY  - UNPB
TI  - E-norms at Ten: What a Decade Taught the Method About Deriving Reference Values from Patient Data
AU  - J. F. Jabre
PY  - 2026
DA  - 2026/08/30
PB  - medRxiv
DO  - 10.64898/2026.08.30.26361424
UR  - https://www.medrxiv.org/content/10.64898/2026.08.30.26361424v1
AB  - Introduction/AimsThe e-norms method derives reference values from a laboratorys own mixed patient data, without recruiting healthy volunteers. A decade of use exposed three things the original description did not address: a dependence on how finely measurements are recorded, a minimum sample size below which derived limits are systematically too permissive, and an extension of the method to engineered composite features. This work characterizes the first two and describes a corrected plateau rule.  MethodsTwenty-six sensory amplitude datasets (18 to 3,301 measurements), three sensory conduction velocity datasets, nine median-nerve datasets and a 2,000-measurement peroneal file were analyzed. Amplitudes were progressively rounded; velocities were made finer by adding noise of 0.05 m/s. Subsamples of 50 to 750 were drawn from thirteen large datasets on raw, logarithmic and square-root scales. The corrected rule was validated against expert visual plateau selection and against the plateau proportions reported in 2015, and limit stability was estimated by resampling.  ResultsAt n = 50, amplitudes as recorded returned 0.28 of the full-sample limit with 73.1% failing; rounded to whole numbers they returned 1.03 with none failing. Integer velocities returned 1.03 with no failures, but failed in all 300 draws once noise was added. Derived limits were systematically too low below 250 to 300 measurements. The corrected rule returned 1.00 with 0.1% failures and left integer velocities essentially unchanged.  DiscussionThe dependence is intrinsic to locating the plateau from gaps between sorted values. Obtaining 250 normal studies means accumulating 530 to 1,250, depending on a laboratorys normal rate.
ER  -