Operative fire risk and supplemental oxygen use in ophthalmic surgery
- 1University of Texas McGovern Medical School, Houston, Texas, USA
Abstract
Objective: This study aimed to replicate an open system delivery of oxygen while measuring periocular oxygen levels. This study aimed to determine the duration required for periocular oxygen levels to return to a safe threshold following the cessation of supplemental oxygen in the operative setting.
Methods and analysis: A nasal cannula connected to an oxygen tank was placed in the nostrils of an adult volunteer (AV) and adult-size mannequin head (MH). A sensor was secured in the periocular region to measure oxygen levels. After oxygen flow was initiated, periocular oxygen concentration was allowed to stabilise, and this value was recorded as the initial concentration. Oxygen was then discontinued, and levels were recorded every 5 s. Flow rates were 1, 2, 3, 4 and 6 L/min. The effect of surgical draping was evaluated with towel draping (TD) and 1020 eye drapes (EDs).
Results: The highest mean initial periocular oxygen levels across flow rates for the MH were 48.5% (±0.3%) with no draping, 71.5% (±<0.1%) with TD and 56.0% (±<0.1%) with ED. For the AV, they were 21.3% (±0.1%) with no draping, 51.6% (±0.6%) with TD and 54.7% (±1.0%) with ED. Periocular oxygen fell to 25% or below within 20 s for the MH. For the AV, it took up to 45 s to reach 25% with TD and up to 35 s with ED.
Conclusion: Supplemental oxygen during electrocautery can increase the risk of operative fire in ophthalmic surgery. TD and EDs further increased periocular oxygen concentrations and prolonged oxygen clearance.
Related
- Development of OphA-PRISM for Preoperative Prediction of Visual Outcomes and Operative Demands in Ophthalmic-Segment Aneurysm Microsurgery
- Air versus supplemental oxygen for resuscitation of term or late preterm infants at birth
- Postoperative opioid-induced respiratory depression monitoring in the context of supplemental oxygen: current limitations and emerging machine-learning approaches
- Changes in quality of life and exercise capacity following initiation of supplemental oxygen in patients with idiopathic pulmonary fibrosis.
- Comparison of the Effects of 0.05% Difluprednate Ophthalmic Solution and 1% Prednisolone Acetate Ophthalmic Solution on Intraocular Inflammation Following Lens Surgery in Healthy Rabbits.
BibTeX
@article{kozman2026operative,
title = {{Operative fire risk and supplemental oxygen use in ophthalmic surgery}},
author = {Mena Kozman and Dean McCulley and Amelia McCulley and Alice Chuang and Martha Meyers and Agni Kakouri and Timothy McCulley and Ying Chen},
journal = {BMJ Open Ophthalmology},
year = {2026},
month = nov,
doi = {10.1136/bmjophth-2026-003057},
url = {https://doi.org/10.1136/bmjophth-2026-003057},
note = {CC BY-NC 4.0},
}
RIS
TY - JOUR
TI - Operative fire risk and supplemental oxygen use in ophthalmic surgery
AU - Mena Kozman
AU - Dean McCulley
AU - Amelia McCulley
AU - Alice Chuang
AU - Martha Meyers
AU - Agni Kakouri
AU - Timothy McCulley
AU - Ying Chen
PY - 2026
DA - 2026/11/16
JO - BMJ Open Ophthalmology
DO - 10.1136/bmjophth-2026-003057
UR - https://doi.org/10.1136/bmjophth-2026-003057
AB - Objective: This study aimed to replicate an open system delivery of oxygen while measuring periocular oxygen levels. This study aimed to determine the duration required for periocular oxygen levels to return to a safe threshold following the cessation of supplemental oxygen in the operative setting. Methods and analysis: A nasal cannula connected to an oxygen tank was placed in the nostrils of an adult volunteer (AV) and adult-size mannequin head (MH). A sensor was secured in the periocular region to measure oxygen levels. After oxygen flow was initiated, periocular oxygen concentration was allowed to stabilise, and this value was recorded as the initial concentration. Oxygen was then discontinued, and levels were recorded every 5 s. Flow rates were 1, 2, 3, 4 and 6 L/min. The effect of surgical draping was evaluated with towel draping (TD) and 1020 eye drapes (EDs). Results: The highest mean initial periocular oxygen levels across flow rates for the MH were 48.5% (±0.3%) with no draping, 71.5% (±<0.1%) with TD and 56.0% (±<0.1%) with ED. For the AV, they were 21.3% (±0.1%) with no draping, 51.6% (±0.6%) with TD and 54.7% (±1.0%) with ED. Periocular oxygen fell to 25% or below within 20 s for the MH. For the AV, it took up to 45 s to reach 25% with TD and up to 35 s with ED. Conclusion: Supplemental oxygen during electrocautery can increase the risk of operative fire in ophthalmic surgery. TD and EDs further increased periocular oxygen concentrations and prolonged oxygen clearance.
SN - 2397-3269
ER -