Staged Transcatheter Valve-in-Valve Replacement for Simultaneous Failure of 3 Surgical Bioprosthetic Valves
- 1Division of Structural Heart, Xiamen Cardiovascular Hospital Xiamen University, Xiamen, China
Abstract
Objectives: To describe a heart team-driven staged transcatheter strategy for accelerated triple bioprosthetic valve failure in a patient with prohibitive surgical risk.
Key Steps: A 78-year-old woman with systemic lupus erythematosus, stage 4 chronic kidney disease, permanent atrial fibrillation, and severe frailty presented with NYHA functional class III heart failure. She had undergone triple surgical bioprosthetic heart valve replacement 8 years earlier. Imaging demonstrated severe aortic, mitral, and tricuspid bioprosthetic valve failure. Given prohibitive surgical risk, a staged transcatheter strategy was selected. Valve-in-valve transcatheter aortic valve replacement with a 23-mm supra-annular self-expanding valve and bioprosthetic valve fracture was performed first, followed by valve-in-valve (ViV)–transcatheter mitral valve replacement and ViV transcatheter tricuspid valve replacement with 26-mm balloon-expandable valves. All procedures were successful without complications. At 2 months, NYHA functional class improved from III to II, with satisfactory function of all 3 valves.
Potential Pitfalls: Coronary obstruction, patient-prosthesis mismatch, neo–left ventricular outflow tract obstruction, and hemodynamic intolerance.
Take-Home Messages: Staged triple ViV intervention may provide a feasible alternative to redo surgery. Multimodality imaging and individualized sequencing are critical.
Related
- ECMO-assisted valve-in-valve transcatheter aortic valve replacement following surgical aortic valve replacement failure in triple-positive antiphospholipid syndrome and systemic lupus erythematosus-associated libman-sacks endocarditis: a case report
- Valve-in-valve transcatheter mitral valve replacement via totally endoscopic endoclamp-assisted approach after failed bioprosthetic mitral valve: a case report.
- Comparative Outcomes of Transcatheter Versus Surgical Aortic Valve Replacement in Patients With Heart Failure and Chronic Kidney Disease
- Simultaneous biventricular pressure-volume loop analysis during transcatheter tricuspid valve replacement
- Surgical aortic valve replacement-level aortotomy for self-expanding transcatheter aortic valve explantation
BibTeX
@article{castillo2026staged,
title = {{Staged Transcatheter Valve-in-Valve Replacement for Simultaneous Failure of 3 Surgical Bioprosthetic Valves}},
author = {Dominador V. del Castillo and Yan Wang},
journal = {JACC Case Reports},
year = {2026},
month = nov,
doi = {10.1016/j.jaccas.2026.110544},
url = {https://doi.org/10.1016/j.jaccas.2026.110544},
note = {CC BY 4.0},
}
RIS
TY - JOUR
TI - Staged Transcatheter Valve-in-Valve Replacement for Simultaneous Failure of 3 Surgical Bioprosthetic Valves
AU - Dominador V. del Castillo
AU - Yan Wang
PY - 2026
DA - 2026/11/11
JO - JACC Case Reports
DO - 10.1016/j.jaccas.2026.110544
UR - https://doi.org/10.1016/j.jaccas.2026.110544
AB - Objectives: To describe a heart team-driven staged transcatheter strategy for accelerated triple bioprosthetic valve failure in a patient with prohibitive surgical risk. Key Steps: A 78-year-old woman with systemic lupus erythematosus, stage 4 chronic kidney disease, permanent atrial fibrillation, and severe frailty presented with NYHA functional class III heart failure. She had undergone triple surgical bioprosthetic heart valve replacement 8 years earlier. Imaging demonstrated severe aortic, mitral, and tricuspid bioprosthetic valve failure. Given prohibitive surgical risk, a staged transcatheter strategy was selected. Valve-in-valve transcatheter aortic valve replacement with a 23-mm supra-annular self-expanding valve and bioprosthetic valve fracture was performed first, followed by valve-in-valve (ViV)–transcatheter mitral valve replacement and ViV transcatheter tricuspid valve replacement with 26-mm balloon-expandable valves. All procedures were successful without complications. At 2 months, NYHA functional class improved from III to II, with satisfactory function of all 3 valves. Potential Pitfalls: Coronary obstruction, patient-prosthesis mismatch, neo–left ventricular outflow tract obstruction, and hemodynamic intolerance. Take-Home Messages: Staged triple ViV intervention may provide a feasible alternative to redo surgery. Multimodality imaging and individualized sequencing are critical.
SN - 2666-0849
ER -