Bradycardia-Dependent Loss of Atrial Capture After Single-Chamber Leadless Atrial Pacemaker Implantation
- 1Division of Cardiology, San Antonio Military Medical Center, San Antonio, Texas, USA
- 2Military Cardiovascular Outcomes Research (MiCOR), Uniformed Services University of the Health Sciences, Bethesda, Maryland, USA
Abstract
Background: Bradycardia-dependent pacing threshold elevation is a rarely reported cause of early postimplant loss of capture, with only 1 prior case reported in an atrial leadless pacemaker.
Case Summary: 88-year-old woman with sinus pauses underwent Aveir AR implantation. On postoperative day 1, intermittent loss of atrial capture occurred at the programmed lower rate limit. Interrogation revealed stable impedance, current of injury, and sensing. Threshold testing revealed marked rate-dependence: 2.75 V/0.4 milliseconds at 60 beats/min vs 0.75 V/0.4 milliseconds at 90 beats/min. Fluoroscopy confirmed device stability, and output was increased conservatively.
Discussion: Proposed mechanisms of this phenomenon include rate-dependent variation in the electrode-tissue interface and phase 4 block from diastolic depolarization in injured myocardium. When early intermittent capture loss occurs with stable device parameters, threshold testing across pacing rates should be performed to assess for bradycardia-dependent threshold elevation.
Take-Home Message: Bradycardia-dependent capture threshold elevation is a possible etiology of early postimplant loss of atrial capture after Aveir AR implantation.
Related
- Leadless Pacemaker Implantation in Fabry Disease With Permanent Atrial Fibrillation and Bradycardia: Diagnostic and Implantation Guidance by Preprocedural Cardiac Magnetic Resonance Imaging.
- Challenging case of leadless pacemaker implantation for a patient with left-sided inferior vena cava.
- Internal jugular vein approach facilitating second leadless pacemaker implantation after battery depletion
- Direct-visualization implantation of an atrial leadless pacemaker at tricuspid valve surgery.
- Contrast-Free Dual-Chamber Leadless Pacemaker Implantation Using Intracardiac Echocardiography and 3-Dimensional Electroanatomical Mapping
BibTeX
@article{soderstrom2026bradycardiadependent,
title = {{Bradycardia-Dependent Loss of Atrial Capture After Single-Chamber Leadless Atrial Pacemaker Implantation}},
author = {Matthew A. Soderstrom and Ariana Ramirez and John Henry Dean and Mark Haigney and Eric Martinez and Cory Madigan},
journal = {JACC Case Reports},
year = {2026},
month = sep,
doi = {10.1016/j.jaccas.2026.110232},
url = {https://doi.org/10.1016/j.jaccas.2026.110232},
note = {CC BY-NC-ND 4.0},
}
RIS
TY - JOUR
TI - Bradycardia-Dependent Loss of Atrial Capture After Single-Chamber Leadless Atrial Pacemaker Implantation
AU - Matthew A. Soderstrom
AU - Ariana Ramirez
AU - John Henry Dean
AU - Mark Haigney
AU - Eric Martinez
AU - Cory Madigan
PY - 2026
DA - 2026/09/15
JO - JACC Case Reports
DO - 10.1016/j.jaccas.2026.110232
UR - https://doi.org/10.1016/j.jaccas.2026.110232
AB - Background: Bradycardia-dependent pacing threshold elevation is a rarely reported cause of early postimplant loss of capture, with only 1 prior case reported in an atrial leadless pacemaker. Case Summary: 88-year-old woman with sinus pauses underwent Aveir AR implantation. On postoperative day 1, intermittent loss of atrial capture occurred at the programmed lower rate limit. Interrogation revealed stable impedance, current of injury, and sensing. Threshold testing revealed marked rate-dependence: 2.75 V/0.4 milliseconds at 60 beats/min vs 0.75 V/0.4 milliseconds at 90 beats/min. Fluoroscopy confirmed device stability, and output was increased conservatively. Discussion: Proposed mechanisms of this phenomenon include rate-dependent variation in the electrode-tissue interface and phase 4 block from diastolic depolarization in injured myocardium. When early intermittent capture loss occurs with stable device parameters, threshold testing across pacing rates should be performed to assess for bradycardia-dependent threshold elevation. Take-Home Message: Bradycardia-dependent capture threshold elevation is a possible etiology of early postimplant loss of atrial capture after Aveir AR implantation.
SN - 2666-0849
ER -