Transient Pneumopericardium During Implantation of an Extravascular Implantable Cardioverter-Defibrillator
- 1Department of Cardiology, Caen University Hospital, Normandie Univ, UniCaen, Caen, France
- 2Normandy Heart Rhythm Association, Caen, France
Abstract
Background: Extravascular implantable cardioverter-defibrillators (EV-ICDs) use a substernal route to avoid intravascular lead placement, but substernal tunneling carries a risk of pericardial injury not encountered with transvenous systems.
Case Summary: A 48-year-old man with ischemic cardiomyopathy (ejection fraction 14%) survived an out-of-hospital cardiac arrest due to monomorphic ventricular tachycardia at 240 beats/min. Cardiac magnetic resonance confirmed an extensive nonviable inferior scar. An EV-ICD was implanted for secondary prevention. Substernal tunneling was complicated by pericardial puncture and pneumopericardium. After lead repositioning, intraoperative defibrillation testing (DFT) failed twice, with subsequent sensing deterioration. Therapies were deactivated. Serial chest radiographs confirmed complete pneumopericardium resorption by day 30 with stable lead position. Repeat DFT at 5 weeks succeeded without reintervention. Device parameters remained normal at 1-year follow-up.
Discussion: Pneumopericardium is an under-recognized complication of substernal EV-ICD implantation that can transiently impair sensing and DFT. Conservative management with deferred re-testing illustrated a safe pathway, avoiding unnecessary system revisions.
Related
- Transient defibrillation threshold elevation associated with air in the generator pocket after extravascular implantable cardioverter-defibrillator implantation.
- Successful extravascular implantable cardioverter defibrillator implantation in a patient with ankylosing spondylitis and severe thoracic deformity: a case report.
- Inappropriate ventricular fibrillation detection by extravascular implantable cardioverter-defibrillator.
- Approach to epicardial access for ventricular tachycardia ablation in a patient with an extravascular implantable cardioverter-defibrillator.
- Management of sensing issues with an extravascular and subcutaneous implantable cardioverter-defibrillator in a patient with polymorphic ventricular tachycardia: A case report.
BibTeX
@article{ollitrault2026transient,
title = {{Transient Pneumopericardium During Implantation of an Extravascular Implantable Cardioverter-Defibrillator}},
author = {Pierre Ollitrault and Arnaud Pellissier and Virginie Ferchaud and Jonaz Font and Laure Champ-Rigot},
journal = {JACC Case Reports},
year = {2026},
month = sep,
doi = {10.1016/j.jaccas.2026.110238},
url = {https://doi.org/10.1016/j.jaccas.2026.110238},
note = {CC BY-NC-ND 4.0},
}
RIS
TY - JOUR
TI - Transient Pneumopericardium During Implantation of an Extravascular Implantable Cardioverter-Defibrillator
AU - Pierre Ollitrault
AU - Arnaud Pellissier
AU - Virginie Ferchaud
AU - Jonaz Font
AU - Laure Champ-Rigot
PY - 2026
DA - 2026/09/03
JO - JACC Case Reports
DO - 10.1016/j.jaccas.2026.110238
UR - https://doi.org/10.1016/j.jaccas.2026.110238
AB - Background: Extravascular implantable cardioverter-defibrillators (EV-ICDs) use a substernal route to avoid intravascular lead placement, but substernal tunneling carries a risk of pericardial injury not encountered with transvenous systems. Case Summary: A 48-year-old man with ischemic cardiomyopathy (ejection fraction 14%) survived an out-of-hospital cardiac arrest due to monomorphic ventricular tachycardia at 240 beats/min. Cardiac magnetic resonance confirmed an extensive nonviable inferior scar. An EV-ICD was implanted for secondary prevention. Substernal tunneling was complicated by pericardial puncture and pneumopericardium. After lead repositioning, intraoperative defibrillation testing (DFT) failed twice, with subsequent sensing deterioration. Therapies were deactivated. Serial chest radiographs confirmed complete pneumopericardium resorption by day 30 with stable lead position. Repeat DFT at 5 weeks succeeded without reintervention. Device parameters remained normal at 1-year follow-up. Discussion: Pneumopericardium is an under-recognized complication of substernal EV-ICD implantation that can transiently impair sensing and DFT. Conservative management with deferred re-testing illustrated a safe pathway, avoiding unnecessary system revisions.
SN - 2666-0849
ER -