Assay or astray? discrepancies in efanesoctocog alfa monitoring

  1. 1Laboratoire d’Hématologie, Unité Hémostase, Hôpitaux Universitaires de Strasbourg, Strasbourg, France
  2. 2Centre de Ressource et Compétences des Maladies Hémorragiques Congénitales, Hôpitaux Universitaires de Strasbourg, Strasbourg, France

Abstract

Background

Efanesoctocog alfa (EFA) is an ultraextended half-life factor VIII (FVIII) with a unique structure that may influence laboratory monitoring. Although assay discrepancies have been described in spiked samples, data in patient samples are limited.Objectives

To assess the performance of seven FVIII assays for measuring EFA in patient plasma and compare results with those from spiked samples.Methods

FVIII activity was measured using four one-stage assays (OSAs) and three chromogenic assays (CSAs) on two analyzers. Assays were calibrated either with human plasma or an EFA-specific calibrator. Measurements were performed in FVIII-deficient plasma spiked with EFA and in samples from patients with hemophilia A treated with EFA. Results were compared with a reference assay (Actin FSL) and expressed as recovery percentages (acceptable range: 80%-120%).Results

In spiked samples, plasma-calibrated CSA overestimated FVIII activity (+37% to +94%), whereas EFA-calibrated CSA showed good agreement with the reference assay (+2% to +16%). OSA results were generally acceptable, with slight underestimation depending on the reagent. In patient samples, patterns differed markedly: most assays deviated significantly from the reference, and none remained within the acceptable recovery range across the full FVIII spectrum. Compared with spiked samples, discrepancies were more pronounced at later time points and trough levels, where OSA tended to overestimate FVIII activity, whereas EFA-calibrated CSA underestimated activity.Conclusion

Assay performance for EFA differs between spiked and patient samples, underscoring the limitations of in vitro evaluations. No assay provided consistent agreement with the reference assay across the full FVIII range, highlighting the need for cautious interpretation and improved standardization of EFA monitoring in clinical practice.

Related

  1. Efanesoctocog Alfa for the Management of a Life-Threatening Intracranial Haemorrhage in Mild Haemophilia A: A Case Report.
    Laura C Seoane, Paula S Fuentes, Patricia S Diaz +1
    CureusPeer-reviewedCC BYFull textGeneral medicineEurope PMCPMC
  2. Unmet needs despite standard-of-care prophylaxis in hemophilia A: improved outcomes after switching to efanesoctocog alfa
    Antoine Rauch, Hortense Maynadié, Stéphanie Desage +25
    Blood AdvancesPeer-reviewedCC BY-NC-ND 4.0Full textGeneral medicinePMC
  3. Restoration of spermatogenesis in a male with prolactinoma-induced hypogonadotropic hypogonadism using follitropin delta and recombinant choriogonadotropin alfa: a case report
    Moises Abraham Adel Domínguez, Walter D. Cardona Maya, Andrés Mora Topete
    Journal of Assisted Reproduction and GeneticsPeer-reviewedLicence not statedFull textObstetrics, gynaecology & paediatricsPMC
  4. Comparative outcomes of conventional follitropin delta 15 µg/day versus follitropin alfa 225 IU/day for ovarian stimulation: results from an exploratory post hoc analysis of Spanish participants from the ADAPT-1 randomized controlled trial
    Marta García Sánchez, Andrea Bernabeu, Ana Hernández Cesteros +7
    Frontiers in reproductive healthPeer-reviewedCC BYFull textGeneral medicineEurope PMC
  5. 514 Initial clinical results from START-002: a phase 1b/2 study of invikafusp alfa, a first-in-class dual T-cell agonist, in combination with sacituzumab govitecan in triple-negative or HR+/HER2- mBC
    Ke Liu, Kevin Chin, Zhen Su +14
    Journal for ImmunoTherapy of CancerPeer-reviewedCC BYFull textOncologyDOAJ
BibTeX
@article{herb2026assay,
  title = {{Assay or astray? discrepancies in efanesoctocog alfa monitoring}},
  author = {Agathe Herb and Jordan Wimmer and Sarra Derbal and Manon Dolt and Léa Pierre and Olivier Feugeas and Laurent Mauvieux and Dominique Desprez and Laurent Sattler},
  journal = {Research and practice in thrombosis and haemostasis},
  year = {2026},
  month = oct,
  doi = {10.1016/j.rpth.2026.106994},
  url = {https://europepmc.org/article/PMC/PMC13717027},
  note = {CC BY},
}
RIS
TY  - JOUR
TI  - Assay or astray? discrepancies in efanesoctocog alfa monitoring
AU  - Agathe Herb
AU  - Jordan Wimmer
AU  - Sarra Derbal
AU  - Manon Dolt
AU  - Léa Pierre
AU  - Olivier Feugeas
AU  - Laurent Mauvieux
AU  - Dominique Desprez
AU  - Laurent Sattler
PY  - 2026
DA  - 2026/10/01
JO  - Research and practice in thrombosis and haemostasis
DO  - 10.1016/j.rpth.2026.106994
UR  - https://europepmc.org/article/PMC/PMC13717027
AB  - Background  Efanesoctocog alfa (EFA) is an ultraextended half-life factor VIII (FVIII) with a unique structure that may influence laboratory monitoring. Although assay discrepancies have been described in spiked samples, data in patient samples are limited.Objectives  To assess the performance of seven FVIII assays for measuring EFA in patient plasma and compare results with those from spiked samples.Methods  FVIII activity was measured using four one-stage assays (OSAs) and three chromogenic assays (CSAs) on two analyzers. Assays were calibrated either with human plasma or an EFA-specific calibrator. Measurements were performed in FVIII-deficient plasma spiked with EFA and in samples from patients with hemophilia A treated with EFA. Results were compared with a reference assay (Actin FSL) and expressed as recovery percentages (acceptable range: 80%-120%).Results  In spiked samples, plasma-calibrated CSA overestimated FVIII activity (+37% to +94%), whereas EFA-calibrated CSA showed good agreement with the reference assay (+2% to +16%). OSA results were generally acceptable, with slight underestimation depending on the reagent. In patient samples, patterns differed markedly: most assays deviated significantly from the reference, and none remained within the acceptable recovery range across the full FVIII spectrum. Compared with spiked samples, discrepancies were more pronounced at later time points and trough levels, where OSA tended to overestimate FVIII activity, whereas EFA-calibrated CSA underestimated activity.Conclusion  Assay performance for EFA differs between spiked and patient samples, underscoring the limitations of in vitro evaluations. No assay provided consistent agreement with the reference assay across the full FVIII range, highlighting the need for cautious interpretation and improved standardization of EFA monitoring in clinical practice.
SN  - 2475-0379
ER  -