Dataset and reproducibility package for spectral-window ATR-MIR milk-protein predictions during passive warming of reconstituted skim milk.

  1. 1University of Southern Queensland, 487-535 West Street, Toowoomba, Queensland, 4350, Australia.

Abstract

This data article documents the calibration, spectral, laboratory, prediction, diagnostic and computational records generated for an attenuated total reflectance mid-infrared (ATR-MIR) data collection for protein indications in reconstituted skim milk. Eight independently prepared calibration formulations were represented by one retained spectrum per formulation and by Kjeldahl protein values obtained from separate laboratory aliquots drawn from the same preparations. One additional physical preparation was measured by three technical scans at 4.10 °C and three at 20.70 °C, and a second additional preparation was measured sequentially during one passive-warming trajectory at nominal milestones of 2, 5, 8, 11, 14, 17 and 20 °C. Neither temperature-experiment preparation was Kjeldahl analysed; the 1.610% m/m value reported for the separate S2W calibration preparation is therefore identified as a formulation-associated comparator. Open-format CSV exports contain 29 spectral matrices and 52,122 long-format rows, alongside the controlled MATLAB workspace, seven one-latent-variable partial least-squares models, leave-one-formulation-out outputs, prediction tables, figure-source data, laboratory reports, software records, checksums and successful clean-run evidence. The files can be reused to inspect the spectra without proprietary software, test alternative spectral-window or calibration workflows, reproduce the deposited calculations, and examine within-trajectory temperature-associated variation and fixed-condition technical-scan dispersion within the stated experimental boundaries.

Related

  1. Interpreting Public-Dataset SOH Benchmarks: A Systematic Review of Evaluation Protocols and Reproducibility Evidence
    Qingde Li, Renqun Wang
    BatteriesPeer-reviewedCC BY 4.0Electrochemistry & energyCrossref
  2. Interaction of fibrinogen–magnetic nanoparticle bioconjugates with integrin reconstituted into artificial membranes
    Ulrike Martens, Una Janke, Sophie Möller +3
    NanoscalePeer-reviewedCC BYFull text19 citedMaterials & polymersHAL
  3. Liquid State 31P NMR as a Tool to Leverage the Understanding of Reconstituted Milk Systems
    Pernille Koch, Dennis W. Juhl, Kirsten G. Malmos +1
    ChemRxivPreprintCC BYFull textGeneral chemistryChemRxiv
  4. Reproducibility, Benchmarks, and Mechanistic Diversity: Building Trust in Computational Electrocatalysis for the AI Era
    David Hernández-Castillo, Kai S. Exner
    ChemRxivPreprintLicence not statedGeneral chemistryChemRxiv
  5. Rigor & Reproducibility in Electrocatalysis: Best Practices for Kinetic Studies
    Jason Adams, Jason S. Bates, David Raciti +11
    ChemRxivPreprintCC BYFull text1 citedElectrochemistry & energyChemRxiv
BibTeX
@article{stefanidis2026dataset,
  title = {{Dataset and reproducibility package for spectral-window ATR-MIR milk-protein predictions during passive warming of reconstituted skim milk.}},
  author = {Kris Stefanidis and Tobias Low and Derek Long and John Billingsley},
  journal = {Data in brief},
  year = {2026},
  month = aug,
  doi = {10.1016/j.dib.2026.113326},
  url = {https://europepmc.org/article/MED/42926983},
  note = {CC BY},
}
RIS
TY  - JOUR
TI  - Dataset and reproducibility package for spectral-window ATR-MIR milk-protein predictions during passive warming of reconstituted skim milk.
AU  - Kris Stefanidis
AU  - Tobias Low
AU  - Derek Long
AU  - John Billingsley
PY  - 2026
DA  - 2026/08/01
JO  - Data in brief
DO  - 10.1016/j.dib.2026.113326
UR  - https://europepmc.org/article/MED/42926983
AB  - This data article documents the calibration, spectral, laboratory, prediction, diagnostic and computational records generated for an attenuated total reflectance mid-infrared (ATR-MIR) data collection for protein indications in reconstituted skim milk. Eight independently prepared calibration formulations were represented by one retained spectrum per formulation and by Kjeldahl protein values obtained from separate laboratory aliquots drawn from the same preparations. One additional physical preparation was measured by three technical scans at 4.10 °C and three at 20.70 °C, and a second additional preparation was measured sequentially during one passive-warming trajectory at nominal milestones of 2, 5, 8, 11, 14, 17 and 20 °C. Neither temperature-experiment preparation was Kjeldahl analysed; the 1.610% m/m value reported for the separate S2W calibration preparation is therefore identified as a formulation-associated comparator. Open-format CSV exports contain 29 spectral matrices and 52,122 long-format rows, alongside the controlled MATLAB workspace, seven one-latent-variable partial least-squares models, leave-one-formulation-out outputs, prediction tables, figure-source data, laboratory reports, software records, checksums and successful clean-run evidence. The files can be reused to inspect the spectra without proprietary software, test alternative spectral-window or calibration workflows, reproduce the deposited calculations, and examine within-trajectory temperature-associated variation and fixed-condition technical-scan dispersion within the stated experimental boundaries.
SN  - 2352-3409
ER  -